510(k) Report

510(k) · Cleared

Cleared

Disposable Medical Nitrile Examination Gloves (Non sterile)

K212639510(k) TraditionalCleared Dec 8, 2021110-day reviewClass I21 CFR 880.6250

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Jiangsu Nanfang Medical Co., Ltd. · LZC Medical Glove, Specialty

Cleared on Dec 8, 2021 as substantially equivalent to K172525 (Central Medicare Sdn Bhd, 2017) after a 110-day review. Its family tree reaches back 5 generations to 2004.

Decided
Dec 8, 2021
Review
110 days
Class
Class I
Product code
LZC
Recalls
None
Predicates
1

Its longest line of descent runs back 5 generations to K042805 (2004). 6 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 5 generations backK042805 Powder Free Nitrile Patient Examination Gloves, Blue Colored…Perusahaan Getah Asas Sdn. Bhd. · 2004
  2. 4 generations backK081089 Kimberly-Clark Sterling Nitrile & Nitrile-Xtra Powder-Free Exam…Kimberly-Clark Corp. · 2008
  3. 3 generations backK141982 Dermagrip Powder Free Blue Nitrile Patient Examination Gloves…Wrp Asia Pacific Sdn. Bhd. · 2015
  4. 2 generations backK163146 POWDER FREE Blue Nitrile GLOVES, Tested for Use with Chemotherapy…Hebei Hongsen Plastics Technology Co, Ltd. · 2017
  5. 1 generation backK172525 Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for…Central Medicare Sdn Bhd · 2017
  6. This deviceK212639 Disposable Medical Nitrile Examination Gloves (Non sterile)Jiangsu Nanfang Medical Co., Ltd. · 2021
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K042805Powder Free Nitrile Patient Examination Gloves, Blue Colored Non-SterilePerusahaan Getah Asas Sdn. Bhd.2004—
−4K081089Kimberly-Clark Sterling Nitrile & Nitrile-Xtra Powder-Free Exam Gloves with…Kimberly-Clark Corp.2008—
−3K141982Dermagrip Powder Free Blue Nitrile Patient Examination Gloves, Non-Sterile…Wrp Asia Pacific Sdn. Bhd.2015—
−2K163146POWDER FREE Blue Nitrile GLOVES, Tested for Use with Chemotherapy DrugsHebei Hongsen Plastics Technology Co, Ltd.2017—
−1K172525Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with…Central Medicare Sdn Bhd2017—
0 (this device)K212639Disposable Medical Nitrile Examination Gloves (Non sterile)Jiangsu Nanfang Medical Co., Ltd.2021—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200510
FIG. 1 — Lineage of K212639, 2004–2021Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LZC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LZC device

  1. #1 K201390 Medline Powder-Free Light Blue Nitrile Exam…score 19.2 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K172525Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy…
Summary page 4
…tion Gloves Device Class: Class I Device Classification Number: 21 CFR 880.6250 Product Code: LZC LZA 3. IDENTIFICATION OF THE LEGALLY MARKETED DEVICE Predicate Device: K172525 Applicant: Central Medicare Sdn. Bhd Device Name: Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with Che…
Central Medicare Sdn Bhd2017—

Also cleared against K172525: 4 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 17 years and passes through 6 companies; a new generation was cleared every 3.6 years on median.
  • None of the 6 ancestors has a recall.
  • 4 other devices also cite its predicate K172525.

How does its category compare?

All LZC devices

LZC Medical Glove, Specialty: FDA decisions, review times, recalls and adverse event reports over 20 years. 2021, this device’s year, is highlighted.

FDA decisions per year · LZC

1in 2025

2007: 12008: 32009: 52010: 52012: 42015: 12016: 32017: 62018: 62019: 72020: 32021: 32022: 72025: 12026 (partial year): 1
2007high 7 (2019) · low 0 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LZC

85in 2022

2008: 166 days median2009: 253 days median2010: 68 days median2012: 149 days median2016: 258 days median2017: 121 days median2018: 105 days median2019: 200 days median2020: 167 days median2021: 94 days median2022: 85 days median
2007high 258 (2016) · low 68 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LZC

1in 2021

2008: 12011: 42021: 12026 (partial year): 1
2007high 4 (2011) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LZC

10in 2025

2009: 12011: 12012: 12013: 12014: 32016: 52017: 72018: 62019: 82020: 12021: 22022: 22023: 12024: 162025: 10
2007high 16 (2024) · low 0 (2007)2025

MAUDE reports naming product code LZC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—00
2008316610
2009525301
201056800
20110—41
2012414901
20130—01
20140—03
20151—00
2016325805
2017612107
2018610506
2019720008
2020316701
202139412
202278502
20230—01
20240—016
20251—010
20261—1—

What is it indicated for?

Disposable Medical Nitrile Examination Gloves (Non sterile) are a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. In addition, these patient examination gloves were tested for use with the following drug concentrations per ASTM D6978-05(2019): The following drugs had NO breakthrough detected up to 240 minutes: Busulfan, Fluorouracil, 6 mg/ml (6,000 ppm) 50.0 mg/ml…

Read the full statement (183 words)

Disposable Medical Nitrile Examination Gloves (Non sterile) are a disposable device intended for medical

purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

In addition, these patient examination gloves were tested for use with the following drug concentrations per

ASTM D6978-05(2019):

The following drugs had NO breakthrough detected up to 240 minutes:

Busulfan, Fluorouracil,

6 mg/ml (6,000 ppm) 50.0 mg/ml (50,000 ppm)

Carboplatin /Paraplatin, lfosfamide,

10 mg/ml (10,000 ppm) 50 mg/ml (50,000 ppm)

Cisplatin, Mechlorethamine HCI,

1.0 mg/ml (1,000 ppm) 1 mg/ml (1,000 ppm)

Cyclophosphamide (Cytoxan), Methotrexate,

20.0 mg/ml (20,000 ppm) 25 mg/ml (25,000 ppm)

Cytarabine HCI, Mitomycin,

100 mg/ml (100,000 ppm) 0.5 mg/ml (500 ppm)

Dacarbazine, Mitoxantrone,

10.0 mg/ml (10,000 ppm) 2 mg/ml (2,000 ppm)

Docetaxel, Paclitaxel,

10 mg/ml (10,000 ppm) 6.0 mg/ml (6,000 ppm)

Doxorubicin HCI, Vincristine Sulfate,

2.0 mg/ml (2,000 ppm) 1 mg/ml (1,000 ppm)

Etoposide,

20.0 mg/ml (20,000 ppm)

Please note that the following drug has low permeation times of less than 60 minutes:

Carmustine (BCNU) 3.3 mg/ml (3,300 ppm), 44.7minutes

ThioTepa10.0 mg/ml (10,000 ppm),35.7 minutes

WARNING: Do not use Carmustine and Thiotepa

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K212639: Disposable Medical Nitrile Examination Gloves (Non sterile).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K212639

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026