510(k) · Cleared
ClearedOlympus PK Electrosurgical Instruments, PK J-hook, PK Spatula, PK Needle
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Gyrus Acmi, Inc. · GEI Electrosurgical, Cutting & Coagulation & Accessories
Cleared on Jan 4, 2022 as substantially equivalent to K142154 (Gyrus Acmi, Inc., 2014) and 2 other predicates after a 27-day review. Its family tree reaches back 11 generations to 1994.
- Decided
- Jan 4, 2022
- Review
- 27 days
- Class
- Class II
- Product code
- GEI
- Recalls
- None
- Predicates
- 3
Its longest line of descent runs back 11 generations to K936109 (1994). 73 ancestors in all, the oldest K800617 (1980). No later device cites it as a predicate yet.
- Root · 11 generations backK936109 Sharplan Model 1120 CO2 Surgical LaserSharplan Lasers, Inc. · 1994
- 7 generations between
- 3 generations backK041285 Gyrus G3 System (Generator, Assessories, Somnoplasty Electrodes and…Gyrus Medical, Ltd. · 2004 · Recalled
- 2 generations backK050460 Gyrus Plasmacision and Plasmablend ElectrodesGyrus Medical, Inc. · 2005
- 1 generation backK142350 PK J-HookGyrus Acmi, Inc. · 2015 · Recalled
- This deviceK213831 Olympus PK Electrosurgical Instruments, PK J-hook, PK Spatula, PK…Gyrus Acmi, Inc. · 2022
Show all 11 generations
- Root · 11 generations backK936109 Sharplan Model 1120 CO2 Surgical LaserSharplan Lasers, Inc. · 1994
- 10 generations backK960521 Sharplan Lasers, Inc. Silktouch Flash ScannerSharplan Lasers, Inc. · 1996
- 9 generations backK963339 Coherent Ultrapulse CO2 Surgical LasersCoherent Medical Group · 1997
- 8 generations backK974789 Coherent Ultrapulse S Series CO2 Surgical Lasers Model C and LLumenis, Inc. · 1998
- 7 generations backK990947 Modified Versapulse Select Single Wavelength, HO: Yag and Dual…Lumenis, Inc. · 1999
- 6 generations backK011703 Lumenis Versapulse Powersuite Holmium (HO:Yag) and Dual Wavelength…Lumenis · 2001 · Recalled
- 5 generations backK014290 Entec Plasma WandsArthrocare Corp. · 2002
- 4 generations backK021777 Gyrus Ent Cogent 2/G2 System (Generator,Accessories Somnoplasty…Gyrus Ent LLC · 2002
- 3 generations backK041285 Gyrus G3 System (Generator, Assessories, Somnoplasty Electrodes and…Gyrus Medical, Ltd. · 2004 · Recalled
- 2 generations backK050460 Gyrus Plasmacision and Plasmablend ElectrodesGyrus Medical, Inc. · 2005
- 1 generation backK142350 PK J-HookGyrus Acmi, Inc. · 2015 · Recalled
- This deviceK213831 Olympus PK Electrosurgical Instruments, PK J-hook, PK Spatula, PK…Gyrus Acmi, Inc. · 2022
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −11 | K936109 | Sharplan Model 1120 CO2 Surgical Laser | Sharplan Lasers, Inc. | 1994 | — |
| −10 | K960521 | Sharplan Lasers, Inc. Silktouch Flash Scanner | Sharplan Lasers, Inc. | 1996 | — |
| −9 | K963339 | Coherent Ultrapulse CO2 Surgical Lasers | Coherent Medical Group | 1997 | — |
| −8 | K974789 | Coherent Ultrapulse S Series CO2 Surgical Lasers Model C and L | Lumenis, Inc. | 1998 | — |
| −7 | K990947 | Modified Versapulse Select Single Wavelength, HO: Yag and Dual Wavelength, HO… | Lumenis, Inc. | 1999 | — |
| −6 | K011703 | Lumenis Versapulse Powersuite Holmium (HO:Yag) and Dual Wavelength… | Lumenis | 2001 | 2 |
| −5 | K014290 | Entec Plasma Wands | Arthrocare Corp. | 2002 | — |
| −4 | K021777 | Gyrus Ent Cogent 2/G2 System (Generator,Accessories Somnoplasty Electrodes and… | Gyrus Ent LLC | 2002 | — |
| −3 | K041285 | Gyrus G3 System (Generator, Assessories, Somnoplasty Electrodes and… | Gyrus Medical, Ltd. | 2004 | 2 |
| −2 | K050460 | Gyrus Plasmacision and Plasmablend Electrodes | Gyrus Medical, Inc. | 2005 | — |
| −1 | K142350 | PK J-Hook | Gyrus Acmi, Inc. | 2015 | 1 |
| 0 (this device) | K213831 | Olympus PK Electrosurgical Instruments, PK J-hook, PK Spatula, PK Needle | Gyrus Acmi, Inc. | 2022 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1980 | 1 | 0 |
| 1986 | 1 | 0 |
| 1989 | 1 | 0 |
| 1990 | 1 | 0 |
| 1991 | 4 | 0 |
| 1992 | 3 | 0 |
| 1993 | 1 | 0 |
| 1994 | 4 | 0 |
| 1995 | 7 | 0 |
| 1996 | 8 | 0 |
| 1997 | 4 | 0 |
| 1998 | 6 | 0 |
| 1999 | 1 | 0 |
| 2000 | 10 | 0 |
| 2001 | 3 | 0 |
| 2002 | 2 | 0 |
| 2003 | 2 | 0 |
| 2004 | 1 | 0 |
| 2014 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in GEI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new GEI device
- #1 K231790 The InMode System with the Morpheus8 Applicatorsscore 32.7 · 21 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (3)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K142154 | PK NeedleSummary page 4…Devices The Olympus PK Electrosurgical Instruments: PK Needle (model # PK-NE0533), PK Spatula (model # PK-SP0533), and PK J-Hook (model # PK-JH0533), were cleared under K142154 (PK Needle), K142289 (PK Spatula), and K142350 (PK J-Hook). This submission is a modification to the device labeling to add compa… | 2014 |
| * | K142289 | PK SpatulaSummary page 4…PK Electrosurgical Instruments: PK Needle (model # PK-NE0533), PK Spatula (model # PK-SP0533), and PK J-Hook (model # PK-JH0533), were cleared under K142154 (PK Needle), K142289 (PK Spatula), and K142350 (PK J-Hook). This submission is a modification to the device labeling to add compatibility with an Olym… | 2014 |
| * | K142350 | PK J-HookSummary page 4…ents: PK Needle (model # PK-NE0533), PK Spatula (model # PK-SP0533), and PK J-Hook (model # PK-JH0533), were cleared under K142154 (PK Needle), K142289 (PK Spatula), and K142350 (PK J-Hook). This submission is a modification to the device labeling to add compatibility with an Olympus generator, ESG-410 cle… | 2015 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 28 years and passes through 9 companies; a new generation was cleared every 1.1 years on median.
- 4 of the 73 ancestors have a recall of any class.
How does its category compare?
All GEI devicesGEI Electrosurgical, Cutting & Coagulation & Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years. 2022, this device’s year, is highlighted.
FDA decisions per year · GEI
52in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · GEI
113in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · GEI
4in 2025
Class I (14 since 2007) Class II/III not classified
Adverse event reports per year · GEI
7,592in 2025
MAUDE reports naming product code GEI, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 50 | 51 | 21 | 2,795 |
| 2008 | 61 | 71 | 19 | 2,556 |
| 2009 | 42 | 130 | 27 | 3,591 |
| 2010 | 46 | 109 | 51 | 3,377 |
| 2011 | 35 | 135 | 32 | 4,463 |
| 2012 | 39 | 133 | 30 | 5,216 |
| 2013 | 48 | 136 | 13 | 4,897 |
| 2014 | 49 | 103 | 3 | 4,908 |
| 2015 | 47 | 123 | 4 | 4,936 |
| 2016 | 64 | 108 | 3 | 5,616 |
| 2017 | 52 | 72 | 2 | 6,073 |
| 2018 | 47 | 93 | 6 | 6,545 |
| 2019 | 48 | 93 | 5 | 7,067 |
| 2020 | 59 | 105 | 1 | 6,490 |
| 2021 | 46 | 146 | 6 | 6,354 |
| 2022 | 55 | 150 | 49 | 6,103 |
| 2023 | 65 | 104 | 27 (4) | 7,176 |
| 2024 | 59 | 125 | 22 (10) | 7,897 |
| 2025 | 52 | 113 | 4 | 7,592 |
| 2026 | 43 | 183 | 8 | — |
What is it indicated for?
1. The PK Needle is indicated for resection of soft tissue in laparoscopic and general surgical procedures.
2. The PK Spatula is indicated for resection and coagulation of soft tissue and blood vessels in laparoscopic and general surgical procedures.
3. The PK J-Hook is indicated for resection and coagulation of soft tissue and blood vessels in laparoscopic and general surgical procedures.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K213831 · printed Nov 4, 2026
Cite this page
510(k) Report. “K213831: Olympus PK Electrosurgical Instruments, PK J-hook, PK Spatula, PK Needle.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K213831
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026