510(k) Report

510(k) · Cleared

Cleared

Viz SDH

K220439510(k) TraditionalCleared Jul 25, 2022159-day reviewClass II21 CFR 892.2080

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Viz. Ai, Inc. · QAS Radiological Computer-Assisted Triage And Notification Software

Cleared on Jul 25, 2022 as substantially equivalent to K210209 (Viz. Ai, Inc., 2021) after a 159-day review. Its family tree reaches back 4 generations to 2018. 1 later device descends from it.

Decided
Jul 25, 2022
Review
159 days
Class
Class II
Product code
QAS
Recalls
None
Predicates
1

Its longest line of descent runs back 4 generations to DEN170073 (2018). 1 later device descends from it, 1 citing it directly.

Descendants, by year cleared 2025: 1 descendant clearedDEN170073 · ContaCT · Viz.Al, Inc. · 2018K180647 · BriefCase · Aidoc Medical, Ltd. · 2018K193658 · Viz ICH · Viz. Ai, Inc. · 2020K210209 · Viz ICH · Viz. Ai, Inc. · 2021201620182020202220242026
  1. Root · 4 generations backDEN170073 ContaCTViz.Al, Inc. · 2018
  2. 3 generations backK180647 BriefCaseAidoc Medical, Ltd. · 2018
  3. 2 generations backK193658 Viz ICHViz. Ai, Inc. · 2020
  4. 1 generation backK210209 Viz ICHViz. Ai, Inc. · 2021
  5. This deviceK220439 Viz SDHViz. Ai, Inc. · 2022
  6. Descendant · cites it directly · 0 of its ownK243611 JLK-SDHJLK, Inc. · 2025

Most influential descendants

Showing 1 of 1 · See all
  • K243611JLK-SDHJLK, Inc. · 2025 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4DEN170073ContaCTViz.Al, Inc.2018—
−3K180647BriefCaseAidoc Medical, Ltd.2018—
−2K193658Viz ICHViz. Ai, Inc.2020—
−1K210209Viz ICHViz. Ai, Inc.2021—
0 (this device)K220439Viz SDHViz. Ai, Inc.2022—
+1K243611JLK-SDHJLK, Inc.2025—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202510
FIG. 1 — Lineage of K220439, 2018–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in QAS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new QAS device

  1. #1 K203508 BriefCasescore 25.5 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K210209Viz ICH
Summary page 4
…Manufacturer Device Name Application No. Viz.ai, Inc. Viz ICH K210209 Device Description Viz SDH is a software-only, parallel workflow tool for use by hospital networks and trained clin…
Viz. Ai, Inc.2021—

Also cleared against K210209: 1 other device →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K243611JLK-SDHJLK, Inc.20250

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 4 years and passes through 3 companies; a new generation was cleared every 1.3 years on median.
  • None of the 4 ancestors has a recall.
  • Of its descendants, K243611 (JLK, Inc., 2025) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate K210209.

How does its category compare?

All QAS devices

QAS Radiological Computer-Assisted Triage And Notification Software: FDA decisions, review times, recalls and adverse event reports over 20 years. 2022, this device’s year, is highlighted.

FDA decisions per year · QAS

15in 2025

2018: 32019: 52020: 122021: 52022: 152023: 112024: 122025: 152026 (partial year): 11
2007high 15 (2022) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · QAS

101in 2025

2019: 111 days median2020: 112 days median2021: 86 days median2022: 116 days median2023: 98 days median2024: 136 days median2025: 101 days median2026 (partial year): 104 days median
2007high 136 (2024) · low 86 (2021)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · QAS

No recall names QAS since 2007.

Adverse event reports per year · QAS

No MAUDE report names QAS since 2007.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20130—00
20140—00
20150—00
20160—00
20170—00
20183—00
2019511100
20201211200
202158600
20221511600
2023119800
20241213600
20251510100
2026111040—

What is it indicated for?

Viz SDH is a notification-only, parallel workflow tool for use by hospital networks and trained clinicians to identify and communicate images of specific patients to a specialist, independent of standard of care workflow. Viz SDH uses an artificial intelligence algorithm to analyze images for findings suggestive of a prespecified clinical condition and to notify an appropriate medical specialist of these findings in parallel to standard of care image…

Read the full statement (207 words)

Viz SDH is a notification-only, parallel workflow tool for use by hospital networks and trained clinicians to identify and communicate images of specific patients to a specialist, independent of standard of care workflow.

Viz SDH uses an artificial intelligence algorithm to analyze images for findings suggestive of a prespecified clinical condition and to notify an appropriate medical specialist of these findings in parallel to standard of care image interpretation. Identification of suspected findings is not for diagnostic use beyond notification. Specifically, the device analyzes non-contrast CT images of the head for subdural hemorrhage and sends notifications to a neurovascular or neurosurgical specialist that a suspected subdural hemorrhage has been identified and recommends review of those images. Images can be previewed through a mobile application.

Images that are previewed through the mobile application may be compressed and are for informational purposes only and not intended for diagnostic use beyond notification. Notified clinicians are responsible for viewing non-compressed images on a diagnostic viewer and engaging in appropriate patient evaluation and relevant discussion with a treating physician before making care-related decisions or requests. Viz SDH is limited to analysis of imaging data and should not be used in-lieu of full patient evaluation or relied upon to make or confirm diagnosis.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K220439: Viz SDH.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K220439

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026