510(k) Report

510(k) · Cleared

Recalled · Class II · 2022

TriMed Ripcord Device

K220650510(k) TraditionalCleared Jun 1, 202286-day reviewClass II21 CFR 888.3030Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

TriMed, Inc. · HTN Washer, Bolt Nut

Cleared on Jun 1, 2022 as substantially equivalent to K130033 (Biomet Manufacturing Corp, 2013) and 1 other predicate after an 86-day review. Its family tree reaches back 9 generations to 1990. 1 later device descends from it. It has been recalled once.

Decided
Jun 1, 2022
Review
86 days
Class
Class II
Product code
HTN
Recalls
1
Predicates
2

Its longest line of descent runs back 9 generations to K904436 (1990). 62 ancestors in all, the oldest K792022 (1979). 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2020: 1 other ancestor2011: 1 other ancestor2008: 1 other ancestor2005: 1 other ancestor2007: 1 other ancestor2004: 2 other ancestors2003: 3 other ancestors2001: 7 other ancestors1999: 5 other ancestors1998: 5 other ancestors1996: 7 other ancestors2006: 1 other ancestor1997: 3 other ancestors1994: 2 other ancestors1992: 1 other ancestor1982: 1 other ancestor2002: 2 other ancestors1995: 3 other ancestors1993: 1 other ancestor1990: 1 other ancestor2000: 3 other ancestors1979: 1 other ancestor2025: 1 descendant clearedK904436 · Quickanchor (TM) · Mitek Surgical Products, Inc. · 1990K962397 · Statak Soft Tissue Attachment Device · Zimmer, Inc. · 1996K973015 · Multitak SS Suture System · Bonutti Research · 1997K990156 · Multitak SS Buttons · Bonutti Research · 1999K031666 · Arthrex Fiberwire Button Repair Kit, Model AR-8920DS/AR-8921DS · Arthrex, Inc. · 2003 · recalledK062747 · Arthrex Retrobutton, 15MM, 60MM, Retrobutton, Long, 15MM, 60MM · Arthrex, Inc. · 2006K100652 · Acl Tightrope · Arthrex, Inc. · 2010K112990 · Acl Tightrope,Acl Tightrope Duuble Bundle · Arthrex, Inc. · 2011K130033 · Toggleloc Xtender; Toggleloc XL; Implant Kit Toggleloc; Implant Kit Toggleloc BTB, Extended Toggleloc; Toggleloc; BTB to · Biomet Manufacturing Corp · 2013 · recalled19801990200020102020
  1. Root · 9 generations backK904436 Quickanchor (TM)Mitek Surgical Products, Inc. · 1990
  2. 5 generations between
  3. 3 generations backK100652 Acl TightropeArthrex, Inc. · 2010
  4. 2 generations backK112990 Acl Tightrope,Acl Tightrope Duuble BundleArthrex, Inc. · 2011
  5. 1 generation backK130033 Toggleloc Xtender; Toggleloc XL; Implant Kit Toggleloc; Implant Kit…Biomet Manufacturing Corp · 2013 · Recalled
  6. This deviceK220650 TriMed Ripcord DeviceTriMed, Inc. · 2022
  7. Descendant · cites it directly · 0 of its ownK251134 RipCordTriMed, Inc. · 2025
Show all 9 generations
  1. Root · 9 generations backK904436 Quickanchor (TM)Mitek Surgical Products, Inc. · 1990
  2. 8 generations backK962397 Statak Soft Tissue Attachment DeviceZimmer, Inc. · 1996
  3. 7 generations backK973015 Multitak SS Suture SystemBonutti Research · 1997
  4. 6 generations backK990156 Multitak SS ButtonsBonutti Research · 1999
  5. 5 generations backK031666 Arthrex Fiberwire Button Repair Kit, Model AR-8920DS/AR-8921DSArthrex, Inc. · 2003 · Recalled
  6. 4 generations backK062747 Arthrex Retrobutton, 15MM, 60MM, Retrobutton, Long, 15MM, 60MMArthrex, Inc. · 2006
  7. 3 generations backK100652 Acl TightropeArthrex, Inc. · 2010
  8. 2 generations backK112990 Acl Tightrope,Acl Tightrope Duuble BundleArthrex, Inc. · 2011
  9. 1 generation backK130033 Toggleloc Xtender; Toggleloc XL; Implant Kit Toggleloc; Implant Kit…Biomet Manufacturing Corp · 2013 · Recalled
  10. This deviceK220650 TriMed Ripcord DeviceTriMed, Inc. · 2022

Most influential descendants

Showing 1 of 1 · See all
  • K251134RipCordTriMed, Inc. · 2025 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−9K904436Quickanchor (TM)Mitek Surgical Products, Inc.1990—
−8K962397Statak Soft Tissue Attachment DeviceZimmer, Inc.1996—
−7K973015Multitak SS Suture SystemBonutti Research1997—
−6K990156Multitak SS ButtonsBonutti Research1999—
−5K031666Arthrex Fiberwire Button Repair Kit, Model AR-8920DS/AR-8921DSArthrex, Inc.20031
−4K062747Arthrex Retrobutton, 15MM, 60MM, Retrobutton, Long, 15MM, 60MMArthrex, Inc.2006—
−3K100652Acl TightropeArthrex, Inc.2010—
−2K112990Acl Tightrope,Acl Tightrope Duuble BundleArthrex, Inc.2011—
−1K130033Toggleloc Xtender; Toggleloc XL; Implant Kit Toggleloc; Implant Kit Toggleloc…Biomet Manufacturing Corp20133
0 (this device)K220650TriMed Ripcord DeviceTriMed, Inc.2022—
+1K251134RipCordTriMed, Inc.2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
197910
198210
199010
199210
199310
199420
199530
199670
199730
199850
199950
200030
200170
200220
200330
200420
200510
200610
200710
200810
201110
202010

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202510
FIG. 1 — Lineage of K220650, 1990–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in HTN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HTN device

  1. #1 K201522 Arthrex Syndesmosis TightRope XP Buttress…score 16.4 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K130033Toggleloc Xtender; Toggleloc XL; Implant Kit Toggleloc; Implant Kit Toggleloc BTB…
Summary page 4
…II Product Codes: HTN (a)(3). Predicate Device: Primary Predicate K130033 – ToggleLoc System (ToggleLoc with Ziptight), Biomet (USA) (Primary)…
Biomet Manufacturing Corp20133
*K201522Arthrex Syndesmosis TightRope XP Buttress Plate Implant System
Summary page 4
…Biomet (USA) (Primary) Other Predicate Device(s) K201522 – Arthrex Syndesmosis TightRope XP Buttress Plate Implant System, Arthrex Inc. (USA) (a…
Arthrex, Inc.2020—

Also cleared against K130033: 16 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K251134RipCordTriMed, Inc.20250

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 22, 2022Class IIZ-1101-2023Under Investigation by firmOverly aggressive tension on the pull suture, while passing the medial button through a bone/soft tissue tunnel across the fibula and tibia, can exceed its strength causing it to break.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 32 years and passes through 6 companies; a new generation was cleared every 2.9 years on median.
  • 10 of the 62 ancestors have a recall of any class.
  • Of its descendants, K251134 (TriMed, Inc., 2025) has gone furthest, with 0 descendants of its own.
  • 16 other devices also cite its predicate K130033.

How does its category compare?

All HTN devices

HTN Washer, Bolt Nut: FDA decisions, review times, recalls and adverse event reports over 20 years. 2022, this device’s year, is highlighted.

FDA decisions per year · HTN

4in 2025

2007: 12008: 22009: 32011: 42013: 32014: 82015: 22016: 12017: 62018: 22019: 52020: 32021: 22022: 52023: 22024: 62025: 42026 (partial year): 2
2007high 8 (2014) · low 0 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HTN

75in 2025

2009: 86 days median2011: 130 days median2013: 105 days median2014: 98 days median2017: 88 days median2019: 116 days median2020: 87 days median2022: 86 days median2024: 90 days median2025: 75 days median
2007high 130 (2011) · low 75 (2025)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HTN

1in 2019

2013: 12019: 1
2007high 1 (2013) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · HTN

190in 2025

2007: 142008: 112009: 222010: 172011: 122012: 692013: 452014: 332015: 352016: 432017: 302018: 482019: 532020: 812021: 1492022: 852023: 1082024: 2692025: 190
2007high 269 (2024) · low 11 (2008)2025

MAUDE reports naming product code HTN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—014
20082—011
2009386022
20100—017
20114130012
20120—069
20133105145
2014898033
20152—035
20161—043
2017688030
20182—048
20195116153
2020387081
20212—0149
2022586085
20232—0108
20246900269
20254750190
20262—0—

What is it indicated for?

TriMed Ripcord devices are intended for use to supplement repair or reconstruction during healing of ligament injuries or deficiencies of the extremities.

The TriMed RipCord device is indicated to be used as an adjunct in combination with compatible TriMed fixation implants to provide fixation during the healing process in syndesmotic trauma, such as fixation of syndesmosis disruptions in connection with Weber B and C ankle fractures.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K220650: TriMed Ripcord Device.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K220650

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026