510(k) Report

510(k) · Cleared

Cleared

VITROS Immunodiagnostic Products Intact PTH II Reagent Pack

K221197510(k) TraditionalCleared Sep 5, 2023498-day reviewClass II21 CFR 862.1545

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ortho Clinical Diagnostics · CEW Radioimmunoassay, Parathyroid Hormone

Cleared on Sep 5, 2023 as substantially equivalent to K070709 (Roche Diagnostics Corp., 2007) after a 498-day review. Its family tree reaches back 3 generations to 1995.

Decided
Sep 5, 2023
Review
498 days
Class
Class II
Product code
CEW
Recalls
None
Predicates
1

Its longest line of descent runs back 3 generations to K954418 (1995). No later device cites it as a predicate yet.

  1. Root · 3 generations backK954418 Intact Parathyroid Hormone (PTH) Immunoassay (Modification)Nichols Institute Diagnostics · 1995
  2. 2 generations backK992680 Elecsys Parathyroid Hormone Test SystemRoche Diagnostics Corp. · 1999 · Recalled
  3. 1 generation backK070709 Roche Elecsys PTH Test SystemRoche Diagnostics Corp. · 2007
  4. This deviceK221197 VITROS Immunodiagnostic Products Intact PTH II Reagent PackOrtho Clinical Diagnostics · 2023
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K954418Intact Parathyroid Hormone (PTH) Immunoassay (Modification)Nichols Institute Diagnostics1995—
−2K992680Elecsys Parathyroid Hormone Test SystemRoche Diagnostics Corp.19991
−1K070709Roche Elecsys PTH Test SystemRoche Diagnostics Corp.2007—
0 (this device)K221197VITROS Immunodiagnostic Products Intact PTH II Reagent PackOrtho Clinical Diagnostics2023—
FIG. 1 — Lineage of K221197, 1995–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CEW? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CEW device

  1. #1 K231927 Elecsys PTH , Elecsys PTH STATscore 7.7 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K070709Roche Elecsys PTH Test System
Summary page 4
…VITROS Immunodiagnostic Products Intact PTH II Reagent Device Pack is substantially equivalent to the Roche Elecsys PTH assay, K070709 5. Device Description: The VITROS Immunodiagnostic Products Intact PTH II assay is performed using the VITROS Intact PTH II Rea…
Roche Diagnostics Corp.2007—

Also cleared against K070709: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 28 years and passes through 3 companies; a new generation was cleared every 7.8 years on median.
  • 1 of the 3 ancestors has a recall of any class.
  • 1 other device also cites its predicate K070709.

How does its category compare?

All CEW devices

CEW Radioimmunoassay, Parathyroid Hormone: FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.

FDA decisions per year · CEW

2in 2024

2007: 42010: 12012: 12013: 22014: 12015: 12017: 22019: 12023: 12024: 2
2007high 4 (2007) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CEW

160in 2007

2007: 160 days median
2007high 160 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CEW

1in 2025

2007: 12008: 12010: 32013: 12016: 22022: 22024: 22025: 1
2007high 3 (2010) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CEW

37in 2025

2008: 82009: 42010: 232011: 972012: 1242013: 1722014: 1672015: 572016: 982017: 832018: 652019: 772020: 502021: 572022: 1882023: 822024: 672025: 37
2007high 188 (2022) · low 0 (2007)2025

MAUDE reports naming product code CEW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007416010
20080—18
20090—04
20101—323
20110—097
20121—0124
20132—1172
20141—0167
20151—057
20160—298
20172—083
20180—065
20191—077
20200—050
20210—057
20220—2188
20231—082
20242—267
20250—137
20260—0—

What is it indicated for?

VITROS Immunodiagnostic Products Intact PTH II Reagent Pack quantitatively measures intact parathyroid hormone (iPTH) in human serum and plasma (K2-EDTA, lithium heparin or sodium heparin) using the automated VITROS 3600 Immunodiagnostic System.

Intact PTH is indicated to aid in the diagnosis of hyperparathyroidism, hypoparathyroidism, differential diagnosis of hypocalcemia or hypercalcemia and for intraoperative measurement of iPTH levels.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K221197: VITROS Immunodiagnostic Products Intact PTH II Reagent Pack.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K221197

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026