510(k) Report

510(k) · Cleared

Cleared

Voyant Maryland Fusion device with Single-Step Activation (EB212, EB213, EB214)

K222284510(k) SpecialCleared Oct 14, 202274-day reviewClass II21 CFR 878.4400

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Applied Medical Resources Corporation · GEI Electrosurgical, Cutting & Coagulation & Accessories

Cleared on Oct 14, 2022 as substantially equivalent to K200598 (Applied Medical Resources Corp., 2020) after a 74-day review. Its family tree reaches back 9 generations to 1994.

Decided
Oct 14, 2022
Review
74 days
Class
Class II
Product code
GEI
Recalls
None
Predicates
1

Its longest line of descent runs back 9 generations to K932293 (1994). 17 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 9 generations backK932293 Cabot Medical Bipolar Cutting ForcepsCabot Medical Corp. · 1994 · Recalled
  2. 5 generations between
  3. 3 generations backK182653 Voyant Maryland Fusion DeviceApplied Medical Resources Corp. · 2018
  4. 2 generations backK193292 Voyant 5mm Fusion Device, Voyant Maryland Fusion DeviceApplied Medical Resources Corp. · 2019
  5. 1 generation backK200598 Voyant Maryland Fusion DeviceApplied Medical Resources Corp. · 2020
  6. This deviceK222284 Voyant Maryland Fusion device with Single-Step Activation (EB212…Applied Medical Resources Corporation · 2022
Show all 9 generations
  1. Root · 9 generations backK932293 Cabot Medical Bipolar Cutting ForcepsCabot Medical Corp. · 1994 · Recalled
  2. 8 generations backK981916 Ligasure Vessel Sealing SystemValleylab, Inc. · 1998
  3. 7 generations backK051644 Valleylab Forcetriad Electrosurgical GeneratorValleylab · 2005
  4. 6 generations backK070162 Forcetriad Electrosurgical Generator; Ligasure InstrumentsValleylab,A Division of Tyco Healthcare Group LP · 2007
  5. 5 generations backK141288 Voyant Electrosurgical Generator, Voyant 5MM Laparoscopic…Applied Medical Resources · 2014
  6. 4 generations backK172624 Voyant 5mm Fusion DeviceApplied Medical Resources Corporation · 2018
  7. 3 generations backK182653 Voyant Maryland Fusion DeviceApplied Medical Resources Corp. · 2018
  8. 2 generations backK193292 Voyant 5mm Fusion Device, Voyant Maryland Fusion DeviceApplied Medical Resources Corp. · 2019
  9. 1 generation backK200598 Voyant Maryland Fusion DeviceApplied Medical Resources Corp. · 2020
  10. This deviceK222284 Voyant Maryland Fusion device with Single-Step Activation (EB212…Applied Medical Resources Corporation · 2022
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−9K932293Cabot Medical Bipolar Cutting ForcepsCabot Medical Corp.19943
−8K981916Ligasure Vessel Sealing SystemValleylab, Inc.1998—
−7K051644Valleylab Forcetriad Electrosurgical GeneratorValleylab2005—
−6K070162Forcetriad Electrosurgical Generator; Ligasure InstrumentsValleylab,A Division of Tyco Healthcare Group LP2007—
−5K141288Voyant Electrosurgical Generator, Voyant 5MM Laparoscopic Sealer-Divider DeviceApplied Medical Resources2014—
−4K172624Voyant 5mm Fusion DeviceApplied Medical Resources Corporation2018—
−3K182653Voyant Maryland Fusion DeviceApplied Medical Resources Corp.2018—
−2K193292Voyant 5mm Fusion Device, Voyant Maryland Fusion DeviceApplied Medical Resources Corp.2019—
−1K200598Voyant Maryland Fusion DeviceApplied Medical Resources Corp.2020—
0 (this device)K222284Voyant Maryland Fusion device with Single-Step Activation (EB212, EB213, EB214)Applied Medical Resources Corporation2022—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199540
199610
200110
200310
200910
FIG. 1 — Lineage of K222284, 1994–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GEI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GEI device

  1. #1 K231790 The InMode System with the Morpheus8 Applicatorsscore 32.7 · 21 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K200598Voyant Maryland Fusion Device
Summary page 4
…21 CFR 878.4400 Device Class: Class II Product Code: GEI Legally Marketed Voyant® Maryland Fusion Device Device: 510(k)#: K200598 Produce Code: GEI Reference Device: LigaSure Maryland Jaw Thoracic Sealer/Divider One-step Sealing…
Applied Medical Resources Corp.2020—
†K181085LigaSure Maryland Jaw Thoracic Sealer/Divider One-step Sealing, Nano-coatedCovidien2018—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 28 years and passes through 7 companies; a new generation was cleared every 2.5 years on median.
  • 4 of the 17 ancestors have a recall of any class.

How does its category compare?

All GEI devices

GEI Electrosurgical, Cutting & Coagulation & Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years. 2022, this device’s year, is highlighted.

FDA decisions per year · GEI

52in 2025

2007: 502008: 612009: 422010: 462011: 352012: 392013: 482014: 492015: 472016: 642017: 522018: 472019: 482020: 592021: 462022: 552023: 652024: 592025: 522026 (partial year): 43
2007high 65 (2023) · low 35 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GEI

113in 2025

2007: 51 days median2008: 71 days median2009: 130 days median2010: 109 days median2011: 135 days median2012: 133 days median2013: 136 days median2014: 103 days median2015: 123 days median2016: 108 days median2017: 72 days median2018: 93 days median2019: 93 days median2020: 105 days median2021: 146 days median2022: 150 days median2023: 104 days median2024: 125 days median2025: 113 days median2026 (partial year): 183 days median
2007high 183 (2026) · low 51 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GEI

4in 2025

2007: 212008: 192009: 272010: 512011: 322012: 302013: 132014: 32015: 42016: 32017: 22018: 62019: 52020: 12021: 62022: 492023: 27, 4 Class I2024: 22, 10 Class I2025: 42026 (partial year): 8
2007high 51 (2010) · low 1 (2020)2026

Class I (14 since 2007) Class II/III not classified

Adverse event reports per year · GEI

7,592in 2025

2007: 2,7952008: 2,5562009: 3,5912010: 3,3772011: 4,4632012: 5,2162013: 4,8972014: 4,9082015: 4,9362016: 5,6162017: 6,0732018: 6,5452019: 7,0672020: 6,4902021: 6,3542022: 6,1032023: 7,1762024: 7,8972025: 7,592
2007high 7,897 (2024) · low 2,556 (2008)2025

MAUDE reports naming product code GEI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20075051212,795
20086171192,556
200942130273,591
201046109513,377
201135135324,463
201239133305,216
201348136134,897
20144910334,908
20154712344,936
20166410835,616
2017527226,073
2018479366,545
2019489357,067
20205910516,490
20214614666,354
202255150496,103
20236510427 (4)7,176
20245912522 (10)7,897
20255211347,592
2026431838—

What is it indicated for?

The Voyant Maryland Fusion Device with Single-Step Activation is a bipolar, electrosurgical device indicated for use with the Voyant electrosurgical generator in open and laparoscopic procedures where the ligation and division of vessels and tissue bundles is desired. The device can seal and divide vessels up to and including 7mm in diameter and tissue bundles that can be captured in the jaws of the device.

The device has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K222284: Voyant Maryland Fusion device with Single-Step Activation (EB212, EB213, EB214).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K222284

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026