510(k) Report

510(k) · Cleared

Cleared

Well-Life Probe Electrode for Stimulation/EMG Probe (Model: SA), Well-Life Incontinence Stimulation/EMG Electrode Probe, Well-Life Probe Electrode

K222528510(k) TraditionalCleared Dec 29, 2022129-day reviewClass II21 CFR 876.5320

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Well-Life Healthcare Limited · KPI Stimulator, Electrical, Non-Implantable, For Incontinence

Cleared on Dec 29, 2022 as substantially equivalent to K122194 (Everyway Medical Instruments Co.,Ltd, 2013) after a 129-day review. Its family tree reaches back 4 generations to 1989. 1 later device descends from it.

Decided
Dec 29, 2022
Review
129 days
Class
Class II
Product code
KPI
Recalls
None
Predicates
1

Its longest line of descent runs back 4 generations to K891773 (1989). 6 ancestors in all. 1 later device descends from it, 1 citing it directly.

  1. Root · 4 generations backK891773 Microgyn II Elec Stim Device Treat of Urin IncontiHollister, Inc. · 1989
  2. 3 generations backK970602 Vaginal 2 Electrode Stimulation/Emg Probe-SmallHollister, Inc. · 1997
  3. 2 generations backK971541 Vaginal 2 Electrode Stimulation/Emg Probe - TamponHollister, Inc. · 1997
  4. 1 generation backK122194 Life-Care Vaginal Probe Life-Care Anal ProbeEveryway Medical Instruments Co.,Ltd · 2013
  5. This deviceK222528 Well-Life Probe Electrode for Stimulation/EMG Probe (Model: SA)…Well-Life Healthcare Limited · 2022
  6. Descendant · cites it directly · 0 of its ownK231166 Levina, Levina Incontinence Stimulation Electrodes (Sterile), Pelvic…Zmi Electronics, Ltd. · 2024

Most influential descendants

Showing 1 of 1 · See all
  • K231166Levina, Levina Incontinence Stimulation Electrodes (Sterile), Pelvic Floor Muscle Probe…Zmi Electronics, Ltd. · 2024 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K891773Microgyn II Elec Stim Device Treat of Urin IncontiHollister, Inc.1989—
−3K970602Vaginal 2 Electrode Stimulation/Emg Probe-SmallHollister, Inc.1997—
−2K971541Vaginal 2 Electrode Stimulation/Emg Probe - TamponHollister, Inc.1997—
−1K122194Life-Care Vaginal Probe Life-Care Anal ProbeEveryway Medical Instruments Co.,Ltd2013—
0 (this device)K222528Well-Life Probe Electrode for Stimulation/EMG Probe (Model: SA), Well-Life…Well-Life Healthcare Limited2022—
+1K231166Levina, Levina Incontinence Stimulation Electrodes (Sterile), Pelvic Floor…Zmi Electronics, Ltd.2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199510
199910

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202410
FIG. 1 — Lineage of K222528, 1989–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KPI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KPI device

  1. #1 K262353 BTL-299-Fscore 14.8 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K122194Life-Care Vaginal Probe Life-Care Anal Probe
Summary page 4
…KPI & HIR 5. Identification of the Predicate and Reference Devices: Predicate Device: 510(K) No K122194 Predicate Device Name: LIFE-CARE VAGINAL PROBE, LIFE-CARE ANAL PROBE. The predic…
Everyway Medical Instruments Co.,Ltd2013—
†K191312Perfect PFETenscare, Ltd.2019—

Also cleared against K122194: 1 other device →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K231166Levina, Levina Incontinence Stimulation Electrodes (Sterile), Pelvic Floor Muscle Probe…Zmi Electronics, Ltd.20240

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 33 years and passes through 3 companies; a new generation was cleared every 9.7 years on median.
  • None of the 6 ancestors has a recall.
  • Of its descendants, K231166 (Zmi Electronics, Ltd., 2024) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate K122194.

How does its category compare?

All KPI devices

KPI Stimulator, Electrical, Non-Implantable, For Incontinence: FDA decisions, review times, recalls and adverse event reports over 20 years. 2022, this device’s year, is highlighted.

FDA decisions per year · KPI

7so far in 2026, the most since 2007

2007: 12008: 12009: 22011: 12012: 22013: 12014: 32015: 62016: 32017: 12018: 32019: 22020: 12021: 32022: 32023: 32024: 42025: 42026 (partial year): 7
2007high 7 (2026) · low 0 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KPI

263in 2025

2014: 57 days median2015: 260 days median2016: 89 days median2018: 304 days median2021: 217 days median2022: 181 days median2023: 197 days median2024: 172 days median2025: 263 days median2026 (partial year): 248 days median
2007high 304 (2018) · low 57 (2014)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KPI

No recall names KPI since 2007.

Adverse event reports per year · KPI

2in 2024

2013: 12015: 22017: 42019: 22020: 32021: 12022: 22023: 62024: 2
2007high 6 (2023) · low 0 (2007)2025

MAUDE reports naming product code KPI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—00
20081—00
20092—00
20100—00
20111—00
20122—00
20131—01
201435700
2015626002
201638900
20171—04
2018330400
20192—02
20201—03
2021321701
2022318102
2023319706
2024417202
2025426300
202672480—

What is it indicated for?

Well-Life Probe Electrode for Stimulation/EMG Probe (Model: SA) include Vaginal Probe SA-2876, SA-25145, SA-20100, SA-3478, SA-2687 & Rectal Probe SA-1483, SA-19108, SA-1563, SA-2486 are intended to provide electromyographic feedback from pelvic musculature or electrical stimulation to pelvic musculature for the purpose of rehabilitation of week pelvic floor muscles and restoration of neuromuscular control during the treatment of urinary incontinence.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K222528: Well-Life Probe Electrode for Stimulation/EMG Probe (Model: SA), Well-Life Incontinence Stimulation/EMG Electrode Probe, Well-Life Probe Electrode.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K222528

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026