510(k) Report

510(k) · Cleared

Cleared

EmPro EPS (EP4514C-190, EP6514C-190);Nanoparasol EPS (PNP4514C-190,PNP6514C-190)

K222694510(k) TraditionalCleared Apr 27, 2023233-day reviewClass II21 CFR 870.1250

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

MicroVention, Inc. · NTE Temporary Carotid Catheter For Embolic Capture

Cleared on Apr 27, 2023 as substantially equivalent to K063204 (Ev3, Inc., 2006) after a 233-day review. Its family tree reaches back 5 generations to 2001.

Decided
Apr 27, 2023
Review
233 days
Class
Class II
Product code
NTE
Recalls
None
Predicates
1

Its longest line of descent runs back 5 generations to K003992 (2001). 7 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 5 generations backK003992 Percusurge Guardwire Temporary Occlusion and Aspiration SystemPercu Surge, Inc. · 2001
  2. 4 generations backK023691 Boston Scientific Filterwire EX EmbolicBoston Scientific Epi · 2003
  3. 3 generations backK032884 Filter Wire EZ Embolic Protection System and EZ Bent Tip Retrieval…Boston Scientific Epi · 2004
  4. 2 generations backK052659 Spiderx Embolic Protection DeviceEv3, Inc. · 2006
  5. 1 generation backK063204 Spiderfx Embolic Protection DeviceEv3, Inc. · 2006
  6. This deviceK222694 EmPro EPS (EP4514C-190, EP6514C-190);Nanoparasol EPS…MicroVention, Inc. · 2023
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K003992Percusurge Guardwire Temporary Occlusion and Aspiration SystemPercu Surge, Inc.2001—
−4K023691Boston Scientific Filterwire EX EmbolicBoston Scientific Epi2003—
−3K032884Filter Wire EZ Embolic Protection System and EZ Bent Tip Retrieval SheathBoston Scientific Epi2004—
−2K052659Spiderx Embolic Protection DeviceEv3, Inc.2006—
−1K063204Spiderfx Embolic Protection DeviceEv3, Inc.2006—
0 (this device)K222694EmPro EPS (EP4514C-190, EP6514C-190);Nanoparasol EPS (PNP4514C-190,PNP6514C-190)MicroVention, Inc.2023—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200210
200410
FIG. 1 — Lineage of K222694, 2001–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in NTE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NTE device

  1. #1 K200198 Wirionscore 4.4 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K063204Spiderfx Embolic Protection Device
Summary page 4
…Name: Embolic Protection System Classification Name: 21 CFR 870.1250 Regulatory Class: II Product Code: NTE III. PREDICATE DEVICE SpiderFX® Embolic Protection Device, K063204, EV3, Inc. No reference devices were used in this submission IV. DEVICE DESCRIPTION MicroVention’s Embolic Protection System (E…
Ev3, Inc.2006—

Also cleared against K063204: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 22 years and passes through 4 companies; a new generation was cleared every 1.5 years on median.
  • None of the 7 ancestors has a recall.
  • 2 other devices also cite its predicate K063204.

How does its category compare?

All NTE devices

NTE Temporary Carotid Catheter For Embolic Capture: FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.

FDA decisions per year · NTE

2in 2023

2007: 12008: 32009: 32010: 12011: 42012: 32014: 12015: 32016: 12018: 12019: 12020: 12021: 12022: 12023: 2
2007high 4 (2011) · low 0 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NTE

105in 2015

2008: 95 days median2009: 86 days median2011: 105 days median2012: 30 days median2015: 105 days median
2007high 105 (2011) · low 30 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · NTE

3in 2023

2008: 62011: 12021: 1, 1 Class I2023: 32026 (partial year): 2, 2 Class I
2007high 6 (2008) · low 0 (2007)2026

Class I (3 since 2007) Class II/III not classified

Adverse event reports per year · NTE

405in 2025

2007: 1192008: 3682009: 3432010: 2942011: 3042012: 2612013: 2582014: 1992015: 1482016: 1482017: 1412018: 1582019: 2332020: 2382021: 3272022: 3262023: 3542024: 3982025: 405
2007high 405 (2025) · low 119 (2007)2025

MAUDE reports naming product code NTE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—0119
20083956368
20093860343
20101—0294
201141051304
20123300261
20130—0258
20141—0199
201531050148
20161—0148
20170—0141
20181—0158
20191—0233
20201—0238
20211—1 (1)327
20221—0326
20232—3354
20240—0398
20250—0405
20260—2 (2)—

What is it indicated for?

The EmPro™ EPS/Nanoparasol™ EPS is indicated for use as a guidewire to contain and remove embolic material (thrombus/debris) while performing angioplasty and stenting procedures in carotid arteries. The diameter of the artery at the site of the filter placement should be between 3.0 and 6.5 mm.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K222694: EmPro EPS (EP4514C-190, EP6514C-190);Nanoparasol EPS (PNP4514C-190,PNP6514C-190).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K222694

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026