510(k) Report

510(k) · Cleared

Recalled · Class II · 2024

Renasys Edge (66803126)

K223041510(k) TraditionalCleared Nov 22, 202254-day reviewClass II21 CFR 878.4780Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Smith & Nephew Medical Limited · OMP Negative Pressure Wound Therapy Powered Suction Pump

Cleared on Nov 22, 2022 as substantially equivalent to K181822 (Smith & Nephew Medical Limited, 2019) after a 54-day review. Its family tree reaches back 12 generations to 1995. 2 later devices descend from it. It has been recalled 3 times.

Decided
Nov 22, 2022
Review
54 days
Class
Class II
Product code
OMP
Recalls
3
Predicates
1

Its longest line of descent runs back 12 generations to K945062 (1995). 31 ancestors in all, the oldest K933902 (1994). 2 later devices descend from it, 1 citing it directly.

Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor3 generations back: 2 other ancestors4 generations back: 3 other ancestors5 generations back: 4 other ancestors6 generations back: 6 other ancestors7 generations back: 3 other ancestors 1 generation forward: 1 descendant2 generations forward: 1 descendant K181822 · RENASYS Touch; RENASYS Y-Connector · Smith & Nephew Medical Limited · 2019K143133 · RENASYS¿ EZ MAX Negative Pressure Wound Therapy Device, RENASYS¿ AB Abdominal Dressing Kit with Soft Port Device · Smith & Nephew, Inc. · 2015K142979 · RENASYS; EZ MAX Negative Pressure Wound Therapy; Foam NPWT Wound Dressing Kits with Soft Port; Gauze NPWT Wound Dressing Kits with Soft Port; EZ PLUS/EZ MAX Canisters · Smith & Nephew, Inc. · 2015 · recalledK132446 · Renasys EZ Max Negative Pressure Wound Therapy Device · Smith & Nephew, Inc. · 2013K102001 · Modification to Renasys EZ Plus Negative Pressure Wound Therapy · Smith & Nephew, Inc. · 2010 · recalledK091470 · Ranasys EZ Negative Pressure Wound Therapy System, Model 66800059 · Smith & Nephew, Inc. · 2009 · recalledK083375 · Renasys GO Negative Pressure Wound Therapy · Smith & Nephew, Inc. · 2009 · recalledK082426 · Renasys EZ · Smith & Nephew, Inc. · 2008 · recalledK061919 · Versatile 1 Ezcare Wound Vacuum System · Blue Sky Medical Group Incorporated · 2007 · recalledK052456 · Modification to Versatile 1 Wound Vacuum System · Blue Sky Medical Group Incorporated · 2005K042134 · Versatile 1 Wound Vacuum System · Blue Sky Medical Group Incorporated · 2004 · recalledK945062 · Vacplus · Kinetic Concepts, Inc. · 1995−10−50
  1. Root · 12 generations backK945062 VacplusKinetic Concepts, Inc. · 1995
  2. 8 generations between
  3. 3 generations backK142979 RENASYS; EZ MAX Negative Pressure Wound Therapy; Foam NPWT Wound…Smith & Nephew, Inc. · 2015 · Recalled
  4. 2 generations backK143133 RENASYS¿ EZ MAX Negative Pressure Wound Therapy Device, RENASYS¿ AB…Smith & Nephew, Inc. · 2015
  5. 1 generation backK181822 RENASYS Touch; RENASYS Y-ConnectorSmith & Nephew Medical Limited · 2019
  6. This deviceK223041 Renasys Edge (66803126)Smith & Nephew Medical Limited · 2022
  7. Descendant · cites it directly · 1 of its ownK231939 Renasys EdgeSmith & Nephew Medical, Ltd. · 2024
  8. Descendant · 2 generations on · 0 of its ownK260646 Renasys Edge (66803126)Smith & Nephew Medical, Ltd. · 2026
Show all 12 generations
  1. Root · 12 generations backK945062 VacplusKinetic Concepts, Inc. · 1995
  2. 11 generations backK042134 Versatile 1 Wound Vacuum SystemBlue Sky Medical Group Incorporated · 2004 · Recalled
  3. 10 generations backK052456 Modification to Versatile 1 Wound Vacuum SystemBlue Sky Medical Group Incorporated · 2005
  4. 9 generations backK061919 Versatile 1 Ezcare Wound Vacuum SystemBlue Sky Medical Group Incorporated · 2007 · Recalled
  5. 8 generations backK082426 Renasys EZSmith & Nephew, Inc. · 2008 · Recalled
  6. 7 generations backK083375 Renasys GO Negative Pressure Wound TherapySmith & Nephew, Inc. · 2009 · Recalled
  7. 6 generations backK091470 Ranasys EZ Negative Pressure Wound Therapy System, Model 66800059Smith & Nephew, Inc. · 2009 · Recalled
  8. 5 generations backK102001 Modification to Renasys EZ Plus Negative Pressure Wound TherapySmith & Nephew, Inc. · 2010 · Recalled
  9. 4 generations backK132446 Renasys EZ Max Negative Pressure Wound Therapy DeviceSmith & Nephew, Inc. · 2013
  10. 3 generations backK142979 RENASYS; EZ MAX Negative Pressure Wound Therapy; Foam NPWT Wound…Smith & Nephew, Inc. · 2015 · Recalled
  11. 2 generations backK143133 RENASYS¿ EZ MAX Negative Pressure Wound Therapy Device, RENASYS¿ AB…Smith & Nephew, Inc. · 2015
  12. 1 generation backK181822 RENASYS Touch; RENASYS Y-ConnectorSmith & Nephew Medical Limited · 2019
  13. This deviceK223041 Renasys Edge (66803126)Smith & Nephew Medical Limited · 2022

Most influential descendants

Showing 2 of 2 · See all
  • K231939Renasys EdgeSmith & Nephew Medical, Ltd. · 2024 · cites it directly1descendant
  • K260646Renasys Edge (66803126)Smith & Nephew Medical, Ltd. · 2026 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−12K945062VacplusKinetic Concepts, Inc.1995—
−11K042134Versatile 1 Wound Vacuum SystemBlue Sky Medical Group Incorporated20042
−10K052456Modification to Versatile 1 Wound Vacuum SystemBlue Sky Medical Group Incorporated2005—
−9K061919Versatile 1 Ezcare Wound Vacuum SystemBlue Sky Medical Group Incorporated20071
−8K082426Renasys EZSmith & Nephew, Inc.20084
−7K083375Renasys GO Negative Pressure Wound TherapySmith & Nephew, Inc.20092
−6K091470Ranasys EZ Negative Pressure Wound Therapy System, Model 66800059Smith & Nephew, Inc.20091
−5K102001Modification to Renasys EZ Plus Negative Pressure Wound TherapySmith & Nephew, Inc.20103
−4K132446Renasys EZ Max Negative Pressure Wound Therapy DeviceSmith & Nephew, Inc.2013—
−3K142979RENASYS; EZ MAX Negative Pressure Wound Therapy; Foam NPWT Wound Dressing Kits…Smith & Nephew, Inc.20151
−2K143133RENASYS¿ EZ MAX Negative Pressure Wound Therapy Device, RENASYS¿ AB Abdominal…Smith & Nephew, Inc.2015—
−1K181822RENASYS Touch; RENASYS Y-ConnectorSmith & Nephew Medical Limited2019—
0 (this device)K223041Renasys Edge (66803126)Smith & Nephew Medical Limited2022—
+1K231939Renasys EdgeSmith & Nephew Medical, Ltd.2024—
+2K260646Renasys Edge (66803126)Smith & Nephew Medical, Ltd.2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410
199710
199920
200010
200210
200320
200510
200620
200710
200820
201010
201120
201510
201610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202410
202610
FIG. 1 — Lineage of K223041, 1995–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OMP? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OMP device

  1. #1 K260646 Renasys Edge (66803126)score 15.3 · 14 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K181822RENASYS Touch; RENASYS Y-Connector
Summary page 4
…Number Regulatory Class Class II Product Code OMP 21 CFR 807.92 (a)(3): Legally 510(k) Number: K181822 marketed device to Device Name: RENASYS TOUCH which equivalence is claimed…
Smith & Nephew Medical Limited2019—

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K231939Renasys EdgeSmith & Nephew Medical, Ltd.20241

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 2, 2024Class IIZ-0516-2025Process controlDue to issues with battery depletion and inability to recharge if the batteries are not appropriately maintained.
Mar 11, 2024Class IIZ-1596-2024Under Investigation by firmBlack particulate matter from the carbon filter component of the canisters entering the NPWT pump and potentially causing damage to the pump
Mar 11, 2024Class IIZ-1597-2024Under Investigation by firmBlack particulate matter from the carbon filter component of the canisters entering the NPWT pump and potentially causing damage to the pump

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 27 years and passes through 4 companies; a new generation was cleared every 1.5 years on median.
  • 14 of the 31 ancestors have a recall of any class.
  • Of its descendants, K231939 (Smith & Nephew Medical, Ltd., 2024) has gone furthest, with 1 descendant of its own.

How does its category compare?

All OMP devices

OMP Negative Pressure Wound Therapy Powered Suction Pump: FDA decisions, review times, recalls and adverse event reports over 20 years. 2022, this device’s year, is highlighted.

FDA decisions per year · OMP

5in 2025

2007: 62008: 82009: 82010: 62011: 62012: 112013: 112014: 102015: 202016: 192017: 112018: 132019: 152020: 62021: 92022: 102023: 82024: 92025: 52026 (partial year): 9
2007high 20 (2015) · low 5 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OMP

127in 2025

2007: 113 days median2008: 98 days median2009: 87 days median2010: 97 days median2011: 123 days median2012: 292 days median2013: 84 days median2014: 179 days median2015: 189 days median2016: 146 days median2017: 137 days median2018: 214 days median2019: 249 days median2020: 179 days median2021: 238 days median2022: 229 days median2023: 291 days median2024: 181 days median2025: 127 days median2026 (partial year): 141 days median
2007high 292 (2012) · low 84 (2013)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OMP

2in 2024

2009: 72010: 32012: 12013: 22014: 42015: 12016: 22019: 12024: 2
2007high 7 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OMP

140in 2025

2007: 12008: 12009: 792010: 802011: 1302012: 972013: 1692014: 2292015: 3782016: 3982017: 3682018: 4222019: 9942020: 3,6642021: 1,4222022: 2582023: 1432024: 1352025: 140
2007high 3,664 (2020) · low 1 (2007)2025

MAUDE reports naming product code OMP, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007611301
200889801
2009887779
2010697380
201161230130
201211292197
201311842169
2014101794229
2015201891378
2016191462398
2017111370368
2018132140422
2019152491994
2020617903,664
2021923801,422
2022102290258
202382910143
202491812135
202551270140
202691410—

What is it indicated for?

The RENASY'S EDGE pump is indicated for patients who would benefit from a suction pump (NPWT), as it may promote wound healing via removal of fluids, including irrigation fluids and body fluids, wound exudate and infectious materials. Appropriate wound types include: Chronic Acute Traumatic Sub-acute and dehisced wounds Ulcers (such as pressure or diabetic) Partial-thickness burns Flaps Grafts When used with the RENASYS AB Abdominal Kit with Soft Port…

Read the full statement (149 words)

The RENASY'S EDGE pump is indicated for patients who would benefit from a suction pump (NPWT), as it may promote wound healing via removal of fluids, including irrigation fluids and body fluids, wound exudate and infectious materials. Appropriate wound types include:

Chronic

Acute

Traumatic

Sub-acute and dehisced wounds

Ulcers (such as pressure or diabetic)

Partial-thickness burns

Flaps

Grafts

When used with the RENASYS AB Abdominal Kit with Soft Port, the RENASYS EDGE pump is indicated for temporary bridging of abdominal wall openings where primary closure is not possible and/or repeat abdominal entries are necessary. It is intended to be used in open abdominal wounds with exposed viscera, including but not limited to abdominal compartment syndrome. The use of RENASYS AB Abdominal Kit with Soft Port is intended for use in acute hospital care settings (trauma, general and plastic surgery wards) and should ideally be applied in the operating theatre.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K223041: Renasys Edge (66803126).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K223041

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026