510(k) Report

510(k) · Cleared

Cleared

Hilzo Esophageal Stents

K223266510(k) TraditionalCleared Jan 16, 2024449-day reviewClass II21 CFR 878.3610

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Thoracent, Inc. · ESW Prosthesis, Esophageal

Cleared on Jan 16, 2024 as substantially equivalent to K080782 (Taewoong Medical Co., Ltd., 2009) after a 449-day review. Its family tree reaches back 4 generations to 1996.

Decided
Jan 16, 2024
Review
449 days
Class
Class II
Product code
ESW
Recalls
None
Predicates
1

Its longest line of descent runs back 4 generations to K955347 (1996). 8 ancestors in all, the oldest K920218 (1994). No later device cites it as a predicate yet.

  1. Root · 4 generations backK955347 UltraflexBoston Scientific Corp · 1996
  2. 3 generations backK032930 Ultraflex Esophageal NG Stent SystemBoston Scientific Corp · 2003 · Recalled
  3. 2 generations backK041648 Niti-S Stent & Introducer, Model EoxxxxTaewoong Medical Co., Ltd. · 2005
  4. 1 generation backK080782 Niti-S Esophageal StentTaewoong Medical Co., Ltd. · 2009 · Recalled
  5. This deviceK223266 Hilzo Esophageal StentsThoracent, Inc. · 2024
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K955347UltraflexBoston Scientific Corp1996—
−3K032930Ultraflex Esophageal NG Stent SystemBoston Scientific Corp200316
−2K041648Niti-S Stent & Introducer, Model EoxxxxTaewoong Medical Co., Ltd.2005—
−1K080782Niti-S Esophageal StentTaewoong Medical Co., Ltd.20091
0 (this device)K223266Hilzo Esophageal StentsThoracent, Inc.2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199420
200010
200210
FIG. 1 — Lineage of K223266, 1996–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in ESW? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new ESW device

  1. #1 K261978 WallFlex Esophageal Stent System;Agile…score 15.2 · 13 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K080782Niti-S Esophageal Stent
Summary page 5
…eal Stent System, manufactured by Taewoong Medical Co., Prosthesis, Esophageal ESW 878.3610 2 General & Plastic Surgery Gastroenterology/Urology Ltd., cleared under K080782; Product Code: ESW Page 594 of 597 K223266 Page 1 of 4 K223266 HILZO Esophageal Stents - Section 5: 510(k) Sum…
Taewoong Medical Co., Ltd.20091
†K123205Esophageal TTS StentTaewoong Medical Co., Ltd.20131
†K092144Bonastent Esophageal, Models BE and Ber, Bonastent Esophageal U.E.S. Flange and Flare…Endochoice, Inc.2010—

Also cleared against K080782: 4 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 28 years and passes through 3 companies; a new generation was cleared every 7.7 years on median.
  • 3 of the 8 ancestors have a recall of any class.
  • 4 other devices also cite its predicate K080782.

How does its category compare?

All ESW devices

ESW Prosthesis, Esophageal: FDA decisions, review times, recalls and adverse event reports over 20 years. 2024, this device’s year, is highlighted.

FDA decisions per year · ESW

3in 2025

2008: 52009: 32010: 32011: 12012: 22013: 22016: 12018: 22019: 12020: 12021: 42022: 12024: 42025: 32026 (partial year): 1
2007high 5 (2008) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · ESW

90in 2025

2008: 99 days median2009: 35 days median2010: 206 days median2021: 139 days median2024: 130 days median2025: 90 days median
2007high 206 (2010) · low 35 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · ESW

25in 2010

2010: 25
2007high 25 (2010) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · ESW

120in 2025

2007: 882008: 1532009: 2392010: 2552011: 1402012: 1212013: 1452014: 1692015: 1752016: 1532017: 872018: 862019: 1672020: 1552021: 772022: 1092023: 1552024: 1662025: 120
2007high 255 (2010) · low 77 (2021)2025

MAUDE reports naming product code ESW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—088
20085990153
20093350239
2010320625255
20111—0140
20122—0121
20132—0145
20140—0169
20150—0175
20161—0153
20170—087
20182—086
20191—0167
20201—0155
20214139077
20221—0109
20230—0155
202441300166
20253900120
20261—0—

What is it indicated for?

The HILZO Esophageal Stents are indicated for use in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K223266: Hilzo Esophageal Stents.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K223266

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026