# K230236: Lyra Influenza A+B Assay

- Pathway: 510(k) (Special)
- Applicant: Quidel Corporation
- Decided: Mar 3, 2023 after 32 days
- Product code: OZE (Influenza A And Influenza B Multiplex Nucleic Acid Assay)
- Device class: 2
- Regulation: 21 CFR 866.3980
- Recalls of this device: none
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 6 generations back to DEN070013 (2008). 9 ancestors in all.
No later device cites it as a predicate yet.
0 of 10 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K131728: Quidel Molecular Influenza a + B Assay, Quidel Corp. (2013) Summary page 4.

## Indications for use
The Lyra Influenza A+B Assay is a multiplex Real Time RT-PCR assay for the in vitro qualitative detection and differentiation of influenza A and influenza B viral RNA in nasal and nasopharyngeal swabs from patients with signs and symptoms of respiratory infection. This test is intended for use as an aid in the differential diagnosis of influenza A and influenza B viral infections in humans in conjunction with clinical and epidemiological risk factors. The assay does not detect the presence of influenza C virus. Negative results do not preclude influenza virus infection and should not be used as the sole basis for diagnosis, treatment or other patient management decisions. Performance characteristics for influenza A were established during the 2011 and 2013 influenza seasons when influenza A/H3 and 2009 H1N1 influenza were the predominant influenza A viruses in circulation. When other influenza A viruses are emerging, performance characteristics may vary. If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens. The assay can be performed using either the Life Technologies QuantStudio Dx, the Applied Biosystems 7500 Fast Dx or the Cepheid SmartCycler II.

Source page: https://510kreport.com/device/K230236