AVVIGO'+ Multi-Modality Guidance System

K230884510(k) TraditionalCleared Sep 26, 2023180-day reviewClass II21 CFR 870.1425Recalls 22

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Boston Scientific Corporation · DQK Computer, Diagnostic, Programmable

Cleared on Sep 26, 2023 as substantially equivalent to K212490 (Boston Scientific Corporation, 2021) after a 180-day review. Its family tree reaches back 11 generations to 2001. 1 later device descends from it. It has been recalled 22 times.

Decided
Sep 26, 2023
Review
180 days
Class
Class II
Product code
DQK
Recalls
22
Predicates
1

Its longest line of descent runs back 11 generations to K003122 (2001). 23 ancestors in all, the oldest K921750 (1992). 1 later device descends from it, 1 citing it directly.

  1. Root · 11 generations backK003122 Smartflow, Model Sfip 2000Florence Medical, Ltd. · 2001
  2. 7 generations between
  3. 3 generations backK201178 iLab Polaris Multi-Modality Guidance SystemBoston Scientific · 2020
  4. 2 generations backK201713 AVVIGO Guidance SystemBoston Scientific Corporation · 2020
  5. 1 generation backK212490 AVVIGO Guidance System IIBoston Scientific Corporation · 2021
  6. This deviceK230884 AVVIGO'+ Multi-Modality Guidance SystemBoston Scientific Corporation · 2023
  7. Descendant · cites it directly · 0 of its ownK253692 AVVIGO™+ Multi-Modality Guidance SystemBoston Scientific · 2026
Show all 11 generations
  1. Root · 11 generations backK003122 Smartflow, Model Sfip 2000Florence Medical, Ltd. · 2001
  2. 10 generations backK012947 SmartflowFlorence Medical, Ltd. · 2001
  3. 9 generations backK020127 Smartflow Pulse Transmission CoefficientFlorence Medical, Ltd. · 2002
  4. 8 generations backK022188 Radianalyzer, Model 12710Radi Medical Systems AB · 2002
  5. 7 generations backK042628 Modification to: Radianalyzer, Model 12711Radi Medical Systems AB · 2005
  6. 6 generations backK092105 Radianalyzer XpressRadi Medical Systems AB · 2009
  7. 5 generations backK151613 iLab Polaris Multi-Modality Guidance SystemBoston Scientific Corporation · 2015
  8. 4 generations backK191008 iLab Polaris Multi-Modality Guidance SystemBoston Scientific Corporation · 2019 · Recalled
  9. 3 generations backK201178 iLab Polaris Multi-Modality Guidance SystemBoston Scientific · 2020
  10. 2 generations backK201713 AVVIGO Guidance SystemBoston Scientific Corporation · 2020
  11. 1 generation backK212490 AVVIGO Guidance System IIBoston Scientific Corporation · 2021
  12. This deviceK230884 AVVIGO'+ Multi-Modality Guidance SystemBoston Scientific Corporation · 2023

Most influential descendants

Showing 1 of 1 · See all
  • K253692AVVIGO™+ Multi-Modality Guidance SystemBoston Scientific · 2026 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−11K003122Smartflow, Model Sfip 2000Florence Medical, Ltd.2001—
−10K012947SmartflowFlorence Medical, Ltd.2001—
−9K020127Smartflow Pulse Transmission CoefficientFlorence Medical, Ltd.2002—
−8K022188Radianalyzer, Model 12710Radi Medical Systems AB2002—
−7K042628Modification to: Radianalyzer, Model 12711Radi Medical Systems AB2005—
−6K092105Radianalyzer XpressRadi Medical Systems AB2009—
−5K151613iLab Polaris Multi-Modality Guidance SystemBoston Scientific Corporation2015—
−4K191008iLab Polaris Multi-Modality Guidance SystemBoston Scientific Corporation20191
−3K201178iLab Polaris Multi-Modality Guidance SystemBoston Scientific2020—
−2K201713AVVIGO Guidance SystemBoston Scientific Corporation2020—
−1K212490AVVIGO Guidance System IIBoston Scientific Corporation2021—
0 (this device)K230884AVVIGO'+ Multi-Modality Guidance SystemBoston Scientific Corporation2023—
+1K253692AVVIGO™+ Multi-Modality Guidance SystemBoston Scientific2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199210
199320
199810
200120
200210
200510
200710
201320
202110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202610
FIG. 1 — Lineage of K230884, 2001–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DQK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DQK device

  1. #1 K213682 HemoSphere Advanced Monitor, HemoSphere…score 26.7 · 28 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K212490AVVIGO Guidance System II
Summary page 5
…ITX (Transducer, Ultrasonic, Diagnostic) Classification Class II, 21 CFR 870.1425 Predicate Device AVVIGO™ Guidance System II K212490 cleared 02 OCT 2021 Device Description The AVVIGO™+ Multi-Modality Guidance System is used by physicians to obtain physiologic, i…
Boston Scientific Corporation2021—

Also cleared against K212490: 1 other device →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K253692AVVIGO™+ Multi-Modality Guidance SystemBoston Scientific20260

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 5, 2024Class IIZ-0097-2025Under Investigation by firmBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System…
Sep 5, 2024Class IIZ-0101-2025Under Investigation by firmBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System…
Sep 5, 2024Class IIZ-0102-2025Under Investigation by firmBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System…
Sep 5, 2024Class IIZ-0105-2025Under Investigation by firmBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System…
Sep 5, 2024Class IIZ-0108-2025Under Investigation by firmBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System…
Sep 5, 2024Class IIZ-0110-2025Under Investigation by firmBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System…
Sep 5, 2024Class IIZ-0114-2025Under Investigation by firmBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System…
Sep 5, 2024Class IIZ-0103-2025Under Investigation by firmBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System…
Sep 5, 2024Class IIZ-0106-2025Under Investigation by firmBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System…
Sep 5, 2024Class IIZ-0111-2025Under Investigation by firmBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System…
Sep 5, 2024Class IIZ-0095-2025Under Investigation by firmBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System…
Sep 5, 2024Class IIZ-0098-2025Under Investigation by firmBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System…
Sep 5, 2024Class IIZ-0104-2025Under Investigation by firmBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System…
Sep 5, 2024Class IIZ-0113-2025Under Investigation by firmBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System…
Sep 5, 2024Class IIZ-0096-2025Under Investigation by firmBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System…
Sep 5, 2024Class IIZ-0099-2025Under Investigation by firmBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System…
Sep 5, 2024Class IIZ-0100-2025Under Investigation by firmBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System…
Sep 5, 2024Class IIZ-0107-2025Under Investigation by firmBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System…
Sep 5, 2024Class IIZ-0109-2025Under Investigation by firmBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System…
Sep 5, 2024Class IIZ-0112-2025Under Investigation by firmBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System…
Sep 5, 2024Class IIZ-0115-2025Under Investigation by firmBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System…
Sep 5, 2024Class IIZ-0116-2025Under Investigation by firmBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 22 years and passes through 4 companies; a new generation was cleared every 1.2 years on median.
  • 3 of the 23 ancestors have a recall of any class.
  • Of its descendants, K253692 (Boston Scientific, 2026) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate K212490.

How does its category compare?

All DQK devices

DQK Computer, Diagnostic, Programmable: FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.

FDA decisions per year · DQK

23so far in 2026, the most since 2007

2007: 132008: 42009: 122010: 92011: 122012: 72013: 132014: 92015: 72016: 152017: 72018: 142019: 152020: 162021: 162022: 192023: 212024: 132025: 222026 (partial year): 23
2007high 23 (2026) · low 4 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DQK

95in 2025

2007: 83 days median2008: 70 days median2009: 93 days median2010: 64 days median2011: 69 days median2012: 74 days median2013: 68 days median2014: 90 days median2015: 125 days median2016: 190 days median2017: 146 days median2018: 150 days median2019: 113 days median2020: 123 days median2021: 132 days median2022: 136 days median2023: 133 days median2024: 100 days median2025: 95 days median2026 (partial year): 151 days median
2007high 190 (2016) · low 64 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DQK

1in 2025

2007: 42008: 12009: 22010: 32011: 22012: 32013: 12014: 42015: 22017: 22018: 62019: 12020: 12021: 12022: 22023: 22024: 242025: 12026 (partial year): 2
2007high 24 (2024) · low 0 (2016)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DQK

433in 2025

2007: 182008: 342009: 402010: 492011: 552012: 1292013: 1802014: 1152015: 1312016: 3402017: 2522018: 2612019: 2182020: 2152021: 3512022: 3192023: 3402024: 5572025: 433
2007high 557 (2024) · low 18 (2007)2025

MAUDE reports naming product code DQK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071383418
2008470134
20091293240
2010964349
20111269255
20127743129
201313681180
20149904115
201571252131
2016151900340
201771462252
2018141506261
2019151131218
2020161231215
2021161321351
2022191362319
2023211332340
20241310024557
202522951433
2026231512—

What is it indicated for?

The IVUS modality of the System is intended for ultrasound examinations of intravascular pathology. Intravascular ultrasound is indicated in patients who are candidates for transluminal interventional procedures such as angioplasty and atherectomy. FFR and DFR™ are intended for use in catheterization and related cardiovascular specialty laboratories to compute, and display various physiological parameters based on the output from one or more electrodes…

Read the full statement (259 words)

The IVUS modality of the System is intended for ultrasound examinations of intravascular pathology. Intravascular ultrasound is indicated in patients who are candidates for transluminal interventional procedures such as angioplasty and atherectomy.

FFR and DFR™ are intended for use in catheterization and related cardiovascular specialty laboratories to compute, and display various physiological parameters based on the output from one or more electrodes, transducers, or measuring devices. FFR and DFR are indicated to provide hemodynamic information for use in the diagnosis and treatment of patients that undergo measurement of physiological parameters.

Refer to the Catheter Instructions for Use provided with all Boston Scientific Ultrasound Imaging Catheters to determine compatibility with the System. All Ultrasound Imaging Catheters will be referred to as Imaging Catheters throughout the remainder of this User Guide.

The Imaging Catheters generate ultrasound images and are intended for ultrasound examination of vascular and cardiac pathology. Boston Scientific manufactures a wide variety of catheters for different applications. The recommended use of each of these catheters may vary depending on the size and type of the catheter. Please refer to the Imaging Catheter Instructions for Use, packaged with each catheter.

Indications for Auto Pullback Use (IVUS Only)

Automatic Pullback is indicated when the following occurs:

• The physician/operator wants to standardize the method in which intravascular ultrasound images are obtained and documented: procedure-to-procedure, operator-to-operator.

• The physician/operator wants to make linear distance determinations post-procedurally, which requires the imaging core of a catheter to be pulled back at a known uniform speed.

• Two-dimensional, longitudinal reconstruction of the anatomy is desired.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K230884: AVVIGO'+ Multi-Modality Guidance System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K230884

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026