# K230930: enspire 3000 Cleaning and Liquid Chemical Sterile Processing System

- Pathway: 510(k) (Traditional)
- Applicant: STERIS Corporation
- Decided: Jun 30, 2023 after 88 days
- Product code: FEB (Accessories, Cleaning, For Endoscope)
- Device class: 2
- Regulation: 21 CFR 876.1500
- Recalls of this device: none
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 8 generations back to K161683 (2016). 17 ancestors in all.
1 later device descends from it; 1 cites it directly.
0 of 19 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K152189: Evotech Ecr, Advanced Sterilization Products (Asp) (2016) Summary page 5.
- K211607: SYSTEM 1E Liquid Chemical Sterilant Processing System, SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6800, SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6900, S40 Sterilant Concentrate, STERIS Corporation (2021) Summary page 14.

## Most influential descendants
- K243433: enspire 3000 Cleaning and Liquid Chemical Sterilant Processing System (2024), 0 descendants of its own

## Indications for use
The enspire™ 3000 Series Cleaning and Liquid Chemical Sterilant Processing System is intended to effectively provide a pressure monitor, clean, provide liquid chemical sterilization, rinse, and air purge validated immersible, reusable, semi-critical, heat sensitive medical devices such as flexible endoscopes and their accessories. The validated cleaning process replaces cleaning for endoscopes other than duodenoscopes. Manual cleaning of duodenoscopes according to the manufacturer’s written instructions for use is required prior to placement in the enspire™ 3000 Series Processor. The enspire™ 3000 Series Processor uses only Revital-Ox 2X Concentrate Enzymatic detergent to clean and S40 Sterilant Concentrate to liquid chemically sterilize medical devices. It automatically dilutes the S40 Sterilant Concentrate to its use dilution (> 1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 46.6 to 55°C and rinses the load with 0.2-micron filtered water.

Source page: https://510kreport.com/device/K230930
Methodology: https://510kreport.com/methodology