510(k) · Cleared
ClearedTalon Transseptal Sheath
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Inari Medical, Inc. · DYB Introducer, Catheter
Cleared on Sep 27, 2023 as substantially equivalent to K183655 (Baylis Medical Company, Inc., 2019) after a 161-day review. Its family tree reaches back 3 generations to 2002. 1 of the 4 devices in its family, counting itself, has a Class I recall.
- Decided
- Sep 27, 2023
- Review
- 161 days
- Class
- Class II
- Product code
- DYB
- Recalls
- None
- Predicates
- 1
Its longest line of descent runs back 3 generations to K013919 (2002). No later device cites it as a predicate yet. 1 of the 4 devices in this family, counting this one, has a Class I recall.
- Root · 3 generations backK013919 Torflex Transseptal Guiding SheathBaylis Medical Co., Inc. · 2002 · Recalled
- 2 generations backK102948 Torflex Transseptal Guiding SheathBaylis Medical Co., Inc. · 2011 · Recalled (Class I)
- 1 generation backK183655 VersaCross Transseptal SheathBaylis Medical Company, Inc. · 2019 · Recalled
- This deviceK231108 Talon Transseptal SheathInari Medical, Inc. · 2023
Class I recalls in this family
- K102948 Torflex Transseptal Guiding Sheath2 generations back · 2011
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −3 | K013919 | Torflex Transseptal Guiding Sheath | Baylis Medical Co., Inc. | 2002 | 1 |
| −2 | K102948 | Torflex Transseptal Guiding Sheath | Baylis Medical Co., Inc. | 2011 | Class I |
| −1 | K183655 | VersaCross Transseptal Sheath | Baylis Medical Company, Inc. | 2019 | 3 |
| 0 (this device) | K231108 | Talon Transseptal Sheath | Inari Medical, Inc. | 2023 | — |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new DYB device
- #1 K251051 AuST Steerable Sheathscore 18.0 · 10 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K183655 | VersaCross Transseptal SheathSummary page 4…sification number 21 CFR 870.1340 Product code DYB Regulatory class II Predicate device(s) Baylis Medical, VersaCross Transseptal Sheath (K183655) Reference device(s) Cook, Extra Large Check-Flo Introducer (K203670) Inari Medical, Triever24 (K… | 2019 |
| † | K203670 | Extra Large Check-Flo Introducer | 2021 |
| † | K213402 | Triever24, Triever20 | 2022 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (1 Class I)
No FDA recall names this device.
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K102948Torflex Transseptal Guiding SheathBaylis Medical Co., Inc. · 2011 · 2 generations back · ◆ on the line of descent in FIG. 11 Class I recall
What does this lineage say?
- The line of descent spans 21 years and passes through 3 companies; a new generation was cleared every 8.1 years on median.
- 3 of the 3 ancestors have a recall of any class; 1 has a Class I recall.
- 1 other device also cites its predicate K183655.
How does its category compare?
All DYB devicesDYB Introducer, Catheter: FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.
FDA decisions per year · DYB
16in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · DYB
63in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · DYB
3in 2025
Class I (48 since 2007) Class II/III not classified
Adverse event reports per year · DYB
4,539in 2025
MAUDE reports naming product code DYB, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 12 | 95 | 15 | 316 |
| 2008 | 17 | 37 | 9 | 423 |
| 2009 | 14 | 52 | 25 | 399 |
| 2010 | 18 | 123 | 18 | 445 |
| 2011 | 17 | 100 | 4 | 628 |
| 2012 | 22 | 128 | 3 | 898 |
| 2013 | 26 | 118 | 5 (1) | 743 |
| 2014 | 30 | 122 | 2 | 780 |
| 2015 | 17 | 129 | 2 (1) | 913 |
| 2016 | 19 | 95 | 45 (3) | 1,234 |
| 2017 | 24 | 70 | 22 (1) | 1,672 |
| 2018 | 19 | 156 | 1 | 1,903 |
| 2019 | 18 | 47 | 2 | 2,450 |
| 2020 | 9 | 49 | 21 (21) | 2,465 |
| 2021 | 12 | 78 | 6 | 2,835 |
| 2022 | 13 | 143 | 2 | 4,084 |
| 2023 | 15 | 55 | 10 | 3,775 |
| 2024 | 18 | 56 | 2 | 3,931 |
| 2025 | 16 | 63 | 3 | 4,539 |
| 2026 | 17 | 59 | 23 (21) | — |
What is it indicated for?
The Talon Transseptal Sheath is used for the percutaneous introduction of various types of cardiovascular catheters and guidewires to all heart chambers, including the left atrium via transseptal perforation/puncture.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K231108 · printed Nov 5, 2026
Cite this page
510(k) Report. “K231108: Talon Transseptal Sheath.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K231108
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026