510(k) Report

510(k) · Cleared

Cleared

CMO8 Adult PMP Oxygenator with Tubing Pack (Model: CME40009)

K231414510(k) TraditionalCleared Nov 1, 2023169-day reviewClass II21 CFR 870.4350

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Chalice Medical, Ltd. · DTZ Oxygenator, Cardiopulmonary Bypass

Cleared on Nov 1, 2023 as substantially equivalent to K201642 (Chalice Medical, Ltd., 2020) and 1 other predicate after a 169-day review. Its family tree reaches back 7 generations to 1992.

Decided
Nov 1, 2023
Review
169 days
Class
Class II
Product code
DTZ
Recalls
None
Predicates
2

Its longest line of descent runs back 7 generations to K922933 (1992). 11 ancestors in all, the oldest K771692 (1977). No later device cites it as a predicate yet.

  1. Root · 7 generations backK922933 Sorin Monolyrh Integrated Membrane LungSorin Biomedica, Fiat, USA, Inc. · 1992
  2. 6 generations backK980600 D903 Avant Adult Hollow Fiber OxygenatorDideco S.P.A. · 1998
  3. 5 generations backK020351 D 903 Avant 2 PH.I.S.I.O. Adult Hollow Fiber OxygenatorDideco S.P.A. · 2002
  4. 4 generations backK033323 D 903 Avant Adult Hollow Fiber Oxygenator and D 903 Avant 2…Dideco S.P.A. · 2004 · Recalled
  5. 3 generations backK121229 Inspire 8 Dual Hollow Fiber Oxygenator with with Integrated…Sorin Group Italia S.R.L. · 2012
  6. 2 generations backK122254 Inspire 8 Hollow Fiber Oxygenator with Integrated Hardshell…Sorin Group Italia S.R.L. · 2012
  7. 1 generation backK201642 Paragon Adult Maxi PMP Oxygenator with Tubing PackChalice Medical, Ltd. · 2020
  8. This deviceK231414 CMO8 Adult PMP Oxygenator with Tubing Pack (Model: CME40009)Chalice Medical, Ltd. · 2023
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K922933Sorin Monolyrh Integrated Membrane LungSorin Biomedica, Fiat, USA, Inc.1992—
−6K980600D903 Avant Adult Hollow Fiber OxygenatorDideco S.P.A.1998—
−5K020351D 903 Avant 2 PH.I.S.I.O. Adult Hollow Fiber OxygenatorDideco S.P.A.2002—
−4K033323D 903 Avant Adult Hollow Fiber Oxygenator and D 903 Avant 2 PH.I.S.I.O. Adult…Dideco S.P.A.20041
−3K121229Inspire 8 Dual Hollow Fiber Oxygenator with with Integrated Hardshell…Sorin Group Italia S.R.L.2012—
−2K122254Inspire 8 Hollow Fiber Oxygenator with Integrated Hardshell Venous/Cardiotomy…Sorin Group Italia S.R.L.2012—
−1K201642Paragon Adult Maxi PMP Oxygenator with Tubing PackChalice Medical, Ltd.2020—
0 (this device)K231414CMO8 Adult PMP Oxygenator with Tubing Pack (Model: CME40009)Chalice Medical, Ltd.2023—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
197710
199210
200120
FIG. 1 — Lineage of K231414, 1992–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DTZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DTZ device

  1. #1 K240934 Quantum Perfusion Blood Oxygenator ECC…score 14.3 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K201642Paragon Adult Maxi PMP Oxygenator with Tubing Pack
Summary page 4
…ass: Class II 510(k) Review Panel Cardiovascular Predicate Devices Paragon Adult Maxi PMP Oxygenator with Tubing Pack (K201642), MVR Venous Reservoir Bag (K920774) FDA Clearance TBC 5.1 Device Description The CMO8…
Chalice Medical, Ltd.2020—
*K920774MVR Collapsible Venous Reservoir Bag
Summary page 4
…Predicate Devices Paragon Adult Maxi PMP Oxygenator with Tubing Pack (K201642), MVR Venous Reservoir Bag (K920774) FDA Clearance TBC 5.1 Device Description The CMO8 Adult PMP oxygenator (CMO8) is a hollow fiber membrane oxygena…
Medtronic Vascular1992—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 31 years and passes through 4 companies; a new generation was cleared every 3.5 years on median.
  • 1 of the 11 ancestors has a recall of any class.

How does its category compare?

All DTZ devices

DTZ Oxygenator, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.

FDA decisions per year · DTZ

2in 2025

2007: 42008: 32009: 92010: 22011: 42012: 92013: 92014: 52015: 92016: 32017: 22018: 32019: 52020: 52021: 52022: 22023: 52024: 32025: 22026 (partial year): 3
2007high 9 (2009) · low 2 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DTZ

124in 2024

2007: 55 days median2008: 88 days median2009: 210 days median2011: 160 days median2012: 90 days median2013: 28 days median2014: 28 days median2015: 134 days median2016: 62 days median2018: 72 days median2019: 90 days median2020: 93 days median2021: 336 days median2023: 145 days median2024: 124 days median2026 (partial year): 137 days median
2007high 336 (2021) · low 28 (2013)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DTZ

1in 2025

2007: 72008: 52010: 12011: 132012: 142014: 22016: 12018: 2, 2 Class I2019: 12020: 32021: 1, 1 Class I2022: 62025: 1
2007high 14 (2012) · low 0 (2009)2026

Class I (3 since 2007) Class II/III not classified

Adverse event reports per year · DTZ

976in 2025

2007: 1342008: 1052009: 522010: 1192011: 1902012: 2152013: 3462014: 7642015: 6582016: 4432017: 3882018: 3202019: 4892020: 4422021: 5952022: 5592023: 7382024: 5352025: 976
2007high 976 (2025) · low 52 (2009)2025

MAUDE reports naming product code DTZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20074557134
20083885105
20099210052
20102—1119
2011416013190
201299014215
20139280346
20145282764
201591340658
20163621443
20172—0388
20183722 (2)320
20195901489
20205933442
202153361 (1)595
20222—6559
202351450738
202431240535
20252—1976
202631370—

What is it indicated for?

The CMO8 Adult PMP Oxygenator is a hollow fiber membrane oxygenator intended for physiologic gas exchange in adults and small adults undergoing cardiopulmonary bypass surgery. The integrated heat exchanger makes it possible to regulate the blood temperature. The CMO8 Adult PMP Oxygenator is intended for clinical use for up to 6 hours. The Tubing Pack is intended for use in extracorporeal circulation during cardiopulmonary bypass procedures, lasting 6 hours or less.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K231414: CMO8 Adult PMP Oxygenator with Tubing Pack (Model: CME40009).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K231414

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026