510(k) Report

510(k) · Cleared

Cleared

TriALTIS Spine System

K231479510(k) TraditionalCleared Sep 22, 2023123-day reviewClass II21 CFR 888.3070

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medos International SARL · NKB Thoracolumbosacral Pedicle Screw System

Cleared on Sep 22, 2023 as substantially equivalent to K220374 (Medos International SARL, 2022) and 2 other predicates after a 123-day review. Its family tree reaches back 19 generations to 2004. 8 later devices descend from it.

Decided
Sep 22, 2023
Review
123 days
Class
Class II
Product code
NKB
Recalls
None
Predicates
5

Its longest line of descent runs back 19 generations to K040583 (2004). 232 ancestors in all, the oldest K781443 (1978). 8 later devices descend from it, 5 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2016: 4 other ancestors2012: 3 other ancestors2015: 1 other ancestor2014: 1 other ancestor2013: 3 other ancestors2011: 6 other ancestors2010: 9 other ancestors2007: 8 other ancestors2004: 24 other ancestors2003: 14 other ancestors2002: 10 other ancestors2000: 16 other ancestors1999: 10 other ancestors2009: 7 other ancestors2008: 8 other ancestors2006: 16 other ancestors2005: 23 other ancestors2001: 21 other ancestors1998: 10 other ancestors1997: 4 other ancestors1996: 5 other ancestors1995: 5 other ancestors1978: 1 other ancestor1993: 2 other ancestors1984: 1 other ancestor1985: 1 other ancestor2025: 2 descendants cleared2023: 1 descendant cleared2026: 4 descendants cleared2024: 1 descendant cleared×7: 7 generations cleared close together, K040583 (2004), K041118 (2004), K042025 (2004), K043151 (2005), K050809 (2005), K052187 (2005), K053101 (2006) ×2: 2 generations cleared close together, K091442 (2009), K091974 (2009) K102555 · CD Horizon Spinal System · Medtronic Sofamor Danek USA, Inc. · 2010×2: 2 generations cleared close together, K112473 (2011), K113174 (2011) K121680 · CD Horizon Spinal System · Medtronic Sofamor Danek USA, Inc. · 2012×3: 3 generations cleared close together, K131321 (2013), K132471 (2013), K140276 (2014) K142381 · Xia 3 Spinal System · Stryker Corp. · 2014K200245 · EXPEDIUM® Spine System; EXPEDIUM VERSE® Spine System · Medos International Sàrl · 2020K220374 · EXPEDIUM VERSE® Fenestrated Screw System (Sterile Fenestrated Screws) · Medos International SARL · 202219801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 2 other ancestors3 generations back: 3 other ancestors4 generations back: 18 other ancestors5 generations back: 55 other ancestors6 generations back: 67 other ancestors7 generations back: 43 other ancestors8 generations back: 20 other ancestors9 generations back: 3 other ancestors 1 generation forward: 5 descendants2 generations forward: 3 descendants K220374 · EXPEDIUM VERSE® Fenestrated Screw System (Sterile Fenestrated Screws) · Medos International SARL · 2022K200245 · EXPEDIUM® Spine System; EXPEDIUM VERSE® Spine System · Medos International Sàrl · 2020K142381 · Xia 3 Spinal System · Stryker Corp. · 2014K140276 · CD Horizon Spinal System · Medtronic Sofamor Danek · 2014K132471 · CD Horizon Spinal System · Medtronic Sofamor Danek USA, Inc. · 2013K131321 · CD Horizon Spinal System · Medtronic Sofamor Danek USA · 2013K121680 · CD Horizon Spinal System · Medtronic Sofamor Danek USA, Inc. · 2012K113174 · CD Horizon Spinal System · Medtronic Sofamor Danek USA, Inc. · 2011 · recalledK112473 · CD Horizon Spinal System · Medtronic Sofamor Danek, Inc. · 2011K102555 · CD Horizon Spinal System · Medtronic Sofamor Danek USA, Inc. · 2010K091974 · Modification to CD Horizon Spinal System · Medtronic Sofamor Danek USA, Inc. · 2009 · recalledK091442 · CD Horizon Spinal System · Medtronic Sofamor Danek · 2009K053101 · CD Horizon Spinal System · Medtronic Sofamor Danek · 2006K052187 · Modification to: CD Horizon Spinal System · Medtronic Sofamor Danek · 2005K050809 · CD Horizon Spinal System · Medtronic Sofamor Danek USA, Inc. · 2005K043151 · Modification to: CD Horizon Spinal System · Medtronic Sofamor Danek · 2005K042025 · Modification to CD Horizon Spinal System · Medtronic Sofamor Danek, Inc. · 2004 · recalledK041118 · Modification to CD Horizon Spinal System · Medtronic Sofamor Danek · 2004K040583 · Modification to: CD Horizon Spinal System · Medtronic Sofamor Danek · 2004−15−10−50
  1. Root · 19 generations backK040583 Modification to: CD Horizon Spinal SystemMedtronic Sofamor Danek · 2004
  2. 15 generations between
  3. 3 generations backK142381 Xia 3 Spinal SystemStryker Corp. · 2014
  4. 2 generations backK200245 EXPEDIUM® Spine System; EXPEDIUM VERSE® Spine SystemMedos International Sàrl · 2020
  5. 1 generation backK220374 EXPEDIUM VERSE® Fenestrated Screw System (Sterile Fenestrated Screws)Medos International SARL · 2022
  6. This deviceK231479 TriALTIS Spine SystemMedos International SARL · 2023
  7. Descendant · cites it directly · 4 of its ownK233684 TriALTIS™ Spine SystemMedos International SARL · 2023
  8. Descendant · cites it directly · 2 of its ownK253990 KHEIRON® Spinal Fixation System, including patient specific K-RODS.M.A.I.O · 2026
  9. Descendant · 2 generations on · 1 of its ownK260786 KHEIRON® Spinal Fixation System, including patient specific K-RODS.M.A.I.O · 2026
  10. Descendant · cites it directly · 1 of its ownK253249 TriALTIS™ Spine SystemMedos International SARL · 2025
  11. Descendant · 2 generations on · 1 of its ownK242042 TriALTIS™ Spine SystemMedos International SARL · 2024
Show all 19 generations
  1. Root · 19 generations backK040583 Modification to: CD Horizon Spinal SystemMedtronic Sofamor Danek · 2004
  2. 18 generations backK041118 Modification to CD Horizon Spinal SystemMedtronic Sofamor Danek · 2004
  3. 17 generations backK042025 Modification to CD Horizon Spinal SystemMedtronic Sofamor Danek, Inc. · 2004 · Recalled
  4. 16 generations backK043151 Modification to: CD Horizon Spinal SystemMedtronic Sofamor Danek · 2005
  5. 15 generations backK050809 CD Horizon Spinal SystemMedtronic Sofamor Danek USA, Inc. · 2005
  6. 14 generations backK052187 Modification to: CD Horizon Spinal SystemMedtronic Sofamor Danek · 2005
  7. 13 generations backK053101 CD Horizon Spinal SystemMedtronic Sofamor Danek · 2006
  8. 12 generations backK091442 CD Horizon Spinal SystemMedtronic Sofamor Danek · 2009
  9. 11 generations backK091974 Modification to CD Horizon Spinal SystemMedtronic Sofamor Danek USA, Inc. · 2009 · Recalled
  10. 10 generations backK102555 CD Horizon Spinal SystemMedtronic Sofamor Danek USA, Inc. · 2010
  11. 9 generations backK112473 CD Horizon Spinal SystemMedtronic Sofamor Danek, Inc. · 2011
  12. 8 generations backK113174 CD Horizon Spinal SystemMedtronic Sofamor Danek USA, Inc. · 2011 · Recalled
  13. 7 generations backK121680 CD Horizon Spinal SystemMedtronic Sofamor Danek USA, Inc. · 2012
  14. 6 generations backK131321 CD Horizon Spinal SystemMedtronic Sofamor Danek USA · 2013
  15. 5 generations backK132471 CD Horizon Spinal SystemMedtronic Sofamor Danek USA, Inc. · 2013
  16. 4 generations backK140276 CD Horizon Spinal SystemMedtronic Sofamor Danek · 2014
  17. 3 generations backK142381 Xia 3 Spinal SystemStryker Corp. · 2014
  18. 2 generations backK200245 EXPEDIUM® Spine System; EXPEDIUM VERSE® Spine SystemMedos International Sàrl · 2020
  19. 1 generation backK220374 EXPEDIUM VERSE® Fenestrated Screw System (Sterile Fenestrated Screws)Medos International SARL · 2022
  20. This deviceK231479 TriALTIS Spine SystemMedos International SARL · 2023

Showing 5 of 8 descendants. See all descendants

Descendants cleared per year

2023: 12024: 12025: 22026: 4
2023high 4 (2026) · low 1 (2023)2026

Most influential descendants

Showing 5 of 8 · See all
  • K233684TriALTIS™ Spine SystemMedos International SARL · 2023 · cites it directly4descendants
  • K253990KHEIRON® Spinal Fixation System, including patient specific K-RODS.M.A.I.O · 2026 · cites it directly2descendants
  • K260786KHEIRON® Spinal Fixation System, including patient specific K-RODS.M.A.I.O · 2026 · 2 generations on1descendant
  • K253249TriALTIS™ Spine SystemMedos International SARL · 2025 · cites it directly1descendant
  • K242042TriALTIS™ Spine SystemMedos International SARL · 2024 · 2 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−19K040583Modification to: CD Horizon Spinal SystemMedtronic Sofamor Danek2004—
−18K041118Modification to CD Horizon Spinal SystemMedtronic Sofamor Danek2004—
−17K042025Modification to CD Horizon Spinal SystemMedtronic Sofamor Danek, Inc.20044
−16K043151Modification to: CD Horizon Spinal SystemMedtronic Sofamor Danek2005—
−15K050809CD Horizon Spinal SystemMedtronic Sofamor Danek USA, Inc.2005—
−14K052187Modification to: CD Horizon Spinal SystemMedtronic Sofamor Danek2005—
−13K053101CD Horizon Spinal SystemMedtronic Sofamor Danek2006—
−12K091442CD Horizon Spinal SystemMedtronic Sofamor Danek2009—
−11K091974Modification to CD Horizon Spinal SystemMedtronic Sofamor Danek USA, Inc.20091
−10K102555CD Horizon Spinal SystemMedtronic Sofamor Danek USA, Inc.2010—
−9K112473CD Horizon Spinal SystemMedtronic Sofamor Danek, Inc.2011—
−8K113174CD Horizon Spinal SystemMedtronic Sofamor Danek USA, Inc.20111
−7K121680CD Horizon Spinal SystemMedtronic Sofamor Danek USA, Inc.2012—
−6K131321CD Horizon Spinal SystemMedtronic Sofamor Danek USA2013—
−5K132471CD Horizon Spinal SystemMedtronic Sofamor Danek USA, Inc.2013—
−4K140276CD Horizon Spinal SystemMedtronic Sofamor Danek2014—
−3K142381Xia 3 Spinal SystemStryker Corp.2014—
−2K200245EXPEDIUM® Spine System; EXPEDIUM VERSE® Spine SystemMedos International Sàrl2020—
−1K220374EXPEDIUM VERSE® Fenestrated Screw System (Sterile Fenestrated Screws)Medos International SARL2022—
0 (this device)K231479TriALTIS Spine SystemMedos International SARL2023—
+1K233684TriALTIS™ Spine SystemMedos International SARL2023—
+1K253990KHEIRON® Spinal Fixation System, including patient specific K-RODS.M.A.I.O2026—
+2K260786KHEIRON® Spinal Fixation System, including patient specific K-RODS.M.A.I.O2026—
+1K253249TriALTIS™ Spine SystemMedos International SARL2025—
+2K242042TriALTIS™ Spine SystemMedos International SARL2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
197810
198410
198510
199320
199550
199650
199740
1998100
1999100
2000160
2001210
2002100
2003140
2004240
2005230
2006160
200780
200880
200970
201090
201160
201230
201330
201410
201510
201640

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202310
202410
202520
202640
FIG. 1 — Lineage of K231479, 2004–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in NKB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NKB device

  1. #1 K223494 CD Horizon Spinal Systemscore 40.4 · 30 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (5)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary; numbers the summary cites that cannot be its predicate are marked – and kept off the family tree.

TypeNumberDeviceApplicantDecidedRecalls
Kept off the family treeK200524FDA’s K200524 is SOMATOM X.cite, Scan&GO, SOMATOM go.Platform - go.Up, go.Now, go.Top, go.All…
Our OCR of the scanned summary read K200524, FDA’s number for SOMATOM X.cite, Scan&GO, SOMATOM go.Platform - go.Up… (Siemens Medical Solutions USA, Inc.). Its FDA name has none of the distinctive words the summary uses for the device it names, “EXPEDIUM Spine System and EXPEDIUM VERSE Spine System”. Kept off the family tree.
Siemens Medical Solutions USA, Inc.20207
*K220374EXPEDIUM VERSE® Fenestrated Screw System (Sterile Fenestrated Screws)
Summary page 4
…_ ____ EXPEDIUM Spine System and EXPEDIUM VERSE Spine SystEjl �N_KB _ ___� K 2 20 37 4 EXPEDIUM VERSE Spine System fenestrated screws NKB K160879…
Medos International SARL2022—
*K160879CONFIDENCE™ High Viscosity Spinal Cement, VIPER® and EXPEDIUM® Fenestrated Screw Systems
Summary page 4
…_ ___� K 2 20 37 4 EXPEDIUM VERSE Spine System fenestrated screws NKB K160879 VIPER and EXPEDIUM Spine System fenestrated screws PML �K1 _ _60…
Medos International Sàrl2016—
*K160904Expedium Spine System, VIPER and VIPER 2 SystemsMedos International SARL2016—
Kept off the family treeK170353FDA’s K170353 is ABX MICROS ES 60 OT and ABX MICROS ES 60 CT
Not matched in our reading of the summary (scans can blur digits), so it is kept off the family tree
Horiba Abx Sas2017—

What descends from it? (8)

5 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K233684TriALTIS™ Spine SystemMedos International SARL20234
K253990KHEIRON® Spinal Fixation System, including patient specific K-RODS.M.A.I.O20262
K253249TriALTIS™ Spine SystemMedos International SARL20251
K262828KHEIRON® Spinal Fixation System, including patient specific K-RODS.M.A.I.O20260
K243249TriALTIS™ Spine SystemMedos International SARL20250

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 19 years and passes through 7 companies; a new generation was cleared every 5 months on median.
  • 29 of the 232 ancestors have a recall of any class.
  • Of its descendants, K233684 (Medos International SARL, 2023) has gone furthest, with 4 descendants of its own.

How does its category compare?

All NKB devices

NKB Thoracolumbosacral Pedicle Screw System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.

FDA decisions per year · NKB

24in 2025

2007: 382008: 362009: 442010: 372011: 482012: 602013: 352014: 462015: 552016: 502017: 632018: 562019: 392020: 482021: 322022: 342023: 442024: 392025: 242026 (partial year): 30
2007high 63 (2017) · low 24 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NKB

61in 2025

2007: 87 days median2008: 59 days median2009: 71 days median2010: 111 days median2011: 84 days median2012: 79 days median2013: 73 days median2014: 90 days median2015: 90 days median2016: 84 days median2017: 69 days median2018: 61 days median2019: 84 days median2020: 68 days median2021: 61 days median2022: 88 days median2023: 72 days median2024: 57 days median2025: 61 days median2026 (partial year): 91 days median
2007high 111 (2010) · low 57 (2024)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · NKB

1in 2025

2007: 42008: 52009: 332010: 42011: 32012: 132014: 3, 1 Class I2015: 32016: 12017: 12018: 32019: 12020: 12022: 52023: 22024: 32025: 12026 (partial year): 3
2007high 33 (2009) · low 0 (2013)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · NKB

1,171in 2025

2007: 342008: 822009: 1712010: 4622011: 5772012: 1,7472013: 1,9622014: 1,6712015: 1,4462016: 1,5852017: 1,6182018: 1,6842019: 1,9152020: 2,0762021: 1,7452022: 1,3392023: 9192024: 1,6142025: 1,171
2007high 2,076 (2020) · low 34 (2007)2025

MAUDE reports naming product code NKB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20073887434
20083659582
2009447133171
2010371114462
201148843577
20126079131,747
2013357301,962
201446903 (1)1,671
2015559031,446
2016508411,585
2017636911,618
2018566131,684
2019398411,915
2020486812,076
2021326101,745
2022348851,339
202344722919
2024395731,614
2025246111,171
202630913—

What is it indicated for?

The TriALTIS TM Spine System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine. The TriALTIS TM Spine System is intended for pedicle fixation of the thoracic, lumbar, and sacral spine (T1-S2) and nonpedicle fixation for the following indications: degenerative disc…

Read the full statement (369 words)

The TriALTIS TM Spine System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine.

The TriALTIS TM Spine System is intended for pedicle fixation of the thoracic, lumbar, and sacral spine (T1-S2) and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudarthrosis; and failed previous fusion in skeletally mature patients.

When used for posterior pedicle screw fixation of the thoracic, lumbar, and sacral spine (T1-S2) in pediatric patients, the TriALTIS ™ Spine System metallic implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The TriALTIS ™ Spine System is intended to be used with autograft and/or allograft. Pediatric screw fixation is limited to a posterior approach.

When used for posterior pedicle screw fixation of the thoracic, lumbar, and sacral spine (T1-S2) in pediatric patients, the TriALTIS ™ Spine System is indicated as an adjunct to fusion to treat progressive spinal deformities (i.e, scoliosis, kyphosis, or lordosis) including adolescent idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the TriALTIS ™ Spine System is intended to treat pediatric patients diagnosed with: spondylolisthesis/spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. The TriAL TIS ™ Spine System is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

When the TriALTIS ™ Spine System fenestrated screws are used in conjunction with CONFIDENCE ™ High Viscosity Spinal Cement, the TriALTIS ™ Spine System is intended to stabilize the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. The TriALTIS ™ fenestrated screws augmented with the CONFIDENCE ™ High Viscosity Spinal Cement is for use at spinal levels where the structural integrity of the spine is not severely compromised.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K231479: TriALTIS Spine System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K231479

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026