510(k) Report

510(k) · Cleared

Cleared

NeuroAlign software

K231897510(k) TraditionalCleared Oct 21, 2025846-day reviewClass II21 CFR 882.4560

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medivis, Inc. · HAW Neurological Stereotaxic Instrument

Cleared on Oct 21, 2025 as substantially equivalent to K092467 (Brainlab AG, 2010) after an 846-day review. Its family tree reaches back 5 generations to 1999. 1 of the 10 devices in its family, counting itself, has a Class I recall.

Decided
Oct 21, 2025
Review
846 days
Class
Class II
Product code
HAW
Recalls
None
Predicates
1

Its longest line of descent runs back 5 generations to K983831 (1999). 9 ancestors in all. No later device cites it as a predicate yet. 1 of the 10 devices in this family, counting this one, has a Class I recall.

  1. Root · 5 generations backK983831 VECTORVISION2Brainlab AG · 1999
  2. 4 generations backK014256 Kolibri Igs SystemsBrainlab AG · 2002
  3. 3 generations backK042391 Kolibri Image Guided Surgery SystemBrainlab AG · 2004
  4. 2 generations backK082060 Cranial Image Guided Surgery SystemBrainlab AG · 2009
  5. 1 generation backK092467 Vectorvision Cranial, Vectorvision Ent, Kolibri Cranial, Kolibri…Brainlab AG · 2010 · Recalled (Class I)
  6. This deviceK231897 NeuroAlign softwareMedivis, Inc. · 2025

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K983831VECTORVISION2Brainlab AG1999—
−4K014256Kolibri Igs SystemsBrainlab AG2002—
−3K042391Kolibri Image Guided Surgery SystemBrainlab AG2004—
−2K082060Cranial Image Guided Surgery SystemBrainlab AG2009—
−1K092467Vectorvision Cranial, Vectorvision Ent, Kolibri Cranial, Kolibri Ent, Cranial…Brainlab AG2010Class I
0 (this device)K231897NeuroAlign softwareMedivis, Inc.2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200010
200110
200310
200410
FIG. 1 — Lineage of K231897, 1999–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HAW device

  1. #1 K253856 Geniant Cranial (Navigated Neurosurgical…score 15.0 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K092467Vectorvision Cranial, Vectorvision Ent, Kolibri Cranial, Kolibri Ent, Cranial Essential…
Summary page 5
…e Manufacturer: BRAINLAB AG Predicate Device Common Name: Neurological Stereotaxic Instrument Predicate Device Premarket Notification # K092467 Predicate Device Classification & Name Class 2 Stereotaxic instrument Predicate Device Classification & Class…
Brainlab AG2010Class I

Also cleared against K092467: 3 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (1 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K092467Vectorvision Cranial, Vectorvision Ent, Kolibri Cranial, Kolibri Ent, Cranial Essential, Cranial UnlimitedBrainlab AG · 2010 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 26 years and passes through 2 companies; a new generation was cleared every 3.2 years on median.
  • 3 of the 9 ancestors have a recall of any class; 1 has a Class I recall.
  • 3 other devices also cite its predicate K092467.

How does its category compare?

All HAW devices

HAW Neurological Stereotaxic Instrument: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.

FDA decisions per year · HAW

9in 2025

2007: 152008: 122009: 82010: 62011: 42012: 82013: 32014: 52015: 102016: 92017: 72018: 62019: 102020: 52021: 92022: 72023: 142024: 72025: 92026 (partial year): 5
2007high 15 (2007) · low 3 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HAW

163in 2025

2007: 107 days median2008: 112 days median2009: 221 days median2010: 138 days median2011: 66 days median2012: 178 days median2013: 172 days median2014: 238 days median2015: 232 days median2016: 175 days median2017: 226 days median2018: 112 days median2019: 147 days median2020: 88 days median2021: 168 days median2022: 128 days median2023: 142 days median2024: 107 days median2025: 163 days median2026 (partial year): 162 days median
2007high 238 (2014) · low 66 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HAW

2in 2025

2007: 22008: 102009: 62010: 282011: 52012: 42013: 6, 1 Class I2014: 52015: 32016: 52017: 22018: 2, 2 Class I2019: 4, 1 Class I2020: 12021: 4, 2 Class I2022: 12023: 4, 1 Class I2024: 2, 2 Class I2025: 22026 (partial year): 6
2007high 28 (2010) · low 1 (2020)2026

Class I (9 since 2007) Class II/III not classified

Adverse event reports per year · HAW

3,425in 2025

2007: 492008: 532009: 962010: 2992011: 1,7252012: 9862013: 1,2902014: 1,6062015: 1,7672016: 1,8282017: 3,5052018: 4,4392019: 6,0622020: 3,4802021: 2,8272022: 1,9292023: 2,5102024: 3,7552025: 3,425
2007high 6,062 (2019) · low 49 (2007)2025

MAUDE reports naming product code HAW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200715107249
2008121121053
20098221696
2010613828299
201146651,725
201281784986
201331726 (1)1,290
2014523851,606
20151023231,767
2016917551,828
2017722623,505
201861122 (2)4,439
2019101474 (1)6,062
202058813,480
202191684 (2)2,827
2022712811,929
2023141424 (1)2,510
202471072 (2)3,755
2025916323,425
202651626—

What is it indicated for?

The Medivis NeuroAlign device is stereotaxic image guidance system intended to be used with the Microsoft HoloLens headset for the spatial positioning and orientation of neurosurgical instruments employed by surgeons. This device is specifically intended for cranial surgery, where reference to a rigid anatomical structure can be identified, without the need for a frame or fixation of a navigated instrument guide to the patient. The system is intended for use in various surgical settings, including operating rooms, intensive care units, and interventional procedure suites.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K231897: NeuroAlign software.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K231897

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026