510(k) Report

510(k) · Cleared

Cleared

Powder Free Nitrile Patient Examination Gloves, Blue Colored, Non Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate

K232266510(k) TraditionalCleared Nov 3, 202395-day reviewClass I21 CFR 880.6250

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Better Care Plastic Technology Co., Ltd. · LZA Polymer Patient Examination Glove

Cleared on Nov 3, 2023 as substantially equivalent to K221269 (Better Care Plastic Technology Co., Ltd., 2022) after a 95-day review. Its family tree reaches back 3 generations to 2019. 3 later devices descend from it.

Decided
Nov 3, 2023
Review
95 days
Class
Class I
Product code
LZA
Recalls
None
Predicates
1

Its longest line of descent runs back 3 generations to K183287 (2019). 3 later devices descend from it, 3 citing it directly.

  1. Root · 3 generations backK183287 Powder Free Nitrile Patient Examination Glove, Blue Colored…Kossan International Sdn Bhd · 2019
  2. 2 generations backK192954 Power Free Nitrile Examination Gloves Tested for Use with…Comfort Rubber Gloves Industries Sdn. Bhd. · 2020
  3. 1 generation backK221269 Powder Free Nitrile Examination Gloves (Blue),Tested for Use with…Better Care Plastic Technology Co., Ltd. · 2022
  4. This deviceK232266 Powder Free Nitrile Patient Examination Gloves, Blue Colored, Non…Better Care Plastic Technology Co., Ltd. · 2023
  5. Descendant · cites it directly · 0 of its ownK260703 Powder Free Blue Nitrile Examination Glove - Tested for Use with…S&S Glove Corporation · 2026
  6. Descendant · cites it directly · 0 of its ownK242533 Powder Free Nitrile Patient Examination Gloves, Blue Colored, Non…Better Care Plastic Technology Co., Ltd. · 2025
  7. Descendant · cites it directly · 0 of its ownK240629 Nitrile Disposable Examination Gloves, Tested for Use with…Xingyu Medical Tech Co., Ltd. · 2024

Most influential descendants

Showing 3 of 3 · See all
  • K260703Powder Free Blue Nitrile Examination Glove - Tested for Use with Chemotherapy Drugs and…S&S Glove Corporation · 2026 · cites it directly0descendants
  • K242533Powder Free Nitrile Patient Examination Gloves, Blue Colored, Non Sterile, Low…Better Care Plastic Technology Co., Ltd. · 2025 · cites it directly0descendants
  • K240629Nitrile Disposable Examination Gloves, Tested for Use with Chemotherapy Drugs and FentanylXingyu Medical Tech Co., Ltd. · 2024 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K183287Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-Sterile…Kossan International Sdn Bhd2019—
−2K192954Power Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs…Comfort Rubber Gloves Industries Sdn. Bhd.2020—
−1K221269Powder Free Nitrile Examination Gloves (Blue),Tested for Use with Chemotherapy…Better Care Plastic Technology Co., Ltd.2022—
0 (this device)K232266Powder Free Nitrile Patient Examination Gloves, Blue Colored, Non Sterile…Better Care Plastic Technology Co., Ltd.2023—
+1K260703Powder Free Blue Nitrile Examination Glove - Tested for Use with Chemotherapy…S&S Glove Corporation2026—
+1K242533Powder Free Nitrile Patient Examination Gloves, Blue Colored, Non Sterile, Low…Better Care Plastic Technology Co., Ltd.2025—
+1K240629Nitrile Disposable Examination Gloves, Tested for Use with Chemotherapy Drugs…Xingyu Medical Tech Co., Ltd.2024—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202410
202510
202610
FIG. 1 — Lineage of K232266, 2019–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LZA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LZA device

  1. #1 K192333 Blue Nitrile Examination Gloves Powder Freescore 71.0 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K221269Powder Free Nitrile Examination Gloves (Blue),Tested for Use with Chemotherapy Drugs and…
Summary page 6
…evice Information: Better Care Plastic Technology Co., Ltd. Powder Free Nitrile Examination Gloves (Blue) Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (K221269) 4. Device Description: Powder Free Nitrile Patient Examination Gloves, Blue Colored, Non Sterile, Tested for Use with Chemoth…
Better Care Plastic Technology Co., Ltd.2022—

Also cleared against K221269: 3 other devices →

What descends from it? (3)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K260703Powder Free Blue Nitrile Examination Glove - Tested for Use with Chemotherapy Drugs and…S&S Glove Corporation20260
K242533Powder Free Nitrile Patient Examination Gloves, Blue Colored, Non Sterile, Low…Better Care Plastic Technology Co., Ltd.20250
K240629Nitrile Disposable Examination Gloves, Tested for Use with Chemotherapy Drugs and FentanylXingyu Medical Tech Co., Ltd.20240

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 4 years and passes through 3 companies; a new generation was cleared every 1.3 years on median.
  • None of the 3 ancestors has a recall.
  • Of its descendants, K260703 (S&S Glove Corporation, 2026) has gone furthest, with 0 descendants of its own.
  • 3 other devices also cite its predicate K221269.

How does its category compare?

All LZA devices

LZA Polymer Patient Examination Glove: FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.

FDA decisions per year · LZA

24in 2025

2007: 212008: 182009: 282010: 312011: 262012: 332013: 202014: 272015: 202016: 302017: 442018: 222019: 252020: 302021: 1392022: 1462023: 622024: 302025: 242026 (partial year): 15
2007high 146 (2022) · low 18 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LZA

110in 2025

2007: 95 days median2008: 97 days median2009: 112 days median2010: 97 days median2011: 122 days median2012: 110 days median2013: 130 days median2014: 197 days median2015: 112 days median2016: 171 days median2017: 124 days median2018: 178 days median2019: 212 days median2020: 101 days median2021: 111 days median2022: 111 days median2023: 92 days median2024: 89 days median2025: 110 days median2026 (partial year): 103 days median
2007high 212 (2019) · low 89 (2024)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LZA

3so far in 2026, the most since 2007

2015: 12016: 12026 (partial year): 3
2007high 3 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LZA

73in 2025

2007: 72008: 22009: 122010: 122011: 92012: 112013: 102014: 82015: 122016: 252017: 42018: 32019: 212020: 42021: 202022: 192023: 242024: 762025: 73
2007high 76 (2024) · low 2 (2008)2025

MAUDE reports naming product code LZA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007219507
2008189702
200928112012
20103197012
20112612209
201233110011
201320130010
20142719708
201520112112
201630171125
20174412404
20182217803
201925212021
20203010104
2021139111020
2022146111019
20236292024
20243089076
202524110073
2026151033—

What is it indicated for?

The glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Gloves have been tested for use with chemotherapy drugs and Fentanyl Citrate using ASTM D6978-05(2019) Chemotherapy Drug Minimum Breakthrough Detection Time (BDT) in Minutes Bleomycin Sulfate 15mg/ml (15000 ppm) >240 Busulfan 6mg/ml (6,000 ppm) >240 Carboplatin 10mg/ml (10,000 ppm) >240 Carmustine…

Read the full statement (259 words)

The glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent

contamination between patient and examiner. Gloves have been tested for use with chemotherapy drugs and Fentanyl

Citrate using ASTM D6978-05(2019)

Chemotherapy Drug Minimum Breakthrough Detection Time (BDT) in Minutes

Bleomycin Sulfate 15mg/ml (15000 ppm) >240

Busulfan 6mg/ml (6,000 ppm) >240

Carboplatin 10mg/ml (10,000 ppm) >240

Carmustine (BCNU) 3.3 mg/ml (3,300 ppm) 22.8

Chloroquine 50mg/ml (50,000ppm) >240

Cisplatin 1mg/ml (1,000 ppm) >240

Cyclophosphamide 20mg/ml (20,000 ppm) >240

Cyclosporin 100 mg/ml (100,000 ppm) >240

Cytarabine HCL, 100 mg/ml (100,000 ppm) >240

Dacarbazine 10 mg/ml (10,000 ppm) >240

Daunorubicin HCL, 5 mg/ml (5,000 ppm) >240

Docetaxel, 10 mg/ml (10,000 ppm) >240

Doxorubicin HCL, 2 mg/ml (2,000 ppm) >240

Epirubicin HCL, 2 mg/ml (2,000 ppm) >240

Etoposide, 20 mg/ml (20,000 ppm) >240

Fludarabine, 25 mg/ml (25,000 ppm) >240

Fluorouracil, 50mg/ml (50,000ppm) >240

Gemcitabine, 38mg/ml (38,000ppm) >240

Idarubicin HCL, 1mg/ml (1,000ppm) >240

Ifosfamide, 50mg/ml (50,000ppm) >240

Irinotecan, 20mg/ml (20,000ppm) >240

Mechlorethamine HCI, 1mg/ml (1,000ppm) >240

Melphalan, 5mg/ml (5,000ppm) >240

Methotrexate, 25mg/ml (25,000ppm) >240

Mitomycin C, 0.5mg/ml (500ppm) >240

Mitoxantrone HCL, 2mg/ml (2,000ppm) >240

Oxaliplatin, 5mg/ml (5,000ppm) >240

Paclitaxel, 6mg/ml (6,000ppm) >240

Paraplatin, 10mg/ml (10,000ppm) >240

Retrovir, 10mg/ml (10,000ppm) >240

Rituximab, 10mg/ml (10,000ppm) >240

Thiotepa, 10mg/ml (10,000ppm) 46.9

Topotecan, 1mg/ml (1,000ppm) >240

Trisenox, 1mg/ml (1,000ppm) >240

Velcade, 1mg/ml (1,000ppm) >240

Vincristine Sulfate, 1mg/ml (1,000ppm) >240

Fentanyl Citrate Injection (100 mcg/2ml) >240

Please note that the following drugs have extremely low permeation times:

Carmustine: 22.8 minutes, Thiotepa: 46.9 minutes,

*Warning: Do not use with Carmustine and Thiotepa.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K232266: Powder Free Nitrile Patient Examination Gloves, Blue Colored, Non Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K232266

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026