510(k) · Cleared
ClearedPowder Free Nitrile Examination Glove, Pink Colored, Non-Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate; Powder Free Nitrile Examination Glove, Orange Colored, Non-Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate; Powder Free Nitrile Examination Glove, Blue Colored, Non-Sterile, Low Dermatitis Potential, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate; Powder Free Nitrile Examination Glove, Black Colored, Non-Sterile, Low Dermatitis
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Kossan International Sdn Bhd · LZA Polymer Patient Examination Glove
Cleared on Nov 14, 2023 after a 91-day review. FDA’s file for it is a 510(k) statement, which names no predicate.
- Decided
- Nov 14, 2023
- Review
- 91 days
- Class
- Class I
- Product code
- LZA
- Recalls
- None
- Predicates
- 0
FDA’s file for it is a 510(k) statement, which names no predicate.
Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in LZA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new LZA device
- #1 K192333 Blue Nitrile Examination Gloves Powder Freescore 71.0 · 2 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against?
FDA’s file for it is a 510(k) statement, which names no predicate.
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
How does its category compare?
All LZA devicesLZA Polymer Patient Examination Glove: FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.
FDA decisions per year · LZA
24in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · LZA
110in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · LZA
3so far in 2026, the most since 2007
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · LZA
73in 2025
MAUDE reports naming product code LZA, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 21 | 95 | 0 | 7 |
| 2008 | 18 | 97 | 0 | 2 |
| 2009 | 28 | 112 | 0 | 12 |
| 2010 | 31 | 97 | 0 | 12 |
| 2011 | 26 | 122 | 0 | 9 |
| 2012 | 33 | 110 | 0 | 11 |
| 2013 | 20 | 130 | 0 | 10 |
| 2014 | 27 | 197 | 0 | 8 |
| 2015 | 20 | 112 | 1 | 12 |
| 2016 | 30 | 171 | 1 | 25 |
| 2017 | 44 | 124 | 0 | 4 |
| 2018 | 22 | 178 | 0 | 3 |
| 2019 | 25 | 212 | 0 | 21 |
| 2020 | 30 | 101 | 0 | 4 |
| 2021 | 139 | 111 | 0 | 20 |
| 2022 | 146 | 111 | 0 | 19 |
| 2023 | 62 | 92 | 0 | 24 |
| 2024 | 30 | 89 | 0 | 76 |
| 2025 | 24 | 110 | 0 | 73 |
| 2026 | 15 | 103 | 3 | — |
What is it indicated for?
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs and Fentanyl Citrate as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. Chemotherapy Drug and Concentration Minimum Breakthrough Detection Time in…
Read the full statement (259 words)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
These gloves were tested for use with chemotherapy drugs and Fentanyl Citrate as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
Chemotherapy Drug and Concentration Minimum Breakthrough Detection Time in Minutes
Pink Orange Blue Black
Carmustine (BCNU) (3.3 mg/ml) 12.4 11.7 10.1 14.4
Cisplatin (1.0 mg/ml) >240 >240 >240 >240
Cyclophosphamide (Cytoxan) (20.0 mg/ml) >240 >240 >240 >240
Cytarabine(100 mg/ml) >240 Not tested >240 >240
Dacarbazine (DTIC) (10.0 mg/ml) >240 >240 >240 >240
Doxorubicin Hydrochloride (2.0 mg/ml) >240 >240 >240 >240
Etoposide (20.0 mg/ml) >240 >240 >240 >240
Fluorouracil (50.0 mg/ml) >240 >240 >240 >240
lfosfamide (50.0 mg/ml) >240 Not Tested >240 >240
Methotrexate (25.0 mg/ml) >240 Not Tested >240 >240
Mitomycin C (0.5 mg/ml) >240 Not Tested >240 >240
Mitoxantrone (2.0 mg/ml) >240 Not Tested >240 >240
Paclitaxel (Taxol) (6.0 mg/ml) >240 >240 >240 >240
Thiotepa (10.0 mg/ml) 43.1 43.2 30.2 29.2
Vincristine Sulfate (1.0 mg/ml) >240 Not Tested >240 >240
Please note that Carmustine (BCNU) and Thiotepa has low permeation times.
Warning: Do Not Use with Carmustine (BCNU) and Thiotepa
Opioid Drug and Concentration Minimum Breakthrough Detection Time in Minutes
Pink Orange Blue Black
Fentanyl Citrate Injection (100 mcg/2ml) >240 >240 >240 >240
Xylazine HCI (100 mg/ml) Not Tested >240 >240 >240
Simulated Gastric Acid Minimum Breakthrough Detection Time in Minutes
Pink Orange Blue Black
Simulated Gastric Acid >240 >240 >240 >240
Indications for use as stated in the FDA 510(k) statement.
https://510kreport.com/device/K232461 · printed Nov 6, 2026
Cite this page
510(k) Report. “K232461: Powder Free Nitrile Examination Glove, Pink Colored, Non-Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate; Powder Free Nitrile Examination Glove, Orange Colored, Non-Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate; Powder Free Nitrile Examination Glove, Blue Colored, Non-Sterile, Low Dermatitis Potential, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate; Powder Free Nitrile Examination Glove, Black Colored, Non-Sterile, Low Dermatitis.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K232461
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Page last updated Nov 3, 2026