# K232491: CT 5300

- Pathway: 510(k) (Traditional)
- Applicant: Philips Healthcare (Suzhou) Co., Ltd.
- Decided: May 3, 2024 after 260 days
- Product code: JAK (System, X-Ray, Tomography, Computed)
- Device class: 2
- Regulation: 21 CFR 892.1750
- Recalls of this device: 5 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 8 generations back to K930090 (1993). 13 ancestors in all.
1 later device descends from it; 1 cites it directly.
0 of 15 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K212441: Philips Incisive CT, Philips Healthcare (Suzhou) Co., Ltd. (2022) Summary page 4.

## Most influential descendants
- K252992: CT Rembra RT; CT Areta RT; CT Rembra (2026), 0 descendants of its own

## Recalls of this device
- Oct 29, 2025 · Class II · Z-0882-2026: If the screws of tube heat exchanger on the rotating scanner are not tightened well after replacing. This component may become detached and make contact with other components located within the Incisi
- Oct 29, 2025 · Class II · Z-0881-2026: If the screws of tube heat exchanger on the rotating scanner are not tightened well after replacing. This component may become detached and make contact with other components located within the Incisi
- Sep 25, 2025 · Class II · Z-0375-2026: Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axial scan with ORI (Dos
- Sep 25, 2025 · Class II · Z-0376-2026: Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axial scan with ORI (Dos
- Sep 25, 2025 · Class II · Z-0377-2026: Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axial scan with ORI (Dos

## Indications for use
The CT 5300 is a Computed Tomography X-Ray System intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient and equipment supports, components and accessories. The CT 5300 is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages. These scanners are intended to be used for diagnostic imaging and for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer*. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society. *Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.