510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

ProNephro AKI™ (NGAL)

K232761510(k) TraditionalCleared Dec 7, 202390-day reviewClass II21 CFR 862.1220Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Bioporto Diagnostic, Inc. · PIG Acute Kidney Injury Test System

Cleared on Dec 7, 2023 as substantially equivalent to DEN130031 (Astute Medical, Inc., 2014) after a 90-day review. Its family tree reaches back 1 generation to 2014. It has been recalled once.

Decided
Dec 7, 2023
Review
90 days
Class
Class II
Product code
PIG
Recalls
1
Predicates
1

Its longest line of descent runs back 1 generation to DEN130031 (2014). No later device cites it as a predicate yet.

DEN130031 · Nephrocheck Test System · Astute Medical, Inc. · 20142010201520202025
DEN130031 · Nephrocheck Test System · Astute Medical, Inc. · 20140
  1. Root · 1 generation backDEN130031 Nephrocheck Test SystemAstute Medical, Inc. · 2014
  2. This deviceK232761 ProNephro AKI™ (NGAL)Bioporto Diagnostic, Inc. · 2023
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1DEN130031Nephrocheck Test SystemAstute Medical, Inc.2014—
0 (this device)K232761ProNephro AKI™ (NGAL)Bioporto Diagnostic, Inc.2023—
FIG. 1 — Lineage of K232761, 2014–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in PIG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new PIG device

  1. #1 K210793 Vidas Nephrocheckscore 3.9 · 2 generations · no Class I recall
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*DEN130031Nephrocheck Test System
Summary page 9
…510(k) Summary 10. Substantial Equivalence Information 10.1. Predicate Device Name(s) NEPHROCHECK® Test System 10.2. Predicate 510(k) Number(s) DEN130031 10.3. Comparison With Predicate Table 2: Predicate Device Comparison…
Astute Medical, Inc.2014—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 18, 2026Class IIZ-2937-2026Nonconforming Material/ComponentThe use of a non-validated rework process for assay reagents which could lead to missed diagnosis of renal disease, inappropriate treatment decisions, or inappropriate monitoring of renal disease in intended users.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.

How does its category compare?

All PIG devices

PIG Acute Kidney Injury Test System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.

FDA decisions per year · PIG

1in 2023

2014: 12016: 12017: 12022: 12023: 1
2007high 1 (2014) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · PIG

Too few PIG 510(k)s in any year since 2007 to give a median.

Recalls per year · PIG

1so far in 2026, the most since 2007

2026 (partial year): 1
2007high 1 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · PIG

1in 2024

2024: 1
2007high 1 (2024) · low 0 (2007)2025

MAUDE reports naming product code PIG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20130—00
20141—00
20150—00
20161—00
20171—00
20180—00
20190—00
20200—00
20210—00
20221—00
20231—00
20240—01
20250—00
20260—1—

What is it indicated for?

Immunoassay for the in vitro quantitative determination of neutrophil gelatinase-associated lipocalin (NGAL) in human urine. Determination of NGAL is intended to be used in conjunction with clinical evaluation in pediatric patients (≥ 3 months to < 22 years) without underlying kidney disease admitted to the intensive care unit (ICU) for the management of cardiovascular or respiratory compromise or who have had a solid organ or bone marrow transplant.…

Read the full statement (187 words)

Immunoassay for the in vitro quantitative determination of neutrophil gelatinase-associated lipocalin (NGAL) in human urine.

Determination of NGAL is intended to be used in conjunction with clinical evaluation in pediatric patients (≥ 3 months to < 22 years) without underlying kidney disease admitted to the intensive care unit (ICU) for the management of cardiovascular or respiratory compromise or who have had a solid organ or bone marrow transplant.

ProNephro AKI™ (NGAL) is intended to be used in the first 24 hours of ICU admission. In patients with low SCr levels, indicative of no acute kidney injury (AKI) or AKI Stage 1, the test can be used as an aid to identify patients at risk to develop moderate to severe AKI (Stage 2/3) 48 to 72 hours after the assessment. In patients with an elevated SCr level, indicative of Stage 2/3 AKI, the test can be used as an aid to identify patients at risk to have persistent moderate to severe AKI (Stage 2/3) 48 to 72 hours after the assessment.

The particle-enhanced turbidimetric immunoassay is intended for use on the Roche cobas c 501 clinical chemistry analyzer.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K232761: ProNephro AKI™ (NGAL).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K232761

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026