510(k) · Cleared
Recalled · Class I · 2025AXIOS Stent and Electrocautery-Enhanced Delivery System (6mmx8mm); AXIOS Stent and Electrocautery-Enhanced Delivery System (8mmx8mm); AXIOS Stent and Electrocautery-Enhanced Delivery System (10mmx10mm); AXIOS Stent and Electrocautery-Enhanced Delivery System (15mmx10mm); AXIOS Stent and Electrocautery-Enhanced Delivery System (20mmx10mm); AXIOS Stent and Electrocautery-Enhanced Delivery System (15mmx15mm)
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Boston Scientific Corporation · KNS Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
Cleared on Oct 27, 2023 as substantially equivalent to DEN230019 (Boston Scientific Corporation, 2023) after a 28-day review. Its family tree reaches back 1 generation to 2023. 1 of the 3 devices in its family, counting itself, has a Class I recall. 1 later device descends from it. It has been recalled 18 times, including 6 Class I recalls.
- Decided
- Oct 27, 2023
- Review
- 28 days
- Class
- Class II
- Product code
- KNS
- Recalls
- 18 (6 Class I)
- Predicates
- 2
Its longest line of descent runs back 1 generation to DEN230019 (2023). 1 later device descends from it, 1 citing it directly. 1 of the 3 devices in this family, counting this one, has a Class I recall.
- Root · 1 generation backDEN230019 AXIOS Stent and Electrocautery-Enhanced Delivery System (10mm x 10mm…Boston Scientific Corporation · 2023
- This deviceK233318 AXIOS Stent and Electrocautery-Enhanced Delivery System (6mmx8mm)…Boston Scientific Corporation · 2023
- Descendant · cites it directly · 0 of its ownK252648 Niti-S SPAXUS StentTaewoong Medical Co., Ltd. · 2026
Most influential descendants
Showing 1 of 1 · See all- K252648Niti-S SPAXUS StentTaewoong Medical Co., Ltd. · 2026 · cites it directly0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | DEN230019 | AXIOS Stent and Electrocautery-Enhanced Delivery System (10mm x 10mm Stent)… | Boston Scientific Corporation | 2023 | — |
| 0 (this device) | K233318 | AXIOS Stent and Electrocautery-Enhanced Delivery System (6mmx8mm); AXIOS Stent… | Boston Scientific Corporation | 2023 | — |
| +1 | K252648 | Niti-S SPAXUS Stent | Taewoong Medical Co., Ltd. | 2026 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2026 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new KNS device
- #1 K241271 Aqua Medical RF Vapor Ablation Systemscore 17.2 · 20 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary; numbers the summary cites that cannot be its predicate are marked – and kept off the family tree.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | DEN230019 | AXIOS Stent and Electrocautery-Enhanced Delivery System (10mm x 10mm Stent); AXIOS Stent…Summary page 5…21 CFR 807.92(a)(3) Predicate# Predicate Trade Name (Primary Predicate is listed first) Product Code I DEN230019, K2201gl I AXIOS Stent and Electrocautery-Enhanced Delivery System 11 KNS Device Descrip… | 2023 |
| Kept off the family tree | K220191 | FDA’s K220191 is Disposable Face Mask Not matched in our reading of the summary (scans can blur digits), so it is kept off the family tree | 2022 |
What descends from it? (1)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K252648 | Niti-S SPAXUS Stent | 2026 | 0 |
Has anything in its family been recalled? (1 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Dec 19, 2025 | Class I | Z-1162-2026 | Process control | Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most… |
| Dec 19, 2025 | Class I | Z-1160-2026 | Process control | Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most… |
| Dec 19, 2025 | Class I | Z-1163-2026 | Process control | Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most… |
| Dec 19, 2025 | Class I | Z-1158-2026 | Process control | Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most… |
| Dec 19, 2025 | Class I | Z-1159-2026 | Process control | Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most… |
| Dec 19, 2025 | Class I | Z-1161-2026 | Process control | Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most… |
| Dec 20, 2024 | Class II | Z-1025-2025 | Process control | Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one. |
| Dec 20, 2024 | Class II | Z-1026-2025 | Process control | Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one. |
| Dec 20, 2024 | Class II | Z-1027-2025 | Process control | Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one. |
| Dec 20, 2024 | Class II | Z-1029-2025 | Process control | Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one. |
| Dec 20, 2024 | Class II | Z-1032-2025 | Process control | Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one. |
| Dec 20, 2024 | Class II | Z-1033-2025 | Process control | Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one. |
| Dec 20, 2024 | Class II | Z-1037-2025 | Process control | Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one. |
| Dec 20, 2024 | Class II | Z-1024-2025 | Process control | Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one. |
| Dec 20, 2024 | Class II | Z-1028-2025 | Process control | Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one. |
| Dec 20, 2024 | Class II | Z-1034-2025 | Process control | Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one. |
| Dec 20, 2024 | Class II | Z-1035-2025 | Process control | Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one. |
| Dec 20, 2024 | Class II | Z-1036-2025 | Process control | Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one. |
The 5 latest of 18 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
How does its category compare?
All KNS devicesKNS Unit, Electrosurgical, Endoscopic (With Or Without Accessories): FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.
FDA decisions per year · KNS
11in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · KNS
171in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · KNS
14in 2024
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · KNS
1,504in 2025
MAUDE reports naming product code KNS, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 1 | — | 3 | 164 |
| 2008 | 3 | 61 | 5 | 587 |
| 2009 | 3 | 212 | 2 | 1,027 |
| 2010 | 4 | 168 | 0 | 908 |
| 2011 | 1 | — | 2 | 860 |
| 2012 | 2 | — | 0 | 848 |
| 2013 | 2 | — | 0 | 449 |
| 2014 | 1 | — | 0 | 405 |
| 2015 | 3 | 252 | 0 | 406 |
| 2016 | 3 | 30 | 6 | 457 |
| 2017 | 3 | 48 | 1 | 446 |
| 2018 | 5 | 62 | 0 | 563 |
| 2019 | 7 | 66 | 0 | 668 |
| 2020 | 4 | 212 | 0 | 906 |
| 2021 | 4 | 139 | 41 | 814 |
| 2022 | 3 | 182 | 0 | 913 |
| 2023 | 9 | 83 | 0 | 976 |
| 2024 | 3 | 198 | 14 | 1,533 |
| 2025 | 11 | 171 | 0 | 1,504 |
| 2026 | 2 | — | 4 | — |
What is it indicated for?
6mmx8mm and 8mmx8mm stents: The AXIOS Stent and Electrocautery-Enhanced Delivery System is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of symptomatic pancreatic pseudocysts 6 cm in size, that are adherent to the gastric or bowel wall and are free of solid debris. The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst resolution. 1Ommx10mm and 15mmx10mm stents: The…
Read the full statement (280 words)
6mmx8mm and 8mmx8mm stents: The AXIOS Stent and Electrocautery-Enhanced Delivery System is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of symptomatic pancreatic pseudocysts 6 cm in size, that are adherent to the gastric or bowel wall and are free of solid debris. The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst resolution.
1Ommx10mm and 15mmx10mm stents: The AXIOS Electrocautery-Enhanced-Stent and Delivery System is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of symptomatic pancreatic pseudocysts 6 cm in size and walled-off necrosis ! 6 cm in size that are adherent to the gastric or bowel wall. Once placed, the AXIOS Stent functions as an access port allowing passage of standard and therapeutic endoscopes to facilitate debridement, irrigation and cystoscopy. The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst or walled-off necrosis resolution. The AXIOS Stent and Electrocautery-Enhanced Delivery System is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of the gallbladder in patients with acute cholecystitis who are at high risk for surgery.
20mmx10mm and 15mmx15mm stents: The AXIOS Electrocautery-Enhanced Stent and Delivery System is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of symptomatic pancreatic pseudocysts 2 6 cm in size and walled-off necrosis 6 cm in size that are adherent to the gastric or bowel wall. Once placed, the AXIOS Stent functions as an access port allowing passage of standard and therapeutic endoscopes to facilitate debridement, irrigation and cystoscopy. The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst or walled-off necrosis resolution.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K233318 · printed Nov 5, 2026
Cite this page
510(k) Report. “K233318: AXIOS Stent and Electrocautery-Enhanced Delivery System (6mmx8mm); AXIOS Stent and Electrocautery-Enhanced Delivery System (8mmx8mm); AXIOS Stent and Electrocautery-Enhanced Delivery System (10mmx10mm); AXIOS Stent and Electrocautery-Enhanced Delivery System (15mmx10mm); AXIOS Stent and Electrocautery-Enhanced Delivery System (20mmx10mm); AXIOS Stent and Electrocautery-Enhanced Delivery System (15mmx15mm).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K233318
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026