# K233669: OEC 3D

- Pathway: 510(k) (Traditional)
- Applicant: Ge Oec Medical Systems, Inc.
- Decided: Mar 28, 2024 after 134 days
- Product code: OXO (Image-Intensified Fluoroscopic X-Ray System, Mobile)
- Device class: 2
- Regulation: 21 CFR 892.1650
- Recalls of this device: 2 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 11 generations back to K021049 (2002). 12 ancestors in all.
No later device cites it as a predicate yet.
2 of 13 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K203346: OEC 3D, Ge Oec Medical Systems, Inc. (2021) Summary page 4.

## Recalls of this device
- Mar 21, 2025 · Class II · Z-1820-2025: OEC Elite and OEC 3D Mobile C-Arms Exceeding 4% Source to Image Distance in X-ray Field.
- Nov 21, 2024 · Class II · Z-0703-2025: Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.

## Indications for use
The OEC 3D mobile fluoroscopy system is designed to provide fluoroscopic and digital spot images of adult and pediatric populations during diagnostic, interventional, and surgical procedures. Examples of a clinical application may include: orthopedic, gastrointestinal, endoscopic, urologic, neurologic, vascular, cardiac, critical care, and emergency procedures.

Source page: https://510kreport.com/device/K233669