# K233983: Swan-Ganz Pacing Probe and Catheters; Swan-Ganz Bipolar Pacing Catheter (D97120F5); Swan-Ganz Bipolar Pacing Catheter (D97130F5); Swan-Ganz Pacing Thermodilution TD Catheter (D200F7); Chandler Transluminal V-Pacing Probe (D98100)

- Pathway: 510(k) (Traditional)
- Applicant: Edwards Lifesciences
- Decided: Jun 27, 2024 after 192 days
- Product code: DYG (Catheter, Flow Directed)
- Device class: 2
- Regulation: 21 CFR 870.1240
- Recalls of this device: 1 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 2 generations back to K813521 (1982). 4 ancestors in all.
No later device cites it as a predicate yet.
0 of 5 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K791183: Swan-Ganz Flow-Directed Catheter, Edwards Lifesciences, LLC (1979) Summary page 5.
- K822723: Swan-Ganz Bipolar Pacing Catheter, American Edwards Laboratories (1982) Summary page 5.
- K152633: Chandler Transluminal Bipolar Pacing Probe, Flextip Transluminal Bipolar Pacing Probe, Edwards Lifesciences, LLC (2015) Summary page 5.

## Recalls of this device
- Apr 8, 2026 · Class II · Z-2204-2026: Catheters may, due to a manufacturing process and material changes, leak/break from the (blue) proximal injectate lumen hub, which may lead to infection, medication loss, and/or blood loss.

## Indications for use
The Swan-Ganz pacing-TD catheters are indicated for use for temporary, atrial, ventricular, or A-V sequential cardiac pacing of adult surgical and/or critically ill patients. This includes, but is not limited to, patients with arrythmias, acute myocardial infarction, cardiac surgery, and coronary angiography. In addition, these catheters are indicated for use in hemodynamic monitoring of patients experiencing medical conditions such as, but not limited to, post-major surgical recovery, trauma, sepsis, burns, pulmonary disease, pulmonary failure, stroke, and cardiac disease including heart failure. The Swan-Ganz bipolar pacing catheters are indicated for use in surgical and/or critically ill cardiac patients who may benefit from temporary, transvenous right ventricular pacing. The Chandler transluminal V-pacing probe is a transluminal bipolar pacing probe, which must be used with a Swan-Ganz Paceport catheter is indicated for use in adult surgical and/or critically ill cardiac patients. This includes, but is not limited to, patients with arrhythmias, acute myocardial infarction, cardiac surgery and coronary angiography.