510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

CDI OneView Monitoring System

K234065510(k) TraditionalCleared Apr 25, 2024125-day reviewClass II21 CFR 870.4330Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Terumo Cardiovascular Systems Corporation · DRY Monitor, Blood-Gas, On-Line, Cardiopulmonary Bypass

Cleared on Apr 25, 2024 as substantially equivalent to K182110 (Terumo Cardiovascular Systems Corporation, 2018) after a 125-day review. Its family tree reaches back 2 generations to 2014. It has been recalled 2 times.

Decided
Apr 25, 2024
Review
125 days
Class
Class II
Product code
DRY
Recalls
2
Predicates
1

Its longest line of descent runs back 2 generations to K133658 (2014). No later device cites it as a predicate yet.

  1. Root · 2 generations backK133658 Cdi Blood Parameter Monitoring System 500Terumo Cardiovascular Systems Corp. · 2014 · Recalled
  2. 1 generation backK182110 CDI Blood Parameter Monitoring System 550Terumo Cardiovascular Systems Corporation · 2018
  3. This deviceK234065 CDI OneView Monitoring SystemTerumo Cardiovascular Systems Corporation · 2024
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K133658Cdi Blood Parameter Monitoring System 500Terumo Cardiovascular Systems Corp.20144
−1K182110CDI Blood Parameter Monitoring System 550Terumo Cardiovascular Systems Corporation2018—
0 (this device)K234065CDI OneView Monitoring SystemTerumo Cardiovascular Systems Corporation2024—
FIG. 1 — Lineage of K234065, 2014–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DRY? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DRY device

  1. #1 K202154 B-Captascore 4.7 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K182110CDI Blood Parameter Monitoring System 550
Summary page 4
…21 CFR §870.4330 Product Code(s): DRY III. Predicate Device CDI Blood Parameter Monitoring System 550, cleared under K182110. IV. Device Description The CDI OneView Monitoring System is an AC-powered (battery back-up) microprocessor-bas…
Terumo Cardiovascular Systems Corporation2018—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 28, 2025Class IIZ-2018-2025Employee errorDue to an error in assembly of the thermistor within some units, the affected probes will experience inaccuracy in measured shunt sensor temperature. Additionally, following changes in patient temperature without an in vivo recalibration…
Mar 3, 2025Class IIZ-1492-2025Software designDuring some clinical cases, it has been observed that the SO2 value may periodically appear as dashes on the Touchscreen Display indicating unavailable data. In these cases, a yellow low-limit alarm will also be triggered to the user, due…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 10 years and passes through 2 companies; a new generation was cleared every 5.5 years on median.
  • 1 of the 2 ancestors has a recall of any class.

How does its category compare?

All DRY devices

DRY Monitor, Blood-Gas, On-Line, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years. 2024, this device’s year, is highlighted.

FDA decisions per year · DRY

1in 2024

2007: 22009: 22011: 22012: 12014: 22017: 12018: 32020: 12021: 12024: 1
2007high 3 (2018) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DRY

88in 2018

2018: 88 days median
2007high 88 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DRY

1in 2020

2008: 12010: 62011: 142012: 52015: 12020: 1
2007high 14 (2011) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DRY

134in 2025

2008: 12009: 22010: 1,0702011: 2,2682012: 1672013: 1572014: 1562015: 2312016: 2402017: 612018: 1652019: 632020: 352021: 422022: 372023: 412024: 722025: 134
2007high 2,268 (2011) · low 0 (2007)2025

MAUDE reports naming product code DRY, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—00
20080—11
20092—02
20100—61,070
20112—142,268
20121—5167
20130—0157
20142—0156
20150—1231
20160—0240
20171—061
20183880165
20190—063
20201—135
20211—042
20220—037
20230—041
20241—072
20250—0134
20260—0—

What is it indicated for?

The CDI OneView Monitoring System is a patient parameter monitoring system to be used on a single patient during cardiopulmonary bypass procedures. By measurement or acquisition from other devices, it displays and outputs information to provide continuous, in-line monitoring of various patient parameters contained within the extracorporeal perfusion circuit and patient. The following parameters are available, based on configuration: *Potential of…

Read the full statement (126 words)

The CDI OneView Monitoring System is a patient parameter monitoring system to be used on a single patient during cardiopulmonary bypass procedures. By measurement or acquisition from other devices, it displays and outputs information to provide continuous, in-line monitoring of various patient parameters contained within the extracorporeal perfusion circuit and patient. The following parameters are available, based on configuration:

*Potential of Hydrogen (pH)

*Partial Pressure of Carbon Dioxide (pCO2)

*Partial Pressure of Oxygen (p02)

*Potassium Ion (K+)

*Oxygen Saturation (S02)

*Hematocrit (HCT)

*Hemoglobin (Hgb)

*Blood Flow Rate (0)

*Cardiac Index (CI)

*Base Excess (BE)

*Bicarbonate (HCO3-)

*Oxygen Consumption (V02)

*indexed Oxygen Consumption (VO2i)

*Oxygen Delivery (DO2)

*indexed Oxygen Delivery (DO2i)

*Cerebral Regional Oxygen Saturation (rSO2)

*Oxygen Extraction Ratio (O2ER)

*Body Surface Area (BSA)

*Shunt Sensor Temperature

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K234065: CDI OneView Monitoring System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K234065

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026