510(k) Report

510(k) · Cleared

Cleared

Iconix Knotless Anchor

K240213510(k) SpecialCleared Feb 23, 202429-day reviewClass II21 CFR 888.3040

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Riverpoint Medical · MBI Fastener, Fixation, Nondegradable, Soft Tissue

Cleared on Feb 23, 2024 as substantially equivalent to K231278 (Riverpoint Medical, LLC, 2023) after a 29-day review. Its family tree reaches back 11 generations to 1996.

Decided
Feb 23, 2024
Review
29 days
Class
Class II
Product code
MBI
Recalls
None
Predicates
1

Its longest line of descent runs back 11 generations to K961555 (1996). 28 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 11 generations backK961555 Acufex Tag Bioabsorbable AnchorsAcufex Microsurgical, Inc. · 1996
  2. 7 generations between
  3. 3 generations backK133671 Iconix All Suture AnchorsStryker Corp. · 2014
  4. 2 generations backK151201 Stryker ICONIX with NeedlesStryker · 2015
  5. 1 generation backK231278 Knotless Suture AnchorRiverpoint Medical, LLC · 2023 · Recalled
  6. This deviceK240213 Iconix Knotless AnchorRiverpoint Medical · 2024
Show all 11 generations
  1. Root · 11 generations backK961555 Acufex Tag Bioabsorbable AnchorsAcufex Microsurgical, Inc. · 1996
  2. 10 generations backK031685 Tag Rod Bioraptor Suture Anchor 7209317, 7209318Smith & Nephew, Inc. · 2003 · Recalled
  3. 9 generations backK053344 Bioraptor Suture Anchor, Twinfix 2.8 MM Suture Anchor, Twinfix 3.5…Smith & Nephew, Inc. · 2006
  4. 8 generations backK061863 Arthrex Corkscrew Suture Anchor, Tak, and Pushlock FamilyArthrex, Inc. · 2006
  5. 7 generations backK071177 Arthrex Bio- Composite Suture Anchors: Bio-Composite Pushlock…Arthrex, Inc. · 2007
  6. 6 generations backK082810 Arthrex Bio-Composite Suture Anchors: Arthrex Bio-Composite…Arthrex, Inc. · 2009
  7. 5 generations backK101823 Arthrex Swivelock AnchorsArthrex, Inc. · 2011 · Recalled
  8. 4 generations backK110545 Smith & Nephew Next Generation Fully Threaded Peek Suture Anchor…Smith & Nephew, Inc. · 2011
  9. 3 generations backK133671 Iconix All Suture AnchorsStryker Corp. · 2014
  10. 2 generations backK151201 Stryker ICONIX with NeedlesStryker · 2015
  11. 1 generation backK231278 Knotless Suture AnchorRiverpoint Medical, LLC · 2023 · Recalled
  12. This deviceK240213 Iconix Knotless AnchorRiverpoint Medical · 2024
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−11K961555Acufex Tag Bioabsorbable AnchorsAcufex Microsurgical, Inc.1996—
−10K031685Tag Rod Bioraptor Suture Anchor 7209317, 7209318Smith & Nephew, Inc.20035
−9K053344Bioraptor Suture Anchor, Twinfix 2.8 MM Suture Anchor, Twinfix 3.5 MM Suture…Smith & Nephew, Inc.2006—
−8K061863Arthrex Corkscrew Suture Anchor, Tak, and Pushlock FamilyArthrex, Inc.2006—
−7K071177Arthrex Bio- Composite Suture Anchors: Bio-Composite Pushlock, Bio-Composite…Arthrex, Inc.2007—
−6K082810Arthrex Bio-Composite Suture Anchors: Arthrex Bio-Composite Pushlock…Arthrex, Inc.2009—
−5K101823Arthrex Swivelock AnchorsArthrex, Inc.20116
−4K110545Smith & Nephew Next Generation Fully Threaded Peek Suture Anchor (Ngftsa PK)Smith & Nephew, Inc.2011—
−3K133671Iconix All Suture AnchorsStryker Corp.2014—
−2K151201Stryker ICONIX with NeedlesStryker2015—
−1K231278Knotless Suture AnchorRiverpoint Medical, LLC20231
0 (this device)K240213Iconix Knotless AnchorRiverpoint Medical2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199710
199810
200020
200130
200410
200520
200620
200740
201210
FIG. 1 — Lineage of K240213, 1996–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MBI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MBI device

  1. #1 K203268 Arthrex FiberTak Suture Anchorscore 21.4 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K231278Knotless Suture Anchor
Summary page 5
…ices) Division of Health Technology 6C Restorative, Repair and Trauma Devices Predicate Device K231278 – Knotless Suture Anchor No reference devices were used in this submission.…
Riverpoint Medical, LLC20231

Also cleared against K231278: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 28 years and passes through 7 companies; a new generation was cleared every 1.6 years on median.
  • 5 of the 28 ancestors have a recall of any class.
  • 2 other devices also cite its predicate K231278.

How does its category compare?

All MBI devices

MBI Fastener, Fixation, Nondegradable, Soft Tissue: FDA decisions, review times, recalls and adverse event reports over 20 years. 2024, this device’s year, is highlighted.

FDA decisions per year · MBI

27in 2025

2007: 162008: 102009: 192010: 172011: 302012: 242013: 312014: 232015: 232016: 172017: 172018: 162019: 312020: 352021: 182022: 192023: 522024: 222025: 272026 (partial year): 30
2007high 52 (2023) · low 10 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MBI

58in 2025

2007: 80 days median2008: 69 days median2009: 82 days median2010: 76 days median2011: 75 days median2012: 94 days median2013: 100 days median2014: 101 days median2015: 85 days median2016: 94 days median2017: 70 days median2018: 91 days median2019: 90 days median2020: 83 days median2021: 107 days median2022: 64 days median2023: 57 days median2024: 30 days median2025: 58 days median2026 (partial year): 63 days median
2007high 107 (2021) · low 30 (2024)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MBI

1in 2025

2007: 242008: 62009: 22010: 32011: 232012: 62013: 22017: 42018: 22020: 12023: 12024: 12025: 1
2007high 24 (2007) · low 0 (2014)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MBI

3,826in 2025

2007: 2402008: 3382009: 3552010: 2912011: 3282012: 3712013: 4422014: 5042015: 5072016: 7952017: 6752018: 1,0722019: 1,3032020: 1,4542021: 1,8902022: 1,5912023: 2,5032024: 3,7792025: 3,826
2007high 3,826 (2025) · low 240 (2007)2025

MAUDE reports naming product code MBI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007168024240
200810696338
200919822355
201017763291
2011307523328
201224946371
2013311002442
2014231010504
201523850507
201617940795
201717704675
2018169121,072
2019319001,303
2020358311,454
20211810701,890
2022196401,591
2023525712,503
2024223013,779
2025275813,826
202630630—

What is it indicated for?

The Iconix Knotless Anchors are intended to be used for soft-tissue to bone fixation in the foot, ankle, knee, hip, hand, wrist, elbow and shoulder. Specific indications are listed below. Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Repair…

Read the full statement (150 words)

The Iconix Knotless Anchors are intended to be used for soft-tissue to bone fixation in the foot, ankle, knee, hip, hand, wrist, elbow and shoulder. Specific indications are listed below.

Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction

Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Repair

Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP joints for all digits, Digital Tendon Repair

Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus Reconstruction, Digital Tendon Transfers, Mid-foot Reconstruction

Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis, Medial Patellofemoral Ligament (MPFL) Repair/ Reconstruction, Quadriceps Tendon Repair

Hip: Capsular Repair, Acetabular Labral Repair, Gluteal Tendon Repair, Proximal Hamstring Repair

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K240213: Iconix Knotless Anchor.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K240213

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026