510(k) · Cleared
Recalled · Class II · 2025LIFEPAK® ECG Cables
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Stryker Physio-Control · DSA Cable, Transducer And Electrode, Patient, (Including Connector)
Cleared on May 7, 2024 as substantially equivalent to K211294 (Vyaire Medical, Inc., 2021) after a 90-day review. Its family tree reaches back 6 generations to 1994. It has been recalled once.
- Decided
- May 7, 2024
- Review
- 90 days
- Class
- Class II
- Product code
- DSA
- Recalls
- 1
- Predicates
- 1
Its longest line of descent runs back 6 generations to K933649 (1994). No later device cites it as a predicate yet.
- Root · 6 generations backK933649 Conmed Ecg Patient Cables and LeadwiresConmedcorp · 1994
- 5 generations backK980582 Multi-Link Cable and Lead Wire SystemsMarquette Medical Systems, Inc. · 1998
- 4 generations backK082851 Multlink Cable and Leadwire SystemGe Medical Systems Information Technologies · 2008
- 3 generations backK101660 Ecg Single Patient Use Lead Wire Set, Models 2048801-5 to 2048801-8Kepler Medtec · 2010
- 2 generations backK162432 Multi-Link X2 ECG Cable and Lead Wire SystemCarefusion 2200, Inc. · 2017
- 1 generation backK211294 Multi-Link X2 ECG Cable and Leadwire SystemVyaire Medical, Inc. · 2021
- This deviceK240377 LIFEPAK® ECG CablesStryker Physio-Control · 2024
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −6 | K933649 | Conmed Ecg Patient Cables and Leadwires | Conmedcorp | 1994 | — |
| −5 | K980582 | Multi-Link Cable and Lead Wire Systems | Marquette Medical Systems, Inc. | 1998 | — |
| −4 | K082851 | Multlink Cable and Leadwire System | Ge Medical Systems Information Technologies | 2008 | — |
| −3 | K101660 | Ecg Single Patient Use Lead Wire Set, Models 2048801-5 to 2048801-8 | Kepler Medtec | 2010 | — |
| −2 | K162432 | Multi-Link X2 ECG Cable and Lead Wire System | Carefusion 2200, Inc. | 2017 | — |
| −1 | K211294 | Multi-Link X2 ECG Cable and Leadwire System | Vyaire Medical, Inc. | 2021 | — |
| 0 (this device) | K240377 | LIFEPAK® ECG Cables | Stryker Physio-Control | 2024 | — |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in DSA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new DSA device
- #1 K252827 IBP cable Becton Dickinson (2606488); IBP…score 11.3 · 4 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K211294 | Multi-Link X2 ECG Cable and Leadwire SystemSummary page 4…ce 510(k) Number(s) / Clearance Date Multi-Link™ X2 ECG Cable and Leadwire K211294/August 16, 2021 System 1 of 7 ==… | 2021 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jan 21, 2025 | Class II | Z-1102-2025 | Labeling Change Control | Product shipped with incorrect IFU. IFU does not include proper cleaning/disinfection instructions |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 30 years and passes through 7 companies; a new generation was cleared every 4.6 years on median.
- None of the 6 ancestors has a recall.
How does its category compare?
All DSA devicesDSA Cable, Transducer And Electrode, Patient, (Including Connector): FDA decisions, review times, recalls and adverse event reports over 20 years. 2024, this device’s year, is highlighted.
FDA decisions per year · DSA
4in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · DSA
105in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · DSA
1in 2022
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · DSA
67in 2025
MAUDE reports naming product code DSA, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 2 | — | 0 | 4 |
| 2008 | 6 | 59 | 0 | 11 |
| 2009 | 0 | — | 0 | 5 |
| 2010 | 4 | 55 | 2 | 1 |
| 2011 | 1 | — | 0 | 4 |
| 2012 | 1 | — | 0 | 8 |
| 2013 | 1 | — | 0 | 23 |
| 2014 | 1 | — | 0 | 67 |
| 2015 | 0 | — | 0 | 34 |
| 2016 | 2 | — | 0 | 28 |
| 2017 | 5 | 140 | 0 | 5 |
| 2018 | 3 | 84 | 0 | 16 |
| 2019 | 4 | 140 | 0 | 11 |
| 2020 | 6 | 177 | 0 | 31 |
| 2021 | 4 | 164 | 0 | 7 |
| 2022 | 2 | — | 1 | 49 |
| 2023 | 3 | 240 | 0 | 29 |
| 2024 | 3 | 102 | 0 | 40 |
| 2025 | 4 | 105 | 0 | 67 |
| 2026 | 1 | — | 0 | — |
What is it indicated for?
The LIFEPAK ECG cables are reusable electrode cable systems intended to transmit signals from
patient electrodes to the LIFEPAK monitor/defibrillator for monitoring purposes. LIFEPAK ECG
cables are intended for use in outdoor and indoor emergency care settings.
LIFEPAK ECG Cables:
- LIFEPAK 3-wire ECG Cable
- LIFEPAK 5-wire ECG Cable
- LIFEPAK 4-wire ECG cable
- LIFEPAK 3-wire extended precordial ECG Cable
- LIFEPAK 6-wire expandable precordial ECG cable
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K240377 · printed Nov 6, 2026
Cite this page
510(k) Report. “K240377: LIFEPAK® ECG Cables.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K240377
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026