510(k) · Decided
DecidedWallFlex Biliary RX Stent System; Epic Biliary Stent System; WallFlex Biliary RX Fully Covered Stent System RMV
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Boston Scientific · FGE Stents, Drains And Dilators For The Biliary Ducts
Cleared on Aug 30, 2024 as substantially equivalent to K140630 (Boston Scientific, 2014) and 2 other predicates after a 196-day review. Its family tree reaches back 13 generations to 1992. 1 later device descends from it.
- Decided
- Aug 30, 2024
- Review
- 196 days
- Class
- Class II
- Product code
- FGE
- Recalls
- None
- Predicates
- 3
Its longest line of descent runs back 13 generations to K923119 (1992). 40 ancestors in all, the oldest K834468 (1984). 1 later device descends from it, 1 citing it directly.
- Root · 13 generations backK923119 Wallstent Esophageal ProthesisSchneider U.S. Stent Div. · 1992
- 9 generations between
- 3 generations backK112543 Wallflex Biliary Transhepatic Uncovered Stent System, Wallflex…Boston Scientific Corporation · 2011
- 2 generations backK122072 Wallflextm Biliary RX Stent System (Uncovered, Partially Covered…Boston Scientific Corporation · 2012
- 1 generation backK140630 Wallflex Biliary RX Stent SystemBoston Scientific · 2014 · Recalled
- This deviceK240464 WallFlex Biliary RX Stent System; Epic Biliary Stent System…Boston Scientific · 2024
- Descendant · cites it directly · 0 of its ownK242950 WallFlex Biliary PLUS RX Stent SystemBoston Scientific · 2024
Show all 13 generations
- Root · 13 generations backK923119 Wallstent Esophageal ProthesisSchneider U.S. Stent Div. · 1992
- 12 generations backK945494 Wallstent (R) Tracheobronchial EndoprosthesisBoston Scientific Scimed, Inc. · 1995 · Recalled
- 11 generations backK961296 Wallstent Tracheobronchial EndoprosthesisBoston Scientific Scimed, Inc. · 1996 · Recalled
- 10 generations backK964121 Wallstent Tracheobronchial EndoprosthesisBoston Scientific Scimed, Inc. · 1996 · Recalled
- 9 generations backK980163 Wallstent Tracheobronchial Endoprosthesis and Wallstent…Boston Scientific Scimed, Inc. · 1998 · Recalled
- 8 generations backK982184 Wallstent EndoprosthesisBoston Scientific Scimed, Inc. · 1998
- 7 generations backK012752 Wallstent RX Biliary Endoprostheses, Models 6961-6972 InclusiveBoston Scientific Corp · 2001 · Recalled
- 6 generations backK030107 100MM Wallstent RX Biliary Endoprosthesis, Models M00569800 8MM…Boston Scientific Corp · 2003 · Recalled
- 5 generations backK061231 Wellflex Biliary RX Stent SystemBoston Scientific Corp · 2006
- 4 generations backK081733 10MM X 40MM Uncovered Wallflex Billiary RX Stent System, Model…Boston Scientific Corp · 2008
- 3 generations backK112543 Wallflex Biliary Transhepatic Uncovered Stent System, Wallflex…Boston Scientific Corporation · 2011
- 2 generations backK122072 Wallflextm Biliary RX Stent System (Uncovered, Partially Covered…Boston Scientific Corporation · 2012
- 1 generation backK140630 Wallflex Biliary RX Stent SystemBoston Scientific · 2014 · Recalled
- This deviceK240464 WallFlex Biliary RX Stent System; Epic Biliary Stent System…Boston Scientific · 2024
Most influential descendants
Showing 1 of 1 · See all- K242950WallFlex Biliary PLUS RX Stent SystemBoston Scientific · 2024 · cites it directly0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −13 | K923119 | Wallstent Esophageal Prothesis | Schneider U.S. Stent Div. | 1992 | — |
| −12 | K945494 | Wallstent (R) Tracheobronchial Endoprosthesis | Boston Scientific Scimed, Inc. | 1995 | 1 |
| −11 | K961296 | Wallstent Tracheobronchial Endoprosthesis | Boston Scientific Scimed, Inc. | 1996 | 1 |
| −10 | K964121 | Wallstent Tracheobronchial Endoprosthesis | Boston Scientific Scimed, Inc. | 1996 | 1 |
| −9 | K980163 | Wallstent Tracheobronchial Endoprosthesis and Wallstent Tracheobronchial… | Boston Scientific Scimed, Inc. | 1998 | 1 |
| −8 | K982184 | Wallstent Endoprosthesis | Boston Scientific Scimed, Inc. | 1998 | — |
| −7 | K012752 | Wallstent RX Biliary Endoprostheses, Models 6961-6972 Inclusive | Boston Scientific Corp | 2001 | 1 |
| −6 | K030107 | 100MM Wallstent RX Biliary Endoprosthesis, Models M00569800 8MM, M00569810 10MM | Boston Scientific Corp | 2003 | 1 |
| −5 | K061231 | Wellflex Biliary RX Stent System | Boston Scientific Corp | 2006 | — |
| −4 | K081733 | 10MM X 40MM Uncovered Wallflex Billiary RX Stent System, Model M00570890 | Boston Scientific Corp | 2008 | — |
| −3 | K112543 | Wallflex Biliary Transhepatic Uncovered Stent System, Wallflex Biliary… | Boston Scientific Corporation | 2011 | — |
| −2 | K122072 | Wallflextm Biliary RX Stent System (Uncovered, Partially Covered, and Fully… | Boston Scientific Corporation | 2012 | — |
| −1 | K140630 | Wallflex Biliary RX Stent System | Boston Scientific | 2014 | 1 |
| 0 (this device) | K240464 | WallFlex Biliary RX Stent System; Epic Biliary Stent System; WallFlex Biliary… | Boston Scientific | 2024 | — |
| +1 | K242950 | WallFlex Biliary PLUS RX Stent System | Boston Scientific | 2024 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1984 | 1 | 0 |
| 1991 | 2 | 0 |
| 1994 | 4 | 0 |
| 1995 | 1 | 0 |
| 1996 | 2 | 0 |
| 1997 | 2 | 0 |
| 1998 | 1 | 0 |
| 1999 | 1 | 0 |
| 2000 | 1 | 0 |
| 2001 | 1 | 0 |
| 2002 | 1 | 0 |
| 2004 | 2 | 0 |
| 2005 | 1 | 0 |
| 2008 | 1 | 0 |
| 2009 | 2 | 0 |
| 2010 | 1 | 0 |
| 2016 | 2 | 0 |
| 2017 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2024 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in FGE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new FGE device
- #1 K251123 Niti-S Biliary Stent; Niti-S Biliary Slim M…score 13.6 · 11 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (3)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K140630 | Wallflex Biliary RX Stent SystemSummary page 5…010 Product Code(s) FGE, PNB | Legally Marketed Predicate Devices 21 CFR 807.92(a)(3) Predicate # Predicate Trade Name (Primary Predicate is listed first) Product Code K140630 WallFlex™ Biliary RX Stent System FGE DEN150040 WallFlex™ Biliary RX Fully Covered Stent System RMV PNB K171809 Epic™ Biliary End… | 2014 |
| * | DEN150040 | WallFlex Biliary RX Fully Covered Stent System RMVSummary page 5…ed Predicate Devices 21 CFR 807.92(a)(3) Predicate # Predicate Trade Name (Primary Predicate is listed first) Product Code K140630 WallFlex™ Biliary RX Stent System FGE DEN150040 WallFlex™ Biliary RX Fully Covered Stent System RMV PNB K171809 Epic™ Biliary Endoscopic Stent System FGE Device Description Summ… | 2016 |
| * | K171809 | Epic Biliary Endoscopic Stent SystemSummary page 5…ade Name (Primary Predicate is listed first) Product Code K140630 WallFlex™ Biliary RX Stent System FGE DEN150040 WallFlex™ Biliary RX Fully Covered Stent System RMV PNB K171809 Epic™ Biliary Endoscopic Stent System FGE Device Description Summary 21 CFR 807.92(a)(4) The WallFlex™ Biliary RX Stent System i… | 2017 |
What descends from it? (1)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K242950 | WallFlex Biliary PLUS RX Stent System | 2024 | 0 |
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 32 years and passes through 5 companies; a new generation was cleared every 1.6 years on median.
- 14 of the 40 ancestors have a recall of any class.
- Of its descendants, K242950 (Boston Scientific, 2024) has gone furthest, with 0 descendants of its own.
- 1 other device also cites its predicate K140630.
How does its category compare?
All FGE devicesFGE Stents, Drains And Dilators For The Biliary Ducts: FDA decisions, review times, recalls and adverse event reports over 20 years. 2024, this device’s year, is highlighted.
FDA decisions per year · FGE
7in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · FGE
257in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · FGE
1in 2024
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · FGE
1,923in 2025
MAUDE reports naming product code FGE, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 4 | 316 | 54 | 1,006 |
| 2008 | 11 | 214 | 93 | 1,465 |
| 2009 | 7 | 160 | 18 | 1,655 |
| 2010 | 9 | 81 | 32 | 1,434 |
| 2011 | 9 | 114 | 2 | 1,321 |
| 2012 | 8 | 96 | 4 | 1,183 |
| 2013 | 5 | 161 | 3 | 1,060 |
| 2014 | 8 | 156 | 0 | 1,270 |
| 2015 | 8 | 79 | 0 | 1,543 |
| 2016 | 4 | 49 | 0 | 1,453 |
| 2017 | 9 | 59 | 1 | 1,515 |
| 2018 | 7 | 215 | 43 | 1,681 |
| 2019 | 2 | — | 13 | 1,896 |
| 2020 | 4 | 121 | 1 | 2,182 |
| 2021 | 1 | — | 27 | 2,297 |
| 2022 | 6 | 101 | 21 | 1,749 |
| 2023 | 6 | 199 | 19 | 1,615 |
| 2024 | 6 | 188 | 1 | 1,948 |
| 2025 | 7 | 257 | 0 | 1,923 |
| 2026 | 8 | 153 | 11 | — |
What is it indicated for?
The WallFlex Biliary RX Stent System is indicated for use in the palliative treatment of biliary strictures produced by malignant neoplasms and relief of malignant biliary obstruction prior to surgery.
The WallFlex Biliary RX Fully Covered Stent System RMV is indicated for use in the palliative treatment of biliary strictures produced by malignant neoplasms, relief of malignant biliary obstruction prior to surgery and for indwell up to 12 months in the treatment of benign biliary strictures secondary to chronic pancreatitis.
The Epic Biliary Endoscopic Stent System is indicated for palliation of malignant neoplasms in the biliary tree.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K240464 · printed Nov 5, 2026
Cite this page
510(k) Report. “K240464: WallFlex Biliary RX Stent System; Epic Biliary Stent System; WallFlex Biliary RX Fully Covered Stent System RMV.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K240464
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026