# K241217: CloudHRV™ System (100-01-001)

- Pathway: 510(k) (Traditional)
- Applicant: Inmedix, Inc.
- Decided: Jan 16, 2025 after 260 days
- Product code: DPS (Electrocardiograph)
- Device class: 2
- Regulation: 21 CFR 870.2340
- Recalls of this device: none
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 4 generations back to K950779 (1995). 10 ancestors in all.
No later device cites it as a predicate yet.
0 of 11 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K071168: The Portable Ansiscope, Dyansys, Inc. (2007) Summary page 5.

## Indications for use
The Inmedix® CloudHRV™ System is intended to acquire, display, and record electrocardiographic (ECG) information to measure heart rate variability (HRV) in adult patients (age 22 or above) with normal sinus rhythm and resting heart rate within 40 – 110 bpm. These measurements are not intended for any specific clinical diagnosis. The clinical significance of HRV must be determined by the physician. Assessment is indicated for patients in a healthcare facility including a physician office or a hospital outpatient clinic where the patient is able to remain supine and still. The CloudHRV™ System is not indicated for use in a surgical suite or during transport. Normal sinus rhythm is determined by the physician. The ECG is not intended to be used to diagnose or monitor cardiovascular conditions. The device does not provide assessments of cardiovascular abnormalities/arrhythmias.