510(k) Report

510(k) · Cleared

Cleared

Medline UNITE® MIS Foot Recon Screw System

K241359510(k) TraditionalCleared Aug 2, 202480-day reviewClass II21 CFR 888.3040

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medline Industries, LP · HWC Screw, Fixation, Bone

Cleared on Aug 2, 2024 as substantially equivalent to K130319 (Medline Industries, Inc., 2013) after an 80-day review. Its family tree reaches back 2 generations to 2008. 1 later device descends from it.

Decided
Aug 2, 2024
Review
80 days
Class
Class II
Product code
HWC
Recalls
None
Predicates
1

Its longest line of descent runs back 2 generations to K082320 (2008). 1 later device descends from it, 1 citing it directly.

Descendants, by year cleared 2025: 1 descendant clearedK082320 · Darco Small Screw · Wrightmedicaltechnologyinc · 2008K130319 · Medline Cannulated Screw · Medline Industries, Inc. · 201320052010201520202025
Descendants, by generation 1 generation forward: 1 descendant K130319 · Medline Cannulated Screw · Medline Industries, Inc. · 2013K082320 · Darco Small Screw · Wrightmedicaltechnologyinc · 20080
  1. Root · 2 generations backK082320 Darco Small ScrewWrightmedicaltechnologyinc · 2008
  2. 1 generation backK130319 Medline Cannulated ScrewMedline Industries, Inc. · 2013
  3. This deviceK241359 Medline UNITE® MIS Foot Recon Screw SystemMedline Industries, LP · 2024
  4. Descendant · cites it directly · 0 of its ownK243888 Medline UNITE® REFLEX® Hybrid Nitinol Implant SystemMedline Industries, LP · 2025

Most influential descendants

Showing 1 of 1 · See all
  • K243888Medline UNITE® REFLEX® Hybrid Nitinol Implant SystemMedline Industries, LP · 2025 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K082320Darco Small ScrewWrightmedicaltechnologyinc2008—
−1K130319Medline Cannulated ScrewMedline Industries, Inc.2013—
0 (this device)K241359Medline UNITE® MIS Foot Recon Screw SystemMedline Industries, LP2024—
+1K243888Medline UNITE® REFLEX® Hybrid Nitinol Implant SystemMedline Industries, LP2025—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202510
FIG. 1 — Lineage of K241359, 2008–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in HWC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HWC device

  1. #1 K203011 Monster Screw Systemscore 22.2 · 14 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K130319Medline Cannulated Screw
Summary page 4
…ion fastener Product Code: HWC Classification Panel: Orthopedics Regulatory Class: Class II Regulation Number: 21 CFR 888.3040 Predicate Device Medline Cannulated Screw K130319 Device Description The Medline UNITE® MIS Foot Recon Screw System includes screws manufactured from Titanium Alloy (Ti-6Al-4V EL…
Medline Industries, Inc.2013—

Also cleared against K130319: 8 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K243888Medline UNITE® REFLEX® Hybrid Nitinol Implant SystemMedline Industries, LP20250

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 16 years and passes through 3 companies; a new generation was cleared every 11.2 years on median.
  • None of the 2 ancestors has a recall.
  • Of its descendants, K243888 (Medline Industries, LP, 2025) has gone furthest, with 0 descendants of its own.
  • 8 other devices also cite its predicate K130319.

How does its category compare?

All HWC devices

HWC Screw, Fixation, Bone: FDA decisions, review times, recalls and adverse event reports over 20 years. 2024, this device’s year, is highlighted.

FDA decisions per year · HWC

15in 2025

2007: 352008: 332009: 272010: 232011: 352012: 202013: 252014: 312015: 412016: 262017: 402018: 202019: 302020: 282021: 262022: 262023: 242024: 172025: 152026 (partial year): 21
2007high 41 (2015) · low 15 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HWC

111in 2025

2007: 77 days median2008: 76 days median2009: 113 days median2010: 110 days median2011: 88 days median2012: 99 days median2013: 90 days median2014: 88 days median2015: 85 days median2016: 131 days median2017: 127 days median2018: 106 days median2019: 108 days median2020: 75 days median2021: 84 days median2022: 167 days median2023: 104 days median2024: 90 days median2025: 111 days median2026 (partial year): 125 days median
2007high 167 (2022) · low 75 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HWC

1in 2025

2007: 252008: 332009: 292010: 72011: 122012: 302013: 142014: 22015: 52016: 42017: 42018: 12019: 52020: 42021: 22022: 132025: 12026 (partial year): 2
2007high 33 (2008) · low 0 (2023)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · HWC

2,893in 2025

2007: 6692008: 6602009: 6512010: 7652011: 1,1042012: 3,3392013: 3,8502014: 2,5492015: 2,7112016: 2,2852017: 2,5612018: 3,6622019: 5,4632020: 4,8202021: 4,4462022: 3,9342023: 3,7182024: 3,5752025: 2,893
2007high 5,463 (2019) · low 651 (2009)2025

MAUDE reports naming product code HWC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007357725669
2008337633660
20092711329651
2010231107765
20113588121,104
20122099303,339
20132590143,850
2014318822,549
2015418552,711
20162613142,285
20174012742,561
20182010613,662
20193010855,463
2020287544,820
2021268424,446
202226167133,934
20232410403,718
2024179003,575
20251511112,893
2026211252—

What is it indicated for?

The Medline UNITE® MIS Foot Recon Screw System is indicated for the stabilization and fixation of bone reconstructions, osteotomies, joint fusions (arthrodeses), and fractures of short (tarsals) and long (metatarsals, phalanges, distal tibia and fibula) bones comprising the foot and ankle. Screws are intended for single use only.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K241359: Medline UNITE® MIS Foot Recon Screw System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K241359

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026