510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Straumann® Anatomic Healing Abutments XC (AHA)

K241575510(k) TraditionalCleared Nov 11, 2024161-day reviewClass II21 CFR 872.3630Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Straumann USA, LLC · NHA Abutment, Implant, Dental, Endosseous

Cleared on Nov 11, 2024 as substantially equivalent to K130808 (Straumann USA, 2013) after a 161-day review. Its family tree reaches back 11 generations to 1990. It has been recalled once.

Decided
Nov 11, 2024
Review
161 days
Class
Class II
Product code
NHA
Recalls
1
Predicates
1

Its longest line of descent runs back 11 generations to K894595 (1990). 66 ancestors in all, the oldest K894844 (1990). No later device cites it as a predicate yet.

Other ancestors, by year cleared 2007: 5 other ancestors2006: 1 other ancestor2001: 6 other ancestors1996: 7 other ancestors1990: 2 other ancestors2009: 2 other ancestors2005: 1 other ancestor2003: 6 other ancestors1999: 4 other ancestors2004: 1 other ancestor2002: 5 other ancestors1995: 1 other ancestor2000: 2 other ancestors1998: 4 other ancestors1997: 1 other ancestor1994: 3 other ancestors1993: 3 other ancestors1992: 1 other ancestorReference device: K212730 · BellaTek Encode Emergence Healing Abutments · Biomet 3i, LLC · 2021Reference device: K133421 · Straumann Magellan (TM) Abutment,Protective Cap,Titanium Copings,Gold Chopings,Basal Screw · Straumann USA, LLC · 2014Reference device: K173961 · Straumann BLX Implant System · Institut Straumann AG · 2018Reference device: K190662 · MRI Compatibility for Existing Straumann Dental Implant Systems · Straumann USA, LLC · 2019K894595 · Bonefit R Screw Implants · Institut Straumann AG · 1990K984104 · Iti Dental Implant System · Straumann USA · 1999 · recalledK992937 · Dental Implant · Nobel Biocare Uas, Inc. · 2000 · recalled×2: 2 generations cleared close together, K013067 (2002), K021318 (2002) K032158 · Xive Dental Implant System Multiple · Friadent GmbH · 2003K033984 · Straumann Dental Implant System · Institut Straumann AG · 2004K053088 · Slactive Implants · Institut Straumann AG · 2006K060958 · Straumann Narrow Neck Implants · Straumann USA · 2006K111357 · Straumann Narrow Neck Crossfit (NNC) 03.3MM Dental Implant System · Straumann USA · 2011K130808 · Straumann Healing Abutments, Healing Caps, Closure Screws · Straumann USA · 201319901995200020052010201520202025
  1. Root · 11 generations backK894595 Bonefit R Screw ImplantsInstitut Straumann AG · 1990
  2. 7 generations between
  3. 3 generations backK060958 Straumann Narrow Neck ImplantsStraumann USA · 2006
  4. 2 generations backK111357 Straumann Narrow Neck Crossfit (NNC) 03.3MM Dental Implant SystemStraumann USA · 2011
  5. 1 generation backK130808 Straumann Healing Abutments, Healing Caps, Closure ScrewsStraumann USA · 2013
  6. This deviceK241575 Straumann® Anatomic Healing Abutments XC (AHA)Straumann USA, LLC · 2024
Show all 11 generations
  1. Root · 11 generations backK894595 Bonefit R Screw ImplantsInstitut Straumann AG · 1990
  2. 10 generations backK984104 Iti Dental Implant SystemStraumann USA · 1999 · Recalled
  3. 9 generations backK992937 Dental ImplantNobel Biocare Uas, Inc. · 2000 · Recalled
  4. 8 generations backK013067 Frialoc Dental Implant SystemFriadent GmbH · 2002
  5. 7 generations backK021318 Xive Dental Implant SystemFriadent GmbH · 2002
  6. 6 generations backK032158 Xive Dental Implant System MultipleFriadent GmbH · 2003
  7. 5 generations backK033984 Straumann Dental Implant SystemInstitut Straumann AG · 2004
  8. 4 generations backK053088 Slactive ImplantsInstitut Straumann AG · 2006
  9. 3 generations backK060958 Straumann Narrow Neck ImplantsStraumann USA · 2006
  10. 2 generations backK111357 Straumann Narrow Neck Crossfit (NNC) 03.3MM Dental Implant SystemStraumann USA · 2011
  11. 1 generation backK130808 Straumann Healing Abutments, Healing Caps, Closure ScrewsStraumann USA · 2013
  12. This deviceK241575 Straumann® Anatomic Healing Abutments XC (AHA)Straumann USA, LLC · 2024
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−11K894595Bonefit R Screw ImplantsInstitut Straumann AG1990—
−10K984104Iti Dental Implant SystemStraumann USA19991
−9K992937Dental ImplantNobel Biocare Uas, Inc.20001
−8K013067Frialoc Dental Implant SystemFriadent GmbH2002—
−7K021318Xive Dental Implant SystemFriadent GmbH2002—
−6K032158Xive Dental Implant System MultipleFriadent GmbH2003—
−5K033984Straumann Dental Implant SystemInstitut Straumann AG2004—
−4K053088Slactive ImplantsInstitut Straumann AG2006—
−3K060958Straumann Narrow Neck ImplantsStraumann USA2006—
−2K111357Straumann Narrow Neck Crossfit (NNC) 03.3MM Dental Implant SystemStraumann USA2011—
−1K130808Straumann Healing Abutments, Healing Caps, Closure ScrewsStraumann USA2013—
0 (this device)K241575Straumann® Anatomic Healing Abutments XC (AHA)Straumann USA, LLC2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199020
199210
199330
199430
199510
199670
199710
199840
199940
200020
200160
200250
200360
200410
200510
200610
200750
200920
FIG. 1 — Lineage of K241575, 1990–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in NHA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NHA device

  1. #1 K222288 DESS Dental Smart Solutionsscore 30.7 · 19 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K130808Straumann Healing Abutments, Healing Caps, Closure Screws
Summary page 6
…K241575 Product Code(s): NHA Classification Panel: Dental Predicate Device(s) Primary Predicate: x Straumann Healing Abutments (K130808) Reference Devices: x BellaTek Encode Emergence Healing Abutment (K212730) x Protective caps of the Straumann Magell…
Straumann USA2013—
†K212730BellaTek Encode Emergence Healing AbutmentsBiomet 3i, LLC2021—
†K133421Straumann Magellan (TM) Abutment,Protective Cap,Titanium Copings,Gold Chopings,Basal ScrewStraumann USA, LLC2014—
†K173961Straumann BLX Implant SystemInstitut Straumann AG2018—
†K190662MRI Compatibility for Existing Straumann Dental Implant SystemsStraumann USA, LLC2019—

Also cleared against K130808: 4 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 17, 2025Class IIZ-1671-2025Under Investigation by firmIncorrect blister labelling.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 34 years and passes through 5 companies; a new generation was cleared every 1.7 years on median.
  • 6 of the 66 ancestors have a recall of any class.
  • 4 other devices also cite its predicate K130808.

How does its category compare?

All NHA devices

NHA Abutment, Implant, Dental, Endosseous: FDA decisions, review times, recalls and adverse event reports over 20 years. 2024, this device’s year, is highlighted.

FDA decisions per year · NHA

40in 2025

2007: 382008: 312009: 302010: 192011: 312012: 252013: 222014: 272015: 292016: 382017: 312018: 312019: 282020: 382021: 292022: 332023: 512024: 432025: 402026 (partial year): 23
2007high 51 (2023) · low 19 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NHA

147in 2025

2007: 91 days median2008: 90 days median2009: 95 days median2010: 162 days median2011: 130 days median2012: 171 days median2013: 184 days median2014: 160 days median2015: 151 days median2016: 202 days median2017: 188 days median2018: 200 days median2019: 160 days median2020: 153 days median2021: 211 days median2022: 142 days median2023: 147 days median2024: 158 days median2025: 147 days median2026 (partial year): 169 days median
2007high 211 (2021) · low 90 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · NHA

1in 2025

2007: 12008: 22011: 52012: 12013: 12015: 12016: 362017: 22018: 12020: 72023: 32025: 12026 (partial year): 3
2007high 36 (2016) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · NHA

838in 2025

2007: 12008: 82009: 302010: 612011: 562012: 52013: 1432014: 1462015: 8582016: 3142017: 5772018: 5912019: 5302020: 4582021: 5462022: 4972023: 6182024: 6352025: 838
2007high 858 (2015) · low 1 (2007)2025

MAUDE reports naming product code NHA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007389111
2008319028
20093095030
201019162061
201131130556
20122517115
2013221841143
2014271600146
2015291511858
20163820236314
2017311882577
2018312001591
2019281600530
2020381537458
2021292110546
2022331420497
2023511473618
2024431580635
2025401471838
2026231693—

What is it indicated for?

Straumann® Anatomic Healing Abutments are indicated to be placed in partially edentulous patients after implant placement. The healing components protect the inner configuration of the implant and form, maintain and stabilize the soft tissue during the healing process. Healing components have a maximum duration of usage of 180 days.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K241575: Straumann® Anatomic Healing Abutments XC (AHA).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K241575

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026