510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

uMI Panorama

K241585510(k) TraditionalCleared Aug 30, 202488-day reviewClass II21 CFR 892.1200Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Shanghai United Imaging Healthcare Co., Ltd. · KPS System, Tomography, Computed, Emission

Cleared on Aug 30, 2024 as substantially equivalent to K231572 (Shanghai United Imaging Healthcare Co., Ltd., 2023) after an 88-day review. Its family tree reaches back 16 generations to 2000. It has been recalled once.

Decided
Aug 30, 2024
Review
88 days
Class
Class II
Product code
KPS
Recalls
1
Predicates
1

Its longest line of descent runs back 16 generations to K002715 (2000). 22 ancestors in all, the oldest K843696 (1985). No later device cites it as a predicate yet.

Other ancestors, by year cleared 2004: 2 other ancestors2001: 1 other ancestor1998: 1 other ancestor1985: 2 other ancestorsReference device: K193241 · Emission Computed Tomography System · Shanghai United Imaging Healthcare Co., Ltd. · 2020 · recalledReference device: K232712 · uMI Panorama · Shanghai United Imaging Healthcare Co., Ltd. · 2023 · recalledReference device: K210001 · HYPER AiR · Shanghai United Imaging Healthcare Co., Ltd. · 2021K002715 · Ecat Pet Scanner · Cti Pet Systems, Inc. · 2000 · recalledK013504 · Modification to: Ecat Lso Pet/CT Scanner · Cti Pet Systems, Inc. · 2001 · recalledK023518 · Ecat Lso Pet/CT 16 · Cti Pet Systems, Inc. · 2002 · recalledK033431 · Lso Pet/CT Hirez and Lso Pet/CT 16 Hirez · Cti Pet Systems, Inc. · 2003 · recalledK042467 · Lso Pet/CT Hirez 64 · Cti Pet Systems, Inc. · 2004 · recalledK050509 · Lso Pet/CT Hirez Series, Models 2I, 6I, 16I, 64I · Cti Pet Systems, Inc. · 2005×2: 2 generations cleared close together, K060060 (2006), K060631 (2006) K081453 · Biograph HD Family · Siemens Medical Solutions USA, Inc. · 2008K101550 · Biograph MCT Family · Siemens Medical Solutions USA, Inc. · 2010K113448 · Biograph MCT Family · Siemens Medical Solutions USA, Inc. · 2011K123737 · Biograph MCT Family · Siemens Medical Solutions USA, Inc. · 2013 · recalledK151486 · Biograph mCT and mCT Flow PET/CT Scanners · Siemens Medical Solutions USA, Inc. · 2015 · recalledK172143 · uMI 780 PET/CT System · Shanghai United Imaging Healthcare Co., Ltd. · 2018 · recalled×2: 2 generations cleared close together, K223325 (2023), K231572 (2023) 1990200020102020
Other ancestors, by generation 10 generations back: 1 other ancestor11 generations back: 1 other ancestor12 generations back: 2 other ancestors13 generations back: 2 other ancestors K231572 · uMI Panorama · Shanghai United Imaging Healthcare Co., Ltd. · 2023 · recalledK223325 · uMI Panorama · Shanghai United Imaging Healthcare Co., Ltd. · 2023 · recalledK172143 · uMI 780 PET/CT System · Shanghai United Imaging Healthcare Co., Ltd. · 2018 · recalledK151486 · Biograph mCT and mCT Flow PET/CT Scanners · Siemens Medical Solutions USA, Inc. · 2015 · recalledK123737 · Biograph MCT Family · Siemens Medical Solutions USA, Inc. · 2013 · recalledK113448 · Biograph MCT Family · Siemens Medical Solutions USA, Inc. · 2011K101550 · Biograph MCT Family · Siemens Medical Solutions USA, Inc. · 2010K081453 · Biograph HD Family · Siemens Medical Solutions USA, Inc. · 2008K060631 · Biograph 64 and Biograph 40 · Siemens Medical Solutions USA, Inc. · 2006 · recalledK060060 · Biograph 6 · Siemens Medical Solutions USA, Inc. · 2006K050509 · Lso Pet/CT Hirez Series, Models 2I, 6I, 16I, 64I · Cti Pet Systems, Inc. · 2005K042467 · Lso Pet/CT Hirez 64 · Cti Pet Systems, Inc. · 2004 · recalledK033431 · Lso Pet/CT Hirez and Lso Pet/CT 16 Hirez · Cti Pet Systems, Inc. · 2003 · recalledK023518 · Ecat Lso Pet/CT 16 · Cti Pet Systems, Inc. · 2002 · recalledK013504 · Modification to: Ecat Lso Pet/CT Scanner · Cti Pet Systems, Inc. · 2001 · recalledK002715 · Ecat Pet Scanner · Cti Pet Systems, Inc. · 2000 · recalled−15−10−50
  1. Root · 16 generations backK002715 Ecat Pet ScannerCti Pet Systems, Inc. · 2000 · Recalled
  2. 12 generations between
  3. 3 generations backK172143 uMI 780 PET/CT SystemShanghai United Imaging Healthcare Co., Ltd. · 2018 · Recalled
  4. 2 generations backK223325 uMI PanoramaShanghai United Imaging Healthcare Co., Ltd. · 2023 · Recalled
  5. 1 generation backK231572 uMI PanoramaShanghai United Imaging Healthcare Co., Ltd. · 2023 · Recalled
  6. This deviceK241585 uMI PanoramaShanghai United Imaging Healthcare Co., Ltd. · 2024
Show all 16 generations
  1. Root · 16 generations backK002715 Ecat Pet ScannerCti Pet Systems, Inc. · 2000 · Recalled
  2. 15 generations backK013504 Modification to: Ecat Lso Pet/CT ScannerCti Pet Systems, Inc. · 2001 · Recalled
  3. 14 generations backK023518 Ecat Lso Pet/CT 16Cti Pet Systems, Inc. · 2002 · Recalled
  4. 13 generations backK033431 Lso Pet/CT Hirez and Lso Pet/CT 16 HirezCti Pet Systems, Inc. · 2003 · Recalled
  5. 12 generations backK042467 Lso Pet/CT Hirez 64Cti Pet Systems, Inc. · 2004 · Recalled
  6. 11 generations backK050509 Lso Pet/CT Hirez Series, Models 2I, 6I, 16I, 64ICti Pet Systems, Inc. · 2005
  7. 10 generations backK060060 Biograph 6Siemens Medical Solutions USA, Inc. · 2006
  8. 9 generations backK060631 Biograph 64 and Biograph 40Siemens Medical Solutions USA, Inc. · 2006 · Recalled
  9. 8 generations backK081453 Biograph HD FamilySiemens Medical Solutions USA, Inc. · 2008
  10. 7 generations backK101550 Biograph MCT FamilySiemens Medical Solutions USA, Inc. · 2010
  11. 6 generations backK113448 Biograph MCT FamilySiemens Medical Solutions USA, Inc. · 2011
  12. 5 generations backK123737 Biograph MCT FamilySiemens Medical Solutions USA, Inc. · 2013 · Recalled
  13. 4 generations backK151486 Biograph mCT and mCT Flow PET/CT ScannersSiemens Medical Solutions USA, Inc. · 2015 · Recalled
  14. 3 generations backK172143 uMI 780 PET/CT SystemShanghai United Imaging Healthcare Co., Ltd. · 2018 · Recalled
  15. 2 generations backK223325 uMI PanoramaShanghai United Imaging Healthcare Co., Ltd. · 2023 · Recalled
  16. 1 generation backK231572 uMI PanoramaShanghai United Imaging Healthcare Co., Ltd. · 2023 · Recalled
  17. This deviceK241585 uMI PanoramaShanghai United Imaging Healthcare Co., Ltd. · 2024
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−16K002715Ecat Pet ScannerCti Pet Systems, Inc.20002
−15K013504Modification to: Ecat Lso Pet/CT ScannerCti Pet Systems, Inc.20012
−14K023518Ecat Lso Pet/CT 16Cti Pet Systems, Inc.20021
−13K033431Lso Pet/CT Hirez and Lso Pet/CT 16 HirezCti Pet Systems, Inc.20031
−12K042467Lso Pet/CT Hirez 64Cti Pet Systems, Inc.20041
−11K050509Lso Pet/CT Hirez Series, Models 2I, 6I, 16I, 64ICti Pet Systems, Inc.2005—
−10K060060Biograph 6Siemens Medical Solutions USA, Inc.2006—
−9K060631Biograph 64 and Biograph 40Siemens Medical Solutions USA, Inc.20063
−8K081453Biograph HD FamilySiemens Medical Solutions USA, Inc.2008—
−7K101550Biograph MCT FamilySiemens Medical Solutions USA, Inc.2010—
−6K113448Biograph MCT FamilySiemens Medical Solutions USA, Inc.2011—
−5K123737Biograph MCT FamilySiemens Medical Solutions USA, Inc.201317
−4K151486Biograph mCT and mCT Flow PET/CT ScannersSiemens Medical Solutions USA, Inc.201511
−3K172143uMI 780 PET/CT SystemShanghai United Imaging Healthcare Co., Ltd.20182
−2K223325uMI PanoramaShanghai United Imaging Healthcare Co., Ltd.20232
−1K231572uMI PanoramaShanghai United Imaging Healthcare Co., Ltd.20231
0 (this device)K241585uMI PanoramaShanghai United Imaging Healthcare Co., Ltd.2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198520
199810
200110
200420
FIG. 1 — Lineage of K241585, 2000–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KPS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KPS device

  1. #1 K253564 uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI…score 18.7 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K231572uMI Panorama
Summary page 4
…e: Emission Computed Tomography System Regulatory Class: II Product Code: KPS, JAK Review Panel: Radiology 4. Predicate Device 510(k) Number: K231572 Device Name: uMI Panorama Model(s): uMI Panorama GS Regulation Name: Emission Computed Tomography System Regu…
Shanghai United Imaging Healthcare Co., Ltd.20231
†K193241Emission Computed Tomography SystemShanghai United Imaging Healthcare Co., Ltd.20206
†K232712uMI PanoramaShanghai United Imaging Healthcare Co., Ltd.20232
†K210001HYPER AiRShanghai United Imaging Healthcare Co., Ltd.2021—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 11, 2026Class IIZ-3285-2026Software designPositron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 24 years and passes through 3 companies; a new generation was cleared every 1.1 years on median.
  • 13 of the 22 ancestors have a recall of any class.

How does its category compare?

All KPS devices

KPS System, Tomography, Computed, Emission: FDA decisions, review times, recalls and adverse event reports over 20 years. 2024, this device’s year, is highlighted.

FDA decisions per year · KPS

12in 2025

2007: 82008: 172009: 92010: 32011: 72012: 62013: 112014: 112015: 32016: 112017: 112018: 132019: 32020: 122021: 152022: 52023: 122024: 112025: 122026 (partial year): 6
2007high 17 (2008) · low 3 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KPS

65in 2025

2007: 38 days median2008: 25 days median2009: 42 days median2010: 91 days median2011: 28 days median2012: 94 days median2013: 84 days median2014: 92 days median2015: 112 days median2016: 65 days median2017: 104 days median2018: 59 days median2019: 60 days median2020: 72 days median2021: 66 days median2022: 60 days median2023: 63 days median2024: 59 days median2025: 65 days median2026 (partial year): 67 days median
2007high 112 (2015) · low 25 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KPS

16in 2025

2007: 222008: 162009: 142010: 42011: 342012: 132013: 132014: 182015: 52016: 162017: 112018: 12, 3 Class I2019: 5, 5 Class I2020: 82021: 42022: 8, 6 Class I2023: 6, 3 Class I2024: 152025: 162026 (partial year): 6
2007high 34 (2011) · low 4 (2010)2026

Class I (17 since 2007) Class II/III not classified

Adverse event reports per year · KPS

36in 2025

2007: 382008: 112009: 152010: 442011: 562012: 342013: 2402014: 332015: 242016: 102017: 312018: 262019: 1052020: 222021: 102022: 152023: 222024: 232025: 36
2007high 240 (2013) · low 10 (2016)2025

MAUDE reports naming product code KPS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20078382238
200817251611
20099421415
2010391444
20117283456
20126941334
2013118413240
201411921833
20153112524
201611651610
2017111041131
2018135912 (3)26
20193605 (5)105
20201272822
20211566410
20225608 (6)15
202312636 (3)22
202411591523
202512651636
20266676—

What is it indicated for?

The uMI Panorama is a diagnostic imaging system that combines two existing imaging modalities PET and CT. The quantitative distribution information of PET radiopharmaceuticals within the patient body measured by PET can assist healthcare providers in assessing metabolic and physiological functions. CT provides diagnostic tomographic anatomical information as well as photon attenuation information for the scanned region. The accurate registration and…

Read the full statement (214 words)

The uMI Panorama is a diagnostic imaging system that combines two existing imaging modalities PET and CT. The quantitative distribution information of PET radiopharmaceuticals within the patient body measured by PET can assist healthcare providers in assessing metabolic and physiological functions. CT provides diagnostic tomographic anatomical information as well as photon attenuation information for the scanned region. The accurate registration and fusion of PET and CT images provides anatomical reference for the findings in the PET images.

This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases, inflammation, infection and disorders in, but not limited to oncology, cardiology and neurology. The system maintains independent functionality of the CT device, allowing for single modality CT diagnostic imaging.

This CT system can be used for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.

* Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K241585: uMI Panorama.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K241585

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026