510(k) Report

510(k) · Cleared

Cleared

SureCath Set

K242049510(k) TraditionalCleared Nov 26, 2024137-day reviewClass II21 CFR 876.5130

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Coloplast Corp. · EZD Catheter, Straight

Cleared on Nov 26, 2024 as substantially equivalent to K973070 (Coloplast Corp., 1997) after a 137-day review. Its family tree reaches back 2 generations to 1991.

Decided
Nov 26, 2024
Review
137 days
Class
Class II
Product code
EZD
Recalls
None
Predicates
1

Its longest line of descent runs back 2 generations to K910022 (1991). No later device cites it as a predicate yet.

  1. Root · 2 generations backK910022 MMG/O'neil Urinary Catheterization System KitMedical Marketing Group · 1991
  2. 1 generation backK973070 Conveen Easicath SetColoplast Corp. · 1997
  3. This deviceK242049 SureCath SetColoplast Corp. · 2024
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K910022MMG/O'neil Urinary Catheterization System KitMedical Marketing Group1991—
−1K973070Conveen Easicath SetColoplast Corp.1997—
0 (this device)K242049SureCath SetColoplast Corp.2024—
FIG. 1 — Lineage of K242049, 1991–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in EZD? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new EZD device

  1. #1 K241210 Luja Coudescore 16.2 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K973070Conveen Easicath Set
Our reading of the summary gave K973090, FDA’s number for Medstation (Program S.R.L.). Corrected to the number one digit away whose FDA name carries the summary’s words for the device, “Conveen EasiCath Set”.
Coloplast Corp.1997—
†K221401Self-Cath and Self-Cath PlusColoplast2022—

Also cleared against K973070: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 33 years and passes through 2 companies; a new generation was cleared every 27.1 years on median.
  • None of the 2 ancestors has a recall.
  • 2 other devices also cite its predicate K973070.

How does its category compare?

All EZD devices

EZD Catheter, Straight: FDA decisions, review times, recalls and adverse event reports over 20 years. 2024, this device’s year, is highlighted.

FDA decisions per year · EZD

11so far in 2026, the most since 2007

2007: 42008: 12009: 22010: 22011: 42012: 22013: 42014: 52015: 12016: 32017: 22018: 32020: 12022: 102023: 62024: 52025: 62026 (partial year): 11
2007high 11 (2026) · low 0 (2019)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · EZD

130in 2025

2007: 121 days median2011: 116 days median2013: 110 days median2014: 91 days median2016: 83 days median2018: 122 days median2022: 288 days median2023: 214 days median2024: 147 days median2025: 130 days median2026 (partial year): 146 days median
2007high 288 (2022) · low 83 (2016)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · EZD

1in 2014

2011: 12012: 12014: 1
2007high 1 (2011) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · EZD

288in 2025

2009: 42010: 62011: 42012: 22013: 182014: 172015: 212016: 212017: 1202018: 3902019: 4662020: 5632021: 9942022: 4872023: 5532024: 1162025: 288
2007high 994 (2021) · low 0 (2007)2025

MAUDE reports naming product code EZD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007412100
20081—00
20092—04
20102—06
2011411614
20122—12
20134110018
2014591117
20151—021
2016383021
20172—0120
201831220390
20190—0466
20201—0563
20210—0994
2022102880487
202362140553
202451470116
202561300288
2026111460—

What is it indicated for?

The product is intended for use in male, female, and pediatric patients requiring bladder drainage as determined by their physician. This device is indicated for those individuals unable to promote a natural urine flow or for those individuals who have a significant volume of residual urine following a natural bladder-voiding episode.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K242049: SureCath Set.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K242049

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026