510(k) Report

510(k) · Cleared

Cleared

Agility Symmetric™ Total Knee System

K242412510(k) TraditionalCleared Nov 6, 202484-day reviewClass II21 CFR 888.3565

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medisurge, LLC · MBH Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer

Cleared on Nov 6, 2024 as substantially equivalent to K080199 (Signal Medical Corporation, 2008) and 1 other predicate after an 84-day review. Its family tree reaches back 11 generations to 1981.

Decided
Nov 6, 2024
Review
84 days
Class
Class II
Product code
MBH
Recalls
None
Predicates
2

Its longest line of descent runs back 11 generations to K810258 (1981). 83 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 2008: 5 other ancestors2006: 2 other ancestors1995: 9 other ancestors2012: 1 other ancestor2010: 3 other ancestors2007: 5 other ancestors1999: 2 other ancestors1998: 1 other ancestor1996: 4 other ancestors2011: 1 other ancestor2009: 2 other ancestors2004: 3 other ancestors2003: 4 other ancestors2002: 4 other ancestors1994: 6 other ancestors1992: 4 other ancestors2005: 4 other ancestors2001: 4 other ancestors2000: 3 other ancestors1997: 3 other ancestors1991: 1 other ancestor1993: 1 other ancestorK810258 · Proximal Fermoral Hip Replacement · Biomet, Inc. · 1981K974558 · Salvage/Onocology Hip and Total Femur System · Biomet, Inc. · 1998K032396 · Ringloc 36MM Line and Modular Femoral Heads · Biomet, Inc. · 2003 · recalled×2: 2 generations cleared close together, K042051 (2005), K051411 (2005) K061312 · 36MM Biolox Delta Ceramic Heads · Biomet, Inc. · 2006K073102 · Biolox Delta Ceramic Heads with 100KGY E-Poly Acetabular Liners · Biomet Manufacturing Corp · 2007K100048 · E1 Antioxidant Infused Technology · Biomet Manufacturing Corp · 2010×2: 2 generations cleared close together, K120370 (2012), K121771 (2012) K141635 · Arthrex Ibalance Tka System · Arthrex, Inc. · 201419801990200020102020
  1. Root · 11 generations backK810258 Proximal Fermoral Hip ReplacementBiomet, Inc. · 1981
  2. 7 generations between
  3. 3 generations backK120370 Vivacit-E Vitamin E Highly Crosslinked Polyethylene LinersZimmer, Inc. · 2012
  4. 2 generations backK121771 Zimmer Persona the Personalized Knee SystemZimmer, Inc. · 2012 · Recalled
  5. 1 generation backK141635 Arthrex Ibalance Tka SystemArthrex, Inc. · 2014
  6. This deviceK242412 Agility Symmetric™ Total Knee SystemMedisurge, LLC · 2024
Show all 11 generations
  1. Root · 11 generations backK810258 Proximal Fermoral Hip ReplacementBiomet, Inc. · 1981
  2. 10 generations backK974558 Salvage/Onocology Hip and Total Femur SystemBiomet, Inc. · 1998
  3. 9 generations backK032396 Ringloc 36MM Line and Modular Femoral HeadsBiomet, Inc. · 2003 · Recalled
  4. 8 generations backK042051 Arcomxl Polyethylene LinersBiomet, Inc. · 2005 · Recalled
  5. 7 generations backK051411 Arcomxl Polyethylene Liners and Biolox Delta HeadsBiomet, Inc. · 2005 · Recalled
  6. 6 generations backK061312 36MM Biolox Delta Ceramic HeadsBiomet, Inc. · 2006
  7. 5 generations backK073102 Biolox Delta Ceramic Heads with 100KGY E-Poly Acetabular LinersBiomet Manufacturing Corp · 2007
  8. 4 generations backK100048 E1 Antioxidant Infused TechnologyBiomet Manufacturing Corp · 2010
  9. 3 generations backK120370 Vivacit-E Vitamin E Highly Crosslinked Polyethylene LinersZimmer, Inc. · 2012
  10. 2 generations backK121771 Zimmer Persona the Personalized Knee SystemZimmer, Inc. · 2012 · Recalled
  11. 1 generation backK141635 Arthrex Ibalance Tka SystemArthrex, Inc. · 2014
  12. This deviceK242412 Agility Symmetric™ Total Knee SystemMedisurge, LLC · 2024
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−11K810258Proximal Fermoral Hip ReplacementBiomet, Inc.1981—
−10K974558Salvage/Onocology Hip and Total Femur SystemBiomet, Inc.1998—
−9K032396Ringloc 36MM Line and Modular Femoral HeadsBiomet, Inc.20033
−8K042051Arcomxl Polyethylene LinersBiomet, Inc.20051
−7K051411Arcomxl Polyethylene Liners and Biolox Delta HeadsBiomet, Inc.20051
−6K06131236MM Biolox Delta Ceramic HeadsBiomet, Inc.2006—
−5K073102Biolox Delta Ceramic Heads with 100KGY E-Poly Acetabular LinersBiomet Manufacturing Corp2007—
−4K100048E1 Antioxidant Infused TechnologyBiomet Manufacturing Corp2010—
−3K120370Vivacit-E Vitamin E Highly Crosslinked Polyethylene LinersZimmer, Inc.2012—
−2K121771Zimmer Persona the Personalized Knee SystemZimmer, Inc.20122
−1K141635Arthrex Ibalance Tka SystemArthrex, Inc.2014—
0 (this device)K242412Agility Symmetric™ Total Knee SystemMedisurge, LLC2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199110
199240
199310
199460
199590
199640
199730
199810
199920
200030
200140
200240
200340
200430
200540
200620
200750
200850
200920
201030
201110
201210
FIG. 1 — Lineage of K242412, 1981–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MBH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MBH device

  1. #1 K241597 Freedom® Total Knee System - Porous Tibial…score 19.9 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K080199Symmetric Total Knee System
Summary page 4
…total knee replacement. The femoral component and tibial tray are for cemented use or uncemented use (biological fixation). The patellar component and metaphyseal stem (K080199) are for cemented use only. Indications for Posterior Stabilized Knee Replacement: Posterior stabilized knee system and systems…
Signal Medical Corporation20081
*K141635Arthrex Ibalance Tka System
Summary page 5
…Marketed Predicate Devices 21 CFR 807.92(a)(3) Predicate # Predicate Trade Name (Primary Predicate is listed first) Product Code K080199 Symmetric™ Total Knee System JWH K141635 Arthrex iBalance® TKA System MBH Device Description Summary 21 CFR 807.92(a)(4) The Agility Symmetric™ Total Knee System is inte…
Arthrex, Inc.2014—

Also cleared against K080199: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 43 years and passes through 5 companies; a new generation was cleared every 1.8 years on median.
  • 39 of the 83 ancestors have a recall of any class.
  • 1 other device also cites its predicate K080199.

How does its category compare?

All MBH devices

MBH Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer: FDA decisions, review times, recalls and adverse event reports over 20 years. 2024, this device’s year, is highlighted.

FDA decisions per year · MBH

8in 2025

2007: 102008: 32009: 32010: 42011: 52012: 52013: 62014: 92015: 82016: 42017: 42018: 62019: 12020: 72021: 62022: 72023: 62024: 102025: 82026 (partial year): 7
2007high 10 (2007) · low 1 (2019)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MBH

98in 2025

2007: 59 days median2008: 70 days median2009: 209 days median2010: 84 days median2011: 84 days median2012: 90 days median2013: 138 days median2014: 114 days median2015: 100 days median2016: 74 days median2017: 115 days median2018: 89 days median2020: 112 days median2021: 48 days median2022: 57 days median2023: 94 days median2024: 84 days median2025: 98 days median2026 (partial year): 153 days median
2007high 209 (2009) · low 48 (2021)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MBH

1in 2024

2009: 22011: 152012: 1, 1 Class I2013: 4, 4 Class I2016: 12017: 92018: 12019: 12020: 22023: 22024: 12026 (partial year): 4
2007high 15 (2011) · low 0 (2007)2026

Class I (5 since 2007) Class II/III not classified

Adverse event reports per year · MBH

1,561in 2025

2007: 272008: 212009: 892010: 1212011: 2472012: 3702013: 3082014: 3852015: 9552016: 1,0342017: 1,0072018: 1,1912019: 1,3942020: 1,0892021: 1,5392022: 1,2612023: 1,2632024: 1,4972025: 1,561
2007high 1,561 (2025) · low 21 (2008)2025

MAUDE reports naming product code MBH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071059027
2008370021
20093209289
20104840121
201158415247
20125901 (1)370
201361384 (4)308
201491140385
201581000955
201647411,034
2017411591,007
201868911,191
20191—11,394
2020711221,089
202164801,539
202275701,261
202369421,263
2024108411,497
202589801,561
202671534—

What is it indicated for?

Indications: Indications for Total Knee Replacement: 1. Rheumatoid arthritis, 2. Correction of functional deformity, 3. Osteoarthritis, 4. Degenerative arthritis in older patients whose age, weight and activity level are compatible with an adequate long-term result, 5. Failed osteotomies, unicompartmental replacement, or total knee replacement. The femoral component and tibial tray are for cemented use or uncemented use (biological fixation). The…

Read the full statement (167 words)

Indications:

Indications for Total Knee Replacement:

1. Rheumatoid arthritis,

2. Correction of functional deformity,

3. Osteoarthritis,

4. Degenerative arthritis in older patients whose age, weight and activity level are compatible with an adequate long-term result,

5. Failed osteotomies, unicompartmental replacement, or total knee replacement.

The femoral component and tibial tray are for cemented use or uncemented use (biological fixation).

The patellar component and metaphyseal stem (K080199) are for cemented use only.

Indications for Posterior Stabilized Knee Replacement:

Posterior stabilized knee system and systems with a deep articular surface are designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact. Total Knee indications still apply.

Indications for Constrained Knee Replacement:

Constrained knee systems are designed for use in patients in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral and/or lateral collateral ligament) are absent or incompetent. Total Knee indications still apply.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K242412: Agility Symmetric™ Total Knee System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K242412

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026