510(k) · Cleared
ClearedRapid Neuro3D
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Ischemaview, Inc. · QIH Automated Radiological Image Processing Software
Cleared on Jan 22, 2025 as substantially equivalent to K233582 (Ischemaview, Inc., 2024) after an 86-day review. Its family tree reaches back 15 generations to 1990. 1 later device descends from it.
- Decided
- Jan 22, 2025
- Review
- 86 days
- Class
- Class II
- Product code
- QIH
- Recalls
- None
- Predicates
- 1
Its longest line of descent runs back 15 generations to K896095 (1990). 21 ancestors in all. 1 later device descends from it, 1 citing it directly.
- Root · 15 generations backK896095 Images-on-Call Teleradiology SystemDevices & Services Co. · 1990
- 11 generations between
- 3 generations backK182130 iSchemaView RAPIDIschemaview, Inc. · 2018
- 2 generations backK213165 RapidIschemaview, Inc. · 2022
- 1 generation backK233582 RapidIschemaview, Inc. · 2024
- This deviceK243350 Rapid Neuro3DIschemaview, Inc. · 2025
- Descendant · cites it directly · 0 of its ownK251987 Rapid Aortic MeasurementsIschemaview, Inc. · 2025
Show all 15 generations
- Root · 15 generations backK896095 Images-on-Call Teleradiology SystemDevices & Services Co. · 1990
- 14 generations backK962699 Radworks Medical Imaging SoftwareApplicare Medical Imaging, B.V. · 1996
- 13 generations backK982862 Radworks Medical Imaging Software with Quality Control ModuleApplicare Medical Imaging, B.V. · 1998
- 12 generations backK012211 Efilm WorkstationEfilm Medical, Inc. · 2001 · Recalled
- 11 generations backK020995 Efilm Workstation with ModulesEfilm Medical, Inc. · 2002 · Recalled
- 10 generations backK063539 Nordic Image Control and Evaluation (Nice) Software, Version 2.1Nordicicemedical AS · 2007
- 9 generations backK082441 Nordicice SoftwareNordicimaginglab AS · 2008 · Recalled
- 8 generations backK090546 Modification to Nordicice SoftwareNordicimaginglab AS · 2009
- 7 generations backK111161 Perfscape V2.0Olea Medical · 2011
- 6 generations backK120196 Olea SphereOlea Medical · 2012
- 5 generations backK121447 Ischema View RapidIschemaview, Inc. · 2013
- 4 generations backK172477 iSchemaView RAPIDIschemaview, Inc. · 2018
- 3 generations backK182130 iSchemaView RAPIDIschemaview, Inc. · 2018
- 2 generations backK213165 RapidIschemaview, Inc. · 2022
- 1 generation backK233582 RapidIschemaview, Inc. · 2024
- This deviceK243350 Rapid Neuro3DIschemaview, Inc. · 2025
Most influential descendants
Showing 1 of 1 · See all- K251987Rapid Aortic MeasurementsIschemaview, Inc. · 2025 · cites it directly0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −15 | K896095 | Images-on-Call Teleradiology System | Devices & Services Co. | 1990 | — |
| −14 | K962699 | Radworks Medical Imaging Software | Applicare Medical Imaging, B.V. | 1996 | — |
| −13 | K982862 | Radworks Medical Imaging Software with Quality Control Module | Applicare Medical Imaging, B.V. | 1998 | — |
| −12 | K012211 | Efilm Workstation | Efilm Medical, Inc. | 2001 | 1 |
| −11 | K020995 | Efilm Workstation with Modules | Efilm Medical, Inc. | 2002 | 2 |
| −10 | K063539 | Nordic Image Control and Evaluation (Nice) Software, Version 2.1 | Nordicicemedical AS | 2007 | — |
| −9 | K082441 | Nordicice Software | Nordicimaginglab AS | 2008 | 2 |
| −8 | K090546 | Modification to Nordicice Software | Nordicimaginglab AS | 2009 | — |
| −7 | K111161 | Perfscape V2.0 | Olea Medical | 2011 | — |
| −6 | K120196 | Olea Sphere | Olea Medical | 2012 | — |
| −5 | K121447 | Ischema View Rapid | Ischemaview, Inc. | 2013 | — |
| −4 | K172477 | iSchemaView RAPID | Ischemaview, Inc. | 2018 | — |
| −3 | K182130 | iSchemaView RAPID | Ischemaview, Inc. | 2018 | — |
| −2 | K213165 | Rapid | Ischemaview, Inc. | 2022 | — |
| −1 | K233582 | Rapid | Ischemaview, Inc. | 2024 | — |
| 0 (this device) | K243350 | Rapid Neuro3D | Ischemaview, Inc. | 2025 | — |
| +1 | K251987 | Rapid Aortic Measurements | Ischemaview, Inc. | 2025 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1991 | 1 | 0 |
| 2006 | 1 | 0 |
| 2016 | 1 | 0 |
| 2018 | 2 | 0 |
| 2021 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2025 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in QIH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new QIH device
- #1 K213998 cvi42 Auto Imaging Software Applicationscore 23.5 · 7 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K233582 | RapidSummary page 6…vices Panel: Predicate Devices: iSchemaView’s Rapid Neuro3D is claimed to be substantially equivalent to the following legally marketed predicate device: Rapid (K233582). Device Description: Rapid Neuro 3D (RN3D) is a Software as a Medical Device (SaMD) image processing module and is part of the… | 2024 |
What descends from it? (1)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K251987 | Rapid Aortic Measurements | 2025 | 0 |
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 35 years and passes through 7 companies; a new generation was cleared every 2.2 years on median.
- 3 of the 21 ancestors have a recall of any class.
- Of its descendants, K251987 (Ischemaview, Inc., 2025) has gone furthest, with 0 descendants of its own.
- 1 other device also cites its predicate K233582.
How does its category compare?
All QIH devicesQIH Automated Radiological Image Processing Software: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.
FDA decisions per year · QIH
94in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · QIH
150in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · QIH
1so far in 2026, the most since 2007
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · QIH
4in 2025
MAUDE reports naming product code QIH, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 0 |
| 2008 | 0 | — | 0 | 0 |
| 2009 | 0 | — | 0 | 0 |
| 2010 | 0 | — | 0 | 0 |
| 2011 | 0 | — | 0 | 0 |
| 2012 | 0 | — | 0 | 0 |
| 2013 | 0 | — | 0 | 0 |
| 2014 | 0 | — | 0 | 0 |
| 2015 | 0 | — | 0 | 0 |
| 2016 | 0 | — | 0 | 0 |
| 2017 | 0 | — | 0 | 0 |
| 2018 | 0 | — | 0 | 0 |
| 2019 | 2 | — | 0 | 0 |
| 2020 | 11 | 145 | 0 | 0 |
| 2021 | 19 | 125 | 0 | 0 |
| 2022 | 25 | 134 | 0 | 0 |
| 2023 | 45 | 151 | 0 | 0 |
| 2024 | 64 | 141 | 0 | 2 |
| 2025 | 94 | 150 | 0 | 4 |
| 2026 | 78 | 130 | 1 | — |
What is it indicated for?
Rapid Neuro3D (RN3D) is an image analysis software for imaging datasets acquired with conventional CT Angiography (CTA) from the aortic arch to the vertex of the head. The module removes bone, tissue, and venous vessels, providing a 3D and 2D visualization of the neurovasculature supplying arterial blood to the brain. Outputs of the device include 3D rotational maximum intensity projections (MIPS), volume renders (VR), along with the curved planar…
Read the full statement (299 words)
Rapid Neuro3D (RN3D) is an image analysis software for imaging datasets acquired with
conventional CT Angiography (CTA) from the aortic arch to the vertex of the head. The module
removes bone, tissue, and venous vessels, providing a 3D and 2D visualization of the
neurovasculature supplying arterial blood to the brain.
Outputs of the device include 3D rotational maximum intensity projections (MIPS), volume renders
(VR), along with the curved planar reformation (CPR) of the isolated left and right internal carotid
and vertebral arteries.
Rapid Neuro3D is designed to support the physician in confirming the presence or absence of
physician-identified lesions and evaluation, documentation, and follow-up of any such lesion and
treatment planning.
Its results are not intended to be used on a stand-alone basis for clinical decision-making or
otherwise preclude clinical assessment.
RN3D is indicated for adults.
Precautions/Exclusions:
o Series containing excessive patient motion or metal implants may impact module output quality.
o The RN3D module will not process series that meet the following module exclusion criteria:
* Series containing inadequate contrast agent (<0.3 mL of right-hemisphere intracranial arterial
contrast media or <0.3 mL of left-hemisphere intracranial arterial contrast media, above 120 HU)
* Series acquired w/cone-beam CT scanners (c-arm CT)
» Series that are non-axial
Series with a non-supine patient position
* Series containing missing or improperly ordered slices (e.g., as a result of manual correction by
an imaging technician)
* CTA datasets with:
1) in-plane X and Y FOV < 160mm or > 400mm.
2) Z FOV (cranio-caudal transverse anatomical coverage) < 90 mm.
3) in-plane pixel spacing (X & Y resolution) < 0.2 mm or > 1.0 mm.
4) Z slice spacing of < 0.2 mm or > 1.25 mm.
5) slice thickness > 1.5mm.
6) data acquired at x-ray tube voltage < 70kVp or > 150kVp.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K243350 · printed Nov 5, 2026
Cite this page
510(k) Report. “K243350: Rapid Neuro3D.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K243350
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026