510(k) Report

510(k) · Cleared

Cleared

Rapid Neuro3D

K243350510(k) TraditionalCleared Jan 22, 202586-day reviewClass II21 CFR 892.2050

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ischemaview, Inc. · QIH Automated Radiological Image Processing Software

Cleared on Jan 22, 2025 as substantially equivalent to K233582 (Ischemaview, Inc., 2024) after an 86-day review. Its family tree reaches back 15 generations to 1990. 1 later device descends from it.

Decided
Jan 22, 2025
Review
86 days
Class
Class II
Product code
QIH
Recalls
None
Predicates
1

Its longest line of descent runs back 15 generations to K896095 (1990). 21 ancestors in all. 1 later device descends from it, 1 citing it directly.

  1. Root · 15 generations backK896095 Images-on-Call Teleradiology SystemDevices & Services Co. · 1990
  2. 11 generations between
  3. 3 generations backK182130 iSchemaView RAPIDIschemaview, Inc. · 2018
  4. 2 generations backK213165 RapidIschemaview, Inc. · 2022
  5. 1 generation backK233582 RapidIschemaview, Inc. · 2024
  6. This deviceK243350 Rapid Neuro3DIschemaview, Inc. · 2025
  7. Descendant · cites it directly · 0 of its ownK251987 Rapid Aortic MeasurementsIschemaview, Inc. · 2025
Show all 15 generations
  1. Root · 15 generations backK896095 Images-on-Call Teleradiology SystemDevices & Services Co. · 1990
  2. 14 generations backK962699 Radworks Medical Imaging SoftwareApplicare Medical Imaging, B.V. · 1996
  3. 13 generations backK982862 Radworks Medical Imaging Software with Quality Control ModuleApplicare Medical Imaging, B.V. · 1998
  4. 12 generations backK012211 Efilm WorkstationEfilm Medical, Inc. · 2001 · Recalled
  5. 11 generations backK020995 Efilm Workstation with ModulesEfilm Medical, Inc. · 2002 · Recalled
  6. 10 generations backK063539 Nordic Image Control and Evaluation (Nice) Software, Version 2.1Nordicicemedical AS · 2007
  7. 9 generations backK082441 Nordicice SoftwareNordicimaginglab AS · 2008 · Recalled
  8. 8 generations backK090546 Modification to Nordicice SoftwareNordicimaginglab AS · 2009
  9. 7 generations backK111161 Perfscape V2.0Olea Medical · 2011
  10. 6 generations backK120196 Olea SphereOlea Medical · 2012
  11. 5 generations backK121447 Ischema View RapidIschemaview, Inc. · 2013
  12. 4 generations backK172477 iSchemaView RAPIDIschemaview, Inc. · 2018
  13. 3 generations backK182130 iSchemaView RAPIDIschemaview, Inc. · 2018
  14. 2 generations backK213165 RapidIschemaview, Inc. · 2022
  15. 1 generation backK233582 RapidIschemaview, Inc. · 2024
  16. This deviceK243350 Rapid Neuro3DIschemaview, Inc. · 2025

Most influential descendants

Showing 1 of 1 · See all
  • K251987Rapid Aortic MeasurementsIschemaview, Inc. · 2025 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−15K896095Images-on-Call Teleradiology SystemDevices & Services Co.1990—
−14K962699Radworks Medical Imaging SoftwareApplicare Medical Imaging, B.V.1996—
−13K982862Radworks Medical Imaging Software with Quality Control ModuleApplicare Medical Imaging, B.V.1998—
−12K012211Efilm WorkstationEfilm Medical, Inc.20011
−11K020995Efilm Workstation with ModulesEfilm Medical, Inc.20022
−10K063539Nordic Image Control and Evaluation (Nice) Software, Version 2.1Nordicicemedical AS2007—
−9K082441Nordicice SoftwareNordicimaginglab AS20082
−8K090546Modification to Nordicice SoftwareNordicimaginglab AS2009—
−7K111161Perfscape V2.0Olea Medical2011—
−6K120196Olea SphereOlea Medical2012—
−5K121447Ischema View RapidIschemaview, Inc.2013—
−4K172477iSchemaView RAPIDIschemaview, Inc.2018—
−3K182130iSchemaView RAPIDIschemaview, Inc.2018—
−2K213165RapidIschemaview, Inc.2022—
−1K233582RapidIschemaview, Inc.2024—
0 (this device)K243350Rapid Neuro3DIschemaview, Inc.2025—
+1K251987Rapid Aortic MeasurementsIschemaview, Inc.2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199110
200610
201610
201820
202110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202510
FIG. 1 — Lineage of K243350, 1990–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in QIH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new QIH device

  1. #1 K213998 cvi42 Auto Imaging Software Applicationscore 23.5 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K233582Rapid
Summary page 6
…vices Panel: Predicate Devices: iSchemaView’s Rapid Neuro3D is claimed to be substantially equivalent to the following legally marketed predicate device: Rapid (K233582). Device Description: Rapid Neuro 3D (RN3D) is a Software as a Medical Device (SaMD) image processing module and is part of the…
Ischemaview, Inc.2024—

Also cleared against K233582: 1 other device →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K251987Rapid Aortic MeasurementsIschemaview, Inc.20250

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 35 years and passes through 7 companies; a new generation was cleared every 2.2 years on median.
  • 3 of the 21 ancestors have a recall of any class.
  • Of its descendants, K251987 (Ischemaview, Inc., 2025) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate K233582.

How does its category compare?

All QIH devices

QIH Automated Radiological Image Processing Software: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.

FDA decisions per year · QIH

94in 2025

2019: 22020: 112021: 192022: 252023: 452024: 642025: 942026 (partial year): 78
2007high 94 (2025) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · QIH

150in 2025

2020: 145 days median2021: 125 days median2022: 134 days median2023: 151 days median2024: 141 days median2025: 150 days median2026 (partial year): 130 days median
2007high 151 (2023) · low 125 (2021)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · QIH

1so far in 2026, the most since 2007

2026 (partial year): 1
2007high 1 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · QIH

4in 2025

2024: 22025: 4
2007high 4 (2025) · low 0 (2007)2025

MAUDE reports naming product code QIH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20130—00
20140—00
20150—00
20160—00
20170—00
20180—00
20192—00
20201114500
20211912500
20222513400
20234515100
20246414102
20259415004
2026781301—

What is it indicated for?

Rapid Neuro3D (RN3D) is an image analysis software for imaging datasets acquired with conventional CT Angiography (CTA) from the aortic arch to the vertex of the head. The module removes bone, tissue, and venous vessels, providing a 3D and 2D visualization of the neurovasculature supplying arterial blood to the brain. Outputs of the device include 3D rotational maximum intensity projections (MIPS), volume renders (VR), along with the curved planar…

Read the full statement (299 words)

Rapid Neuro3D (RN3D) is an image analysis software for imaging datasets acquired with

conventional CT Angiography (CTA) from the aortic arch to the vertex of the head. The module

removes bone, tissue, and venous vessels, providing a 3D and 2D visualization of the

neurovasculature supplying arterial blood to the brain.

Outputs of the device include 3D rotational maximum intensity projections (MIPS), volume renders

(VR), along with the curved planar reformation (CPR) of the isolated left and right internal carotid

and vertebral arteries.

Rapid Neuro3D is designed to support the physician in confirming the presence or absence of

physician-identified lesions and evaluation, documentation, and follow-up of any such lesion and

treatment planning.

Its results are not intended to be used on a stand-alone basis for clinical decision-making or

otherwise preclude clinical assessment.

RN3D is indicated for adults.

Precautions/Exclusions:

o Series containing excessive patient motion or metal implants may impact module output quality.

o The RN3D module will not process series that meet the following module exclusion criteria:

* Series containing inadequate contrast agent (<0.3 mL of right-hemisphere intracranial arterial

contrast media or <0.3 mL of left-hemisphere intracranial arterial contrast media, above 120 HU)

* Series acquired w/cone-beam CT scanners (c-arm CT)

» Series that are non-axial

Series with a non-supine patient position

* Series containing missing or improperly ordered slices (e.g., as a result of manual correction by

an imaging technician)

* CTA datasets with:

1) in-plane X and Y FOV < 160mm or > 400mm.

2) Z FOV (cranio-caudal transverse anatomical coverage) < 90 mm.

3) in-plane pixel spacing (X & Y resolution) < 0.2 mm or > 1.0 mm.

4) Z slice spacing of < 0.2 mm or > 1.25 mm.

5) slice thickness > 1.5mm.

6) data acquired at x-ray tube voltage < 70kVp or > 150kVp.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K243350: Rapid Neuro3D.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K243350

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026