510(k) Report

510(k) · Cleared

Cleared

Nutrifit

K243361510(k) TraditionalCleared Dec 19, 202451-day reviewClass II21 CFR 876.5980

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Vygon USA · FPD Tube, Feeding

Cleared on Dec 19, 2024 as substantially equivalent to K060944 (Vygon Corp., 2006) after a 51-day review. Its family tree reaches back 2 generations to 1982.

Decided
Dec 19, 2024
Review
51 days
Class
Class II
Product code
FPD
Recalls
None
Predicates
1

Its longest line of descent runs back 2 generations to K820176 (1982). 4 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 2002: 1 other ancestor1999: 1 other ancestorK820176 · Vygon Enternal Nutrition Tubes · Exco, Inc. · 1982K060944 · Nutrisafe 2 · Vygon Corp. · 2006 · recalled19801990200020102020
Other ancestors, by generation 2 generations back: 2 other ancestors K060944 · Nutrisafe 2 · Vygon Corp. · 2006 · recalledK820176 · Vygon Enternal Nutrition Tubes · Exco, Inc. · 19820
  1. Root · 2 generations backK820176 Vygon Enternal Nutrition TubesExco, Inc. · 1982
  2. 1 generation backK060944 Nutrisafe 2Vygon Corp. · 2006 · Recalled
  3. This deviceK243361 NutrifitVygon USA · 2024
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K820176Vygon Enternal Nutrition TubesExco, Inc.1982—
−1K060944Nutrisafe 2Vygon Corp.20063
0 (this device)K243361NutrifitVygon USA2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199910
200210
FIG. 1 — Lineage of K243361, 1982–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FPD? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K060944Nutrisafe 2
Summary page 5
…Product Code(s) FPD, KNT | Legally Marketed Predicate Devices Predicate # Predicate Trade Name (Primary Predicate is listed first) 21 CFR 807.92(a)(3) Product Code K060944 Nutrisafe 2 FPD Device Description Summary 21 CFR 807.92(a)(4) enteral nutrition and for a short-term use, not longer than 29…
Vygon Corp.20063

Also cleared against K060944: 6 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 42 years and passes through 3 companies; a new generation was cleared every 24.6 years on median.
  • 2 of the 4 ancestors have a recall of any class.
  • 6 other devices also cite its predicate K060944.

How does its category compare?

All FPD devices

FPD Tube, Feeding: FDA decisions, review times, recalls and adverse event reports over 20 years. 2024, this device’s year, is highlighted.

FDA decisions per year · FPD

1in 2024

2007: 12008: 42009: 42010: 42011: 22012: 12013: 22019: 12024: 1
2007high 4 (2008) · low 0 (2014)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FPD

118in 2010

2008: 76 days median2009: 69 days median2010: 118 days median
2007high 118 (2010) · low 69 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FPD

1in 2019

2015: 12018: 32019: 1
2007high 3 (2018) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FPD

10in 2025

2007: 192008: 242009: 402010: 212011: 152012: 322013: 382014: 642015: 552016: 1022017: 342018: 302019: 192020: 312021: 242022: 192023: 552024: 412025: 10
2007high 102 (2016) · low 10 (2025)2025

MAUDE reports naming product code FPD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—019
2008476024
2009469040
20104118021
20112—015
20121—032
20132—038
20140—064
20150—155
20160—0102
20170—034
20180—330
20191—119
20200—031
20210—024
20220—019
20230—055
20241—041
20250—010
20260—0—

What is it indicated for?

For nasogastric/ oralgastric enteral feeding, incorporating safety connectors which eliminates the risk of IV administration through the feeding tube.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K243361: Nutrifit.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K243361

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026