510(k) Report

510(k) · Cleared

Cleared

LIAISON PLEX Gram-Positive Blood Culture Assay

K243490510(k) TraditionalCleared Jun 6, 2025206-day reviewClass II21 CFR 866.3365

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Luminex Corporation · PAM Gram-Positive Bacteria And Their Resistance Markers

Cleared on Jun 6, 2025 as substantially equivalent to K122514 (Nanosphere, Inc., 2012) after a 206-day review. Its family tree reaches back 2 generations to 2012.

Decided
Jun 6, 2025
Review
206 days
Class
Class II
Product code
PAM
Recalls
None
Predicates
1

Its longest line of descent runs back 2 generations to DEN120014 (2012). No later device cites it as a predicate yet.

  1. Root · 2 generations backDEN120014 Verigene Gram Positive Blood Culture Nucleic Acid Test (BC-GP)Nanosphere, Inc. · 2012
  2. 1 generation backK122514 Verigene Gram-Positive Blood Culture (GC-GP) Nucleic Acid TestNanosphere, Inc. · 2012 · Recalled
  3. This deviceK243490 LIAISON PLEX Gram-Positive Blood Culture AssayLuminex Corporation · 2025
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2DEN120014Verigene Gram Positive Blood Culture Nucleic Acid Test (BC-GP)Nanosphere, Inc.2012—
−1K122514Verigene Gram-Positive Blood Culture (GC-GP) Nucleic Acid TestNanosphere, Inc.20128
0 (this device)K243490LIAISON PLEX Gram-Positive Blood Culture AssayLuminex Corporation2025—
FIG. 1 — Lineage of K243490, 2012–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in PAM? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new PAM device

  1. #1 K254194 QIAstat-Dx BCID GPF Plus AMR Panelscore 12.7 · 4 generations · no Class I recall
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K122514Verigene Gram-Positive Blood Culture (GC-GP) Nucleic Acid Test
Summary page 9
…uivalence Information: 1. Predicate device name(s): VERIGENE Blood Culture Gram-Positive (BC-GP) Nucleic Acid Test 2. Predicate 510(k) number(s): K122514 3. Comparison with predicate: The following tables compare the LIAISON PLEX® Gram-Positive Blood Culture Assay to the…
Nanosphere, Inc.20128

Also cleared against K122514: 3 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 13 years and passes through 2 companies; a new generation was cleared every 12.7 years on median.
  • 1 of the 2 ancestors has a recall of any class.
  • 3 other devices also cite its predicate K122514.

How does its category compare?

All PAM devices

PAM Gram-Positive Bacteria And Their Resistance Markers: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.

FDA decisions per year · PAM

1in 2025

2012: 22016: 12017: 12018: 12025: 12026 (partial year): 1
2007high 2 (2012) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · PAM

Too few PAM 510(k)s in any year since 2007 to give a median.

Recalls per year · PAM

1in 2023

2016: 12018: 32020: 12023: 1
2007high 3 (2018) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · PAM

75in 2025

2015: 12020: 22021: 22022: 12023: 1062024: 4222025: 75
2007high 422 (2024) · low 0 (2007)2025

MAUDE reports naming product code PAM, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20122—00
20130—00
20140—00
20150—01
20161—10
20171—00
20181—30
20190—00
20200—12
20210—02
20220—01
20230—1106
20240—0422
20251—075
20261—0—

What is it indicated for?

The LIAISON PLEX® Gram-Positive Blood Culture Assay (BCP), performed using the automated, sample-to-result LIAISON PLEX® System, is a qualitative multiplexed in vitro diagnostic test for the simultaneous detection and identification of selected gram-positive pathogens and/or selected genetic determinants associated with antimicrobial resistance in positive blood culture bottles. BCP is performed directly on blood culture media using blood culture bottles…

Read the full statement (359 words)

The LIAISON PLEX® Gram-Positive Blood Culture Assay (BCP), performed using the automated, sample-to-result LIAISON PLEX® System, is a qualitative multiplexed in vitro diagnostic test for the simultaneous detection and identification of selected gram-positive pathogens and/or selected genetic determinants associated with antimicrobial resistance in positive blood culture bottles. BCP is performed directly on blood culture media using blood culture bottles identified as positive by a continuous monitoring blood culture system and which contain gram-positive bacteria as determined by Gram stain.

The BCP Assay detects and identifies the following:

Gram Positive Resistance Markers(a):

mecA/mecC

vanA

vanB

Genera and Species:

Bacillus spp.

Enterococcus faecalis

Enterococcus faecium

Listeria spp.

Staphylococcus spp.

Staphylococcus aureus

Staphylococcus epidermidis

Staphylococcus lugdunensis

Streptococcus spp.

Streptococcus agalactiae

Streptococcus anginosus group

Streptococcus pneumoniae

Streptococcus pyogenes

(a) Negative results for antimicrobial resistance genes do not indicate bacterial susceptibility as there are multiple mechanisms that can contribute to resistance.

The LIAISON PLEX® BCP Assay contains targets for the detection of genetic determinants associated with resistance to methicillin (mecA/C) and vancomycin (vanA and vanB) to aid in the identification of potentially antimicrobial-resistant organisms in positive blood culture samples. In mixed growth, the LIAISON PLEX BCP Assay does not specifically attribute vanA/vanB-mediated vancomycin resistance to either E. faecalis or E. faecium, or mecA/mecC-mediated methicillin resistance to either Staphylococcus spp., S. aureus, S. epidermidis or S. lugdunensis.

The antimicrobial resistance gene or marker detected may or may not be associated with the agent responsible for disease. Negative results for these select antimicrobial resistance gene and marker assays do not indicate susceptibility, as multiple mechanisms of methicillin and vancomycin resistance exist.

The LIAISON PLEX® BCP Assay is indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial bloodstream infections (BSI). The LIAISON PLEX® BCP Assay is not intended to monitor these infections. Sub-culturing of positive blood cultures is necessary to recover organisms for antimicrobial susceptibility testing (AST), for identification of organisms not detected by the LIAISON PLEX BCP Assay, to detect mixed infections that may not be detected by the LIAISON PLEX BCP Assay, for association of antimicrobial resistance genes to a specific organism, or for epidemiological typing.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K243490: LIAISON PLEX Gram-Positive Blood Culture Assay.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K243490

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026