510(k) Report

510(k) · Cleared

Cleared

DCM (PW-DCM)

K243513510(k) TraditionalCleared Apr 16, 2025155-day reviewClass II21 CFR 882.5050

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Pneumowave, Ltd. · LEL Device, Sleep Assessment

Cleared on Apr 16, 2025 as substantially equivalent to K231532 (Actigraph, LLC, 2023) after a 155-day review. Its family tree reaches back 7 generations to 1986.

Decided
Apr 16, 2025
Review
155 days
Class
Class II
Product code
LEL
Recalls
None
Predicates
1

Its longest line of descent runs back 7 generations to K854030 (1986). No later device cites it as a predicate yet.

  1. Root · 7 generations backK854030 Wrist ActigraphAmbulatory Monitoring, Inc. · 1986
  2. 6 generations backK983533 ActiwatchMini-Mitter Co., Inc. · 1999 · Recalled
  3. 5 generations backK992410 ActitracIndividual Monitoring Systems, Inc. · 1999
  4. 4 generations backK040554 Actigraph, Model AM7164-2.2 and Actigraph with Extended Memory…Manufacturing Technology, Inc. · 2004
  5. 3 generations backK080545 ActitrainerActigraph, LLC · 2008
  6. 2 generations backK181077 ActiGraph CentrePoint Insight WatchActigraph · 2018
  7. 1 generation backK231532 ActiGraph LEAP activity monitor (ActiGraph LEAP)Actigraph, LLC · 2023
  8. This deviceK243513 DCM (PW-DCM)Pneumowave, Ltd. · 2025
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K854030Wrist ActigraphAmbulatory Monitoring, Inc.1986—
−6K983533ActiwatchMini-Mitter Co., Inc.19993
−5K992410ActitracIndividual Monitoring Systems, Inc.1999—
−4K040554Actigraph, Model AM7164-2.2 and Actigraph with Extended Memory, Model AM71256Manufacturing Technology, Inc.2004—
−3K080545ActitrainerActigraph, LLC2008—
−2K181077ActiGraph CentrePoint Insight WatchActigraph2018—
−1K231532ActiGraph LEAP activity monitor (ActiGraph LEAP)Actigraph, LLC2023—
0 (this device)K243513DCM (PW-DCM)Pneumowave, Ltd.2025—
FIG. 1 — Lineage of K243513, 1986–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LEL? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LEL device

  1. #1 K252981 EmbraceMiniscore 23.3 · 40 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K231532ActiGraph LEAP activity monitor (ActiGraph LEAP)
Summary page 5
…ce Classification Name: Biofeedback Device Classification Codes: LEL Regulation Number: 21 CFR §882.5050 Predicate Devices: Actigraph LEAP (K231532) Actigraph, LLC Indications for Use: DCM is a small worn activity monitor designed for documenting physical movement as…
Actigraph, LLC2023—
†K132764Motion Watch and Pro-DiaryCamntech, Ltd.2014—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 39 years and passes through 7 companies; a new generation was cleared every 4.8 years on median.
  • 1 of the 7 ancestors has a recall of any class.

How does its category compare?

All LEL devices

LEL Device, Sleep Assessment: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.

FDA decisions per year · LEL

3in 2025

2011: 12014: 12016: 12018: 12023: 12024: 22025: 3
2007high 3 (2025) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LEL

96in 2025

2025: 96 days median
2007high 96 (2025)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LEL

No recall names LEL since 2007.

Adverse event reports per year · LEL

1in 2021

2007: 12012: 12018: 32021: 1
2007high 3 (2018) · low 0 (2008)2025

MAUDE reports naming product code LEL, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—01
20080—00
20090—00
20100—00
20111—00
20120—01
20130—00
20141—00
20150—00
20161—00
20170—00
20181—03
20190—00
20200—00
20210—01
20220—00
20231—00
20242—00
202539600
20260—0—

What is it indicated for?

DCM is a small worn activity monitor designed for documenting physical movement associated with applications in physiological monitoring.

The device is intended to monitor the activity associated with movement during sleep.

DCM can be used to analyze circadian rhythms and assess activity in any instance where quantifiable analysis of physical motion is desirable.

DCM is indicated for monitoring of adult patients only.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K243513: DCM (PW-DCM).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K243513

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026