510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Teslatome Bipolar Sphincterotome; Tesla BiCord Active Cord

K243568510(k) SpecialCleared Jan 17, 202560-day reviewClass II21 CFR 876.4300Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Wilson-Cook Medical, Inc. · KNS Unit, Electrosurgical, Endoscopic (With Or Without Accessories)

Cleared on Jan 17, 2025 as substantially equivalent to K192339 (Wilson-Cook Medical, Inc., 2020) after a 60-day review. Its family tree reaches back 7 generations to 1991. It has been recalled once.

Decided
Jan 17, 2025
Review
60 days
Class
Class II
Product code
KNS
Recalls
1
Predicates
1

Its longest line of descent runs back 7 generations to K910722 (1991). 16 ancestors in all, the oldest K901443 (1990). No later device cites it as a predicate yet.

  1. Root · 7 generations backK910722 450CM Guide Wire for G.I. UseTerumo Corp. · 1991
  2. 6 generations backK971937 Climber - Partially Coated Guide Wire for Endoscopic UseTerumo Medical Corp. · 1997
  3. 5 generations backK021179 Disposable Guidewire, Models G-205-3545S & G-205-3545AOlympus Optical Co., Ltd. · 2002
  4. 4 generations backK091417 Single Use Guidewire, Models G-240-2527S, G-240-2545S, G-240-2527A…Terumo Corp. · 2009
  5. 3 generations backK122505 Single Use Preloaded Sphincterotome VOlympus Medical Systems Corp. · 2013
  6. 2 generations backK141991 Single Use Preloaded Sphincterotome V (Distal Wireguided); Single…Olympus Medical Systems Copr. · 2015 · Recalled
  7. 1 generation backK192339 TeslaTome Bipolar Sphincterotome, Tesla Bipolar Active CordWilson-Cook Medical, Inc. · 2020
  8. This deviceK243568 Teslatome Bipolar Sphincterotome; Tesla BiCord Active CordWilson-Cook Medical, Inc. · 2025
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K910722450CM Guide Wire for G.I. UseTerumo Corp.1991—
−6K971937Climber - Partially Coated Guide Wire for Endoscopic UseTerumo Medical Corp.1997—
−5K021179Disposable Guidewire, Models G-205-3545S & G-205-3545AOlympus Optical Co., Ltd.2002—
−4K091417Single Use Guidewire, Models G-240-2527S, G-240-2545S, G-240-2527A…Terumo Corp.2009—
−3K122505Single Use Preloaded Sphincterotome VOlympus Medical Systems Corp.2013—
−2K141991Single Use Preloaded Sphincterotome V (Distal Wireguided); Single Use SPOlympus Medical Systems Copr.20153
−1K192339TeslaTome Bipolar Sphincterotome, Tesla Bipolar Active CordWilson-Cook Medical, Inc.2020—
0 (this device)K243568Teslatome Bipolar Sphincterotome; Tesla BiCord Active CordWilson-Cook Medical, Inc.2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199010
199210
199310
199510
199710
200110
200310
200510
201810
FIG. 1 — Lineage of K243568, 1991–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KNS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KNS device

  1. #1 K241271 Aqua Medical RF Vapor Ablation Systemscore 17.2 · 20 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K192339TeslaTome Bipolar Sphincterotome, Tesla Bipolar Active Cord
Summary page 5
…876.4300 Product Code(s) KNS I Legally Marketed Predicate Devices 21 CFR 807.92(a)(3) Predicate # Predicate Trade Name (Primary Predicate is listed first) Product Code K192339 TeslaTome Bipolar Sphincterotome, Tesla Bipolar Active Cord KNS Device Description Summary 21 CFR 807.92(a)(4) In a sphincteroto…
Wilson-Cook Medical, Inc.2020—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 18, 2025Class IIZ-1362-2026Under Investigation by firmDue to product that may contain a void in the material that separates the wire guide and cutting wire lumens, which may cause an inappropriate path of current during cutting.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 34 years and passes through 6 companies; a new generation was cleared every 4.9 years on median.
  • 5 of the 16 ancestors have a recall of any class.

How does its category compare?

All KNS devices

KNS Unit, Electrosurgical, Endoscopic (With Or Without Accessories): FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.

FDA decisions per year · KNS

11in 2025

2007: 12008: 32009: 32010: 42011: 12012: 22013: 22014: 12015: 32016: 32017: 32018: 52019: 72020: 42021: 42022: 32023: 92024: 32025: 112026 (partial year): 2
2007high 11 (2025) · low 1 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KNS

171in 2025

2008: 61 days median2009: 212 days median2010: 168 days median2015: 252 days median2016: 30 days median2017: 48 days median2018: 62 days median2019: 66 days median2020: 212 days median2021: 139 days median2022: 182 days median2023: 83 days median2024: 198 days median2025: 171 days median
2007high 252 (2015) · low 30 (2016)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KNS

14in 2024

2007: 32008: 52009: 22011: 22016: 62017: 12021: 412024: 142026 (partial year): 4
2007high 41 (2021) · low 0 (2010)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KNS

1,504in 2025

2007: 1642008: 5872009: 1,0272010: 9082011: 8602012: 8482013: 4492014: 4052015: 4062016: 4572017: 4462018: 5632019: 6682020: 9062021: 8142022: 9132023: 9762024: 1,5332025: 1,504
2007high 1,533 (2024) · low 164 (2007)2025

MAUDE reports naming product code KNS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—3164
20083615587
2009321221,027
201041680908
20111—2860
20122—0848
20132—0449
20141—0405
201532520406
20163306457
20173481446
20185620563
20197660668
202042120906
2021413941814
202231820913
20239830976
20243198141,533
20251117101,504
20262—4—

What is it indicated for?

This device is used for cannulation of the ductal system and for sphincterotomy. If preloaded, the device also aids in bridging difficult strictures during ERCP. Also indicated for sphincterotome-aided, wire-guided selective cannulation of the biliary ducts.

Tesla BiCord Active Cord:

This device is used to connect the Teslatome Bipolar Sphincterotome to compatible electrosurgical generators.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K243568: Teslatome Bipolar Sphincterotome; Tesla BiCord Active Cord.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K243568

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026