510(k) Report

510(k) · Cleared

Cleared

Sterile Powder Free Synthetic Rubber Surgeon’s Gloves, Green Color, Tested For Use With Chemotherapy Drugs and Gastric Acid

K250313510(k) AbbreviatedCleared Aug 19, 2025197-day reviewClass I21 CFR 878.4460

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Harps Europe Manufacturing GmbH · KGO Surgeon'S Gloves

Cleared on Aug 19, 2025 as substantially equivalent to K160781 (Semperit Investments Asia Pte, Ltd., 2017) after a 197-day review. Its family tree reaches back 3 generations to 1998.

Decided
Aug 19, 2025
Review
197 days
Class
Class I
Product code
KGO
Recalls
None
Predicates
1

Its longest line of descent runs back 3 generations to K981096 (1998). No later device cites it as a predicate yet.

  1. Root · 3 generations backK981096 Sempermed Surgeons' Gloves - SupremeSemperit Technische Produkte Gesellshaft · 1998
  2. 2 generations backK013560 Sempermed Supreme, Sizes 5.5, 6, 6.5, 7, 7.5, 8, 8.5, and 9Semperit Technische Produkte Gesellshaft · 2002
  3. 1 generation backK160781 Sempermed Syntegra IR, Sterile Powder Free Synthetic Rubber Surgeons…Semperit Investments Asia Pte, Ltd. · 2017
  4. This deviceK250313 Sterile Powder Free Synthetic Rubber Surgeon’s Gloves, Green Color…Harps Europe Manufacturing GmbH · 2025
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K981096Sempermed Surgeons' Gloves - SupremeSemperit Technische Produkte Gesellshaft1998—
−2K013560Sempermed Supreme, Sizes 5.5, 6, 6.5, 7, 7.5, 8, 8.5, and 9Semperit Technische Produkte Gesellshaft2002—
−1K160781Sempermed Syntegra IR, Sterile Powder Free Synthetic Rubber Surgeons GlovesSemperit Investments Asia Pte, Ltd.2017—
0 (this device)K250313Sterile Powder Free Synthetic Rubber Surgeon’s Gloves, Green Color, Tested For…Harps Europe Manufacturing GmbH2025—
FIG. 1 — Lineage of K250313, 1998–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KGO? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KGO device

  1. #1 K192328 JR.MEDIK Latex Surgeon’s Gloves Powder Freescore 15.2 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K160781Sempermed Syntegra IR, Sterile Powder Free Synthetic Rubber Surgeons Gloves
Summary page 6
…I Product Code: KGO, LZC Classification Panel: General and Plastic Surgery Legally Marketed Predicate Device Primary Predicate: K160781 Sempermed Syntegra IR, Sterile Powder Free Synthetic Rubber Surgeon…
Semperit Investments Asia Pte, Ltd.2017—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 27 years and passes through 3 companies; a new generation was cleared every 8.3 years on median.
  • None of the 3 ancestors has a recall.

How does its category compare?

All KGO devices

KGO Surgeon'S Gloves: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.

FDA decisions per year · KGO

1in 2025

2007: 102008: 22009: 32010: 62011: 112012: 72013: 62014: 72015: 122016: 92017: 122018: 72019: 112020: 152021: 142022: 62023: 142024: 22025: 12026 (partial year): 2
2007high 15 (2020) · low 1 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KGO

185in 2023

2007: 113 days median2009: 48 days median2010: 87 days median2011: 90 days median2012: 133 days median2013: 184 days median2014: 257 days median2015: 225 days median2016: 218 days median2017: 111 days median2018: 194 days median2019: 204 days median2020: 117 days median2021: 199 days median2022: 127 days median2023: 185 days median
2007high 257 (2014) · low 48 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KGO

2in 2023

2009: 152010: 12016: 22017: 12018: 82022: 22023: 2
2007high 15 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KGO

21in 2025

2007: 132008: 82009: 142010: 162011: 142012: 142013: 412014: 282015: 112016: 152017: 202018: 152019: 512020: 152021: 142022: 102023: 282024: 182025: 21
2007high 51 (2019) · low 8 (2008)2025

MAUDE reports naming product code KGO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200710113013
20082—08
20093481514
2010687116
20111190014
20127133014
20136184041
20147257028
201512225011
20169218215
201712111120
20187194815
201911204051
202015117015
202114199014
20226127210
202314185228
20242—018
20251—021
20262—0—

What is it indicated for?

The Sterile Powder Free Synthetic Rubber Surgeon's Gloves, Green Color, Tested For Use With Chemotherapy Drugs and Gastric Acid is a single use disposable device intended to be worn by operating room personnel to protect a surgical wound from contamination. The tested drugs are: Compound Minimum Breakthrough Time (minutes) Carmustine (BCNU) (3.3 mg/ml) 15.8 Cyclophosphamide (20 mg/ml) >240 Doxorubicin (2 mg/ml) >240 Etoposide (Toposar) (20 mg/ml) >240…

Read the full statement (248 words)

The Sterile Powder Free Synthetic Rubber Surgeon's Gloves, Green Color, Tested For Use With Chemotherapy Drugs and Gastric Acid is a single use disposable device intended to be worn by operating room personnel to protect a surgical wound from contamination.

The tested drugs are:

Compound Minimum Breakthrough Time (minutes)

Carmustine (BCNU) (3.3 mg/ml) 15.8

Cyclophosphamide (20 mg/ml) >240

Doxorubicin (2 mg/ml) >240

Etoposide (Toposar) (20 mg/ml) >240

Fluorouracil (50 mg/ml) >240

Paclitaxel (6 mg/ml) >240

Thiotepa (THT) (10 mg/ml) 24.6

Bleomycin sulfate (15 mg/ml) >240

Carboplatin (10 mg/ml) >240

Cisplatin (1 mg/ml) >240

Cytarabine (100 mg/ml) >240

Dacarbazine (10 mg/ml) >240

Daunorubicin HCl (5 mg/ml) >240

Docetaxel (10 mg/ml) >240

Gemcitabine HCl (38 mg/ml) >240

Idarubicin HCl (1 mg/ml) >240

Ifosfamide (50 mg/ml) >240

Irinotecan HCl (20 mg/ml) >240

Mechlorethamine HCl (1 mg/ml) >240

Melphalan HCl (5 mg/ml) >240

Methotrexate (25 mg/ml) >240

Mitomycin C (0.5 mg/ml) >240

Mitoxantrone (2 mg/ml) >240

Vincristine Sulfate (1 mg/ml) >240

Busulfan (6 mg/ml) >240

Chloroquine (50 mg/ml) >240

Cyclosporin A (100 mg/ml) >240

Epirubicin HCl (2 mg/ml) >240

Fludarabine Phosphate (25 mg/ml) >240

Oxaliplatin (2 mg/ml) >240

Retrovir (10 mg/ml) >240

Rituximab (10 mg/ml) >240

Topotecan HCl (1 mg/ml) >240

Trisenox (Arsenic Trioxide) (1 mg/ml) >240

Velcade (Bortezomib) (1 mg/ml) >240

Simulated Gastric Acid Fluid was also tested with a minimum breakthrough greater than 240 minutes.

WARNING:

Please note the following drugs have extremely low permeation times:

Carmustine: 15.8 minutes and Thiotepa: 24.6 minutes. Do not use with Carmustine and Thiotepa.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K250313: Sterile Powder Free Synthetic Rubber Surgeon’s Gloves, Green Color, Tested For Use With Chemotherapy Drugs and Gastric Acid.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K250313

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026