510(k) Report

510(k) · Cleared

Cleared

SnapHammer Hammertoe Correction System

K250613510(k) SpecialCleared Apr 2, 202533-day reviewClass II21 CFR 888.3040

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Nvision Biomedical Technologies, Inc. · HWC Screw, Fixation, Bone

Cleared on Apr 2, 2025 as substantially equivalent to K231453 (Nvision Biomedical Technologies, Inc., 2024) after a 33-day review. Its family tree reaches back 13 generations to 1979.

Decided
Apr 2, 2025
Review
33 days
Class
Class II
Product code
HWC
Recalls
None
Predicates
1

Its longest line of descent runs back 13 generations to K792022 (1979). 99 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 2021: 1 other ancestor2020: 1 other ancestor2019: 1 other ancestor2016: 1 other ancestor2015: 1 other ancestor2014: 3 other ancestors2011: 2 other ancestors2013: 1 other ancestor2012: 5 other ancestors2010: 3 other ancestors2008: 2 other ancestors2007: 4 other ancestors2006: 3 other ancestors2005: 4 other ancestors2002: 5 other ancestors2003: 4 other ancestors2001: 7 other ancestors2000: 2 other ancestors1998: 3 other ancestors1996: 10 other ancestors2004: 4 other ancestors1997: 5 other ancestors1990: 2 other ancestors1980: 1 other ancestor1999: 3 other ancestors1995: 4 other ancestors1994: 2 other ancestors1992: 1 other ancestor1993: 1 other ancestorReference device: K210424 · Quantum Anterior Cervical Plate · Nvision Biomedical Technologies, Inc. · 2021Reference device: K182949 · Healix™ Compression Screw System · Nvision Biomedical Technologies, LLC · 2018K792022 · Herbert Bone Screw · Zimmer, Inc. · 1979 · recalledK974876 · Bionx Distal Radius Screw · Bionx Implants, Inc. · 1998K992947 · Cannulated Smartscrew Models 224540C, 224570C · Bionx Implants, Ltd. · 1999K012001 · Smartscrew · Bionx Implants, Ltd. · 2001K030900 · Inion Otps Biodegradable Fixaton System · Inion, Ltd. · 2003K043248 · Tightrope Syndesmosis Repair Kit, Titanium, Model AR-8920DS; Stainless Steel. Model AR-8921DS · Arthrex, Inc. · 2005K052776 · Tightrope Acromioclavicular (AC) Device, Titanium, Stainless Steel , Models AR-2257, AR-TBD · Arthrex, Inc. · 2005K061925 · Mini Tightrope Repair Kit, Model AR-8911DS · Arthrex, Inc. · 2006K141420 · Fastforward · MedShape, Inc. · 2014K160264 · Additive Orthopaedics Hammertoe Correction System · Additive Orthopaedics, LLC · 2016K190926 · Hammertoe Truss System (HTS) · 4Web, Inc. · 2019K221283 · Hammertoe Truss System (HTS) · 4Web Medical, Inc. · 2022K231453 · SnapHammer Hammertoe Correction System · Nvision Biomedical Technologies, Inc. · 202419801990200020102020
Other ancestors, by generation 2 generations back: 1 other ancestor3 generations back: 2 other ancestors4 generations back: 4 other ancestors5 generations back: 14 other ancestors6 generations back: 16 other ancestors7 generations back: 14 other ancestors8 generations back: 9 other ancestors9 generations back: 10 other ancestors10 generations back: 7 other ancestors11 generations back: 6 other ancestors12 generations back: 3 other ancestors K231453 · SnapHammer Hammertoe Correction System · Nvision Biomedical Technologies, Inc. · 2024K221283 · Hammertoe Truss System (HTS) · 4Web Medical, Inc. · 2022K190926 · Hammertoe Truss System (HTS) · 4Web, Inc. · 2019K160264 · Additive Orthopaedics Hammertoe Correction System · Additive Orthopaedics, LLC · 2016K141420 · Fastforward · MedShape, Inc. · 2014K061925 · Mini Tightrope Repair Kit, Model AR-8911DS · Arthrex, Inc. · 2006K052776 · Tightrope Acromioclavicular (AC) Device, Titanium, Stainless Steel , Models AR-2257, AR-TBD · Arthrex, Inc. · 2005K043248 · Tightrope Syndesmosis Repair Kit, Titanium, Model AR-8920DS; Stainless Steel. Model AR-8921DS · Arthrex, Inc. · 2005K030900 · Inion Otps Biodegradable Fixaton System · Inion, Ltd. · 2003K012001 · Smartscrew · Bionx Implants, Ltd. · 2001K992947 · Cannulated Smartscrew Models 224540C, 224570C · Bionx Implants, Ltd. · 1999K974876 · Bionx Distal Radius Screw · Bionx Implants, Inc. · 1998K792022 · Herbert Bone Screw · Zimmer, Inc. · 1979 · recalled−10−50
  1. Root · 13 generations backK792022 Herbert Bone ScrewZimmer, Inc. · 1979 · Recalled
  2. 9 generations between
  3. 3 generations backK190926 Hammertoe Truss System (HTS)4Web, Inc. · 2019
  4. 2 generations backK221283 Hammertoe Truss System (HTS)4Web Medical, Inc. · 2022
  5. 1 generation backK231453 SnapHammer Hammertoe Correction SystemNvision Biomedical Technologies, Inc. · 2024
  6. This deviceK250613 SnapHammer Hammertoe Correction SystemNvision Biomedical Technologies, Inc. · 2025
Show all 13 generations
  1. Root · 13 generations backK792022 Herbert Bone ScrewZimmer, Inc. · 1979 · Recalled
  2. 12 generations backK974876 Bionx Distal Radius ScrewBionx Implants, Inc. · 1998
  3. 11 generations backK992947 Cannulated Smartscrew Models 224540C, 224570CBionx Implants, Ltd. · 1999
  4. 10 generations backK012001 SmartscrewBionx Implants, Ltd. · 2001
  5. 9 generations backK030900 Inion Otps Biodegradable Fixaton SystemInion, Ltd. · 2003
  6. 8 generations backK043248 Tightrope Syndesmosis Repair Kit, Titanium, Model AR-8920DS…Arthrex, Inc. · 2005
  7. 7 generations backK052776 Tightrope Acromioclavicular (AC) Device, Titanium, Stainless Steel…Arthrex, Inc. · 2005
  8. 6 generations backK061925 Mini Tightrope Repair Kit, Model AR-8911DSArthrex, Inc. · 2006
  9. 5 generations backK141420 FastforwardMedShape, Inc. · 2014
  10. 4 generations backK160264 Additive Orthopaedics Hammertoe Correction SystemAdditive Orthopaedics, LLC · 2016
  11. 3 generations backK190926 Hammertoe Truss System (HTS)4Web, Inc. · 2019
  12. 2 generations backK221283 Hammertoe Truss System (HTS)4Web Medical, Inc. · 2022
  13. 1 generation backK231453 SnapHammer Hammertoe Correction SystemNvision Biomedical Technologies, Inc. · 2024
  14. This deviceK250613 SnapHammer Hammertoe Correction SystemNvision Biomedical Technologies, Inc. · 2025
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−13K792022Herbert Bone ScrewZimmer, Inc.19796
−12K974876Bionx Distal Radius ScrewBionx Implants, Inc.1998—
−11K992947Cannulated Smartscrew Models 224540C, 224570CBionx Implants, Ltd.1999—
−10K012001SmartscrewBionx Implants, Ltd.2001—
−9K030900Inion Otps Biodegradable Fixaton SystemInion, Ltd.2003—
−8K043248Tightrope Syndesmosis Repair Kit, Titanium, Model AR-8920DS; Stainless Steel.…Arthrex, Inc.2005—
−7K052776Tightrope Acromioclavicular (AC) Device, Titanium, Stainless Steel , Models…Arthrex, Inc.2005—
−6K061925Mini Tightrope Repair Kit, Model AR-8911DSArthrex, Inc.2006—
−5K141420FastforwardMedShape, Inc.2014—
−4K160264Additive Orthopaedics Hammertoe Correction SystemAdditive Orthopaedics, LLC2016—
−3K190926Hammertoe Truss System (HTS)4Web, Inc.2019—
−2K221283Hammertoe Truss System (HTS)4Web Medical, Inc.2022—
−1K231453SnapHammer Hammertoe Correction SystemNvision Biomedical Technologies, Inc.2024—
0 (this device)K250613SnapHammer Hammertoe Correction SystemNvision Biomedical Technologies, Inc.2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198010
199020
199210
199310
199420
199540
1996100
199750
199830
199930
200020
200170
200250
200340
200440
200540
200630
200740
200820
201030
201120
201250
201310
201430
201510
201610
201910
202010
202110
FIG. 1 — Lineage of K250613, 1979–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in HWC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HWC device

  1. #1 K203011 Monster Screw Systemscore 22.2 · 14 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K231453SnapHammer Hammertoe Correction System
Summary page 6
…510(k) Number Predicate Device Name / Manufacturer Primary Predicate K231453 Nvision Biomedical’s SnapHammer Hammertoe ✓ Correction System Nvision Biome…
Nvision Biomedical Technologies, Inc.2024—
†K210424Quantum Anterior Cervical PlateNvision Biomedical Technologies, Inc.2021—
†K182949Healix™ Compression Screw SystemNvision Biomedical Technologies, LLC2018—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 46 years and passes through 10 companies; a new generation was cleared every 1.7 years on median.
  • 9 of the 99 ancestors have a recall of any class.

How does its category compare?

All HWC devices

HWC Screw, Fixation, Bone: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.

FDA decisions per year · HWC

15in 2025

2007: 352008: 332009: 272010: 232011: 352012: 202013: 252014: 312015: 412016: 262017: 402018: 202019: 302020: 282021: 262022: 262023: 242024: 172025: 152026 (partial year): 21
2007high 41 (2015) · low 15 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HWC

111in 2025

2007: 77 days median2008: 76 days median2009: 113 days median2010: 110 days median2011: 88 days median2012: 99 days median2013: 90 days median2014: 88 days median2015: 85 days median2016: 131 days median2017: 127 days median2018: 106 days median2019: 108 days median2020: 75 days median2021: 84 days median2022: 167 days median2023: 104 days median2024: 90 days median2025: 111 days median2026 (partial year): 125 days median
2007high 167 (2022) · low 75 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HWC

1in 2025

2007: 252008: 332009: 292010: 72011: 122012: 302013: 142014: 22015: 52016: 42017: 42018: 12019: 52020: 42021: 22022: 132025: 12026 (partial year): 2
2007high 33 (2008) · low 0 (2023)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · HWC

2,893in 2025

2007: 6692008: 6602009: 6512010: 7652011: 1,1042012: 3,3392013: 3,8502014: 2,5492015: 2,7112016: 2,2852017: 2,5612018: 3,6622019: 5,4632020: 4,8202021: 4,4462022: 3,9342023: 3,7182024: 3,5752025: 2,893
2007high 5,463 (2019) · low 651 (2009)2025

MAUDE reports naming product code HWC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007357725669
2008337633660
20092711329651
2010231107765
20113588121,104
20122099303,339
20132590143,850
2014318822,549
2015418552,711
20162613142,285
20174012742,561
20182010613,662
20193010855,463
2020287544,820
2021268424,446
202226167133,934
20232410403,718
2024179003,575
20251511112,893
2026211252—

What is it indicated for?

The SnapHammer Hammertoe Correction System is indicated for the fixation of osteotomies and reconstruction of the lesser toes following correction procedures for hammertoe, claw toe, and mallet toe. Cannulated implants in the SnapHammer Hammertoe Correction System can be used with K-wires for the delivery of implants or the temporary stabilization of outlying joints (e.g. MTP Joint).

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K250613: SnapHammer Hammertoe Correction System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K250613

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026