510(k) Report

510(k) · Cleared

Cleared

Tru Kidney Health Test Panel; Tru Analyzer

K251058510(k) TraditionalCleared Nov 26, 2025236-day reviewClass II21 CFR 862.1225

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Truvian Health · CGX Alkaline Picrate, Colorimetry, Creatinine

Cleared on Nov 26, 2025 as substantially equivalent to K242685 (Siemens Healthcare Diagnostics, Inc., 2024) and 1 other predicate after a 236-day review. Its family tree reaches back 4 generations to 1998.

Decided
Nov 26, 2025
Review
236 days
Class
Class II
Product code
CGX
Recalls
None
Predicates
2

Its longest line of descent runs back 4 generations to K973993 (1998). 9 ancestors in all, the oldest K942782 (1995). No later device cites it as a predicate yet.

  1. Root · 4 generations backK973993 Randox CreatinineRandox Laboratories, Ltd. · 1998
  2. 3 generations backK070727 Advia Chemistry Enzymatic Creatinine_2Siemens Medical Solutions Diagnostics · 2007 · Recalled
  3. 2 generations backK161494 Atellica CH Creatinine_2 (Crea_2), Atellica CH Chemistry Calibrator…Siemens Healthcare Diagnostics, Inc. · 2016 · Recalled
  4. 1 generation backK242685 Atellica® CH Creatinine_3 (Crea3)Siemens Healthcare Diagnostics, Inc. · 2024
  5. This deviceK251058 Tru Kidney Health Test Panel; Tru AnalyzerTruvian Health · 2025
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K973993Randox CreatinineRandox Laboratories, Ltd.1998—
−3K070727Advia Chemistry Enzymatic Creatinine_2Siemens Medical Solutions Diagnostics20075
−2K161494Atellica CH Creatinine_2 (Crea_2), Atellica CH Chemistry Calibrator (CHEM CAL)Siemens Healthcare Diagnostics, Inc.20161
−1K242685Atellica® CH Creatinine_3 (Crea3)Siemens Healthcare Diagnostics, Inc.2024—
0 (this device)K251058Tru Kidney Health Test Panel; Tru AnalyzerTruvian Health2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199520
199910
200510
201810
FIG. 1 — Lineage of K251058, 1998–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CGX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CGX device

  1. #1 K242685 Atellica® CH Creatinine_3 (Crea3)score 12.1 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K242685Atellica® CH Creatinine_3 (Crea3)
Summary page 5
…510(k) Exempt* * Meets limitations of exemption per 21 CFR § 862.9 (c)(9) 3 Predicate Devices Primary Predicate Device: Atellica CH Creatinine_3 (Crea3) (K242685) Additional Predicate Devices: Skyla Comprehensive Metabolic Panel (K171971) Skyla Clin…
Siemens Healthcare Diagnostics, Inc.2024—
*K171971Comprehensive Metabolic Panel, skyla Clinical Chemistry Analyzer, Minicare C300 Clinical…
Summary page 5
…Predicate Devices Primary Predicate Device: Atellica CH Creatinine_3 (Crea3) (K242685) Additional Predicate Devices: Skyla Comprehensive Metabolic Panel (K171971) Skyla Clinical Chemistry Analyzer (K171971)…
Lite-On Technology Corp. H.S.P.B.2018—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 27 years and passes through 4 companies; a new generation was cleared every 9.3 years on median.
  • 6 of the 9 ancestors have a recall of any class.

How does its category compare?

All CGX devices

CGX Alkaline Picrate, Colorimetry, Creatinine: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.

FDA decisions per year · CGX

1in 2025

2007: 42009: 12012: 12013: 12014: 22016: 12022: 12024: 22025: 12026 (partial year): 1
2007high 4 (2007) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CGX

25in 2007

2007: 25 days median
2007high 25 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CGX

2in 2016

2008: 12009: 22010: 22011: 12012: 22016: 22026 (partial year): 1
2007high 2 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CGX

34in 2025

2007: 152008: 272009: 52010: 52011: 322012: 132013: 242014: 132015: 132016: 32017: 212018: 142019: 252020: 192021: 132022: 232023: 192024: 322025: 34
2007high 34 (2025) · low 3 (2016)2025

MAUDE reports naming product code CGX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007425015
20080—127
20091—25
20100—25
20110—132
20121—213
20131—024
20142—013
20150—013
20161—23
20170—021
20180—014
20190—025
20200—019
20210—013
20221—023
20230—019
20242—032
20251—034
20261—1—

What is it indicated for?

The Tru Kidney Health Test Panel is part of the TruWellness Panel™ and is intended for use on the Tru Analyzer. The Tru Kidney Health Test Panel (part of the TruWellness Panel™) is an in vitro diagnostic device and intended to be used for the quantitative determination of Creatinine (CRE) and Blood Urea Nitrogen (BUN) in lithium-heparinized venous whole blood in clinical laboratory or point-of-care settings. From the CRE determination, Estimated…

Read the full statement (194 words)

The Tru Kidney Health Test Panel is part of the TruWellness Panel™ and is intended for use on the Tru Analyzer. The Tru Kidney Health Test Panel (part of the TruWellness Panel™) is an in vitro diagnostic device and intended to be used for the quantitative determination of Creatinine (CRE) and Blood Urea Nitrogen (BUN) in lithium-heparinized venous whole blood in clinical laboratory or point-of-care settings. From the CRE determination, Estimated Glomerular Filtration Rate (eGFR) is calculated by the analyzer.

The Tru Kidney Health Test Panel (part of the TruWellness Panel™) is an in vitro diagnostic test system that aids the physician in diagnosing the following disorders in adults 18 years of age or older:

Creatinine (CRE): Renal disease and monitoring of renal dialysis.

Blood Urea Nitrogen (BUN): Certain renal and metabolic diseases.

The Tru Analyzer is an automated, multi-assay integrated system, containing an absorbance reader, a bead scanner, and a cell imager for in vitro diagnostic use. This analyzer is intended for the quantitative determination of various analyte concentrations found in lithium-heparinized venous whole blood. It is for clinical laboratory or point-of-care use.

Use only TruWellness Panel™ test kits with the Tru Analyzer.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K251058: Tru Kidney Health Test Panel; Tru Analyzer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K251058

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026