The Acumed Wrist Fixation System

K251296510(k) TraditionalCleared Jun 18, 202554-day reviewClass II21 CFR 888.3030Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Acumed, LLC · HRS Plate, Fixation, Bone

Cleared on Jun 18, 2025 as substantially equivalent to K220757 (Conmed Corporation, 2022) after a 54-day review. Its family tree reaches back 11 generations to 1991. It has been recalled once.

Decided
Jun 18, 2025
Review
54 days
Class
Class II
Product code
HRS
Recalls
1
Predicates
1

Its longest line of descent runs back 11 generations to K911837 (1991). 71 ancestors in all, the oldest K792022 (1979). No later device cites it as a predicate yet.

  1. Root · 11 generations backK911837 Suretac Polyglyconate Absorbable Fixation DeviceAcufex Microsurgical, Inc. · 1991
  2. 7 generations between
  3. 3 generations backK091549 Conmed Linvatec Soft Tissue to Bone SystemLinvatec Corp. · 2009 · Recalled
  4. 2 generations backK163258 CrossFT Knotless Suture Anchor with Disposable DriverConmed Corporation · 2017
  5. 1 generation backK220757 Argo Knotless SP AnchorConmed Corporation · 2022
  6. This deviceK251296 The Acumed Wrist Fixation SystemAcumed, LLC · 2025
Show all 11 generations
  1. Root · 11 generations backK911837 Suretac Polyglyconate Absorbable Fixation DeviceAcufex Microsurgical, Inc. · 1991
  2. 10 generations backK973849 Bioabsorbable Bankart TackHogan & Hartson · 1998
  3. 9 generations backK992567 Anatomical Bankart Tack, Model 523520ABionx Implants, Ltd. · 1999
  4. 8 generations backK003970 Contour Labral Nail, Model 533520ABionx Implants, Inc. · 2001
  5. 7 generations backK020056 Duet Suture AnchorBionx Implants, Ltd. · 2002
  6. 6 generations backK030388 Impact Suture Anchor. Model 433510Bionx Implants, Ltd. · 2003
  7. 5 generations backK042966 Duet Suture Anchor, Models BXS4216, BXS4216H; Impact Suture Anchor…Linvatec Biomaterials, Ltd. · 2004
  8. 4 generations backK053561 Bio-Mini Revo Suture AnchorConmed Linvatec · 2006
  9. 3 generations backK091549 Conmed Linvatec Soft Tissue to Bone SystemLinvatec Corp. · 2009 · Recalled
  10. 2 generations backK163258 CrossFT Knotless Suture Anchor with Disposable DriverConmed Corporation · 2017
  11. 1 generation backK220757 Argo Knotless SP AnchorConmed Corporation · 2022
  12. This deviceK251296 The Acumed Wrist Fixation SystemAcumed, LLC · 2025
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−11K911837Suretac Polyglyconate Absorbable Fixation DeviceAcufex Microsurgical, Inc.1991—
−10K973849Bioabsorbable Bankart TackHogan & Hartson1998—
−9K992567Anatomical Bankart Tack, Model 523520ABionx Implants, Ltd.1999—
−8K003970Contour Labral Nail, Model 533520ABionx Implants, Inc.2001—
−7K020056Duet Suture AnchorBionx Implants, Ltd.2002—
−6K030388Impact Suture Anchor. Model 433510Bionx Implants, Ltd.2003—
−5K042966Duet Suture Anchor, Models BXS4216, BXS4216H; Impact Suture Anchor, Models…Linvatec Biomaterials, Ltd.2004—
−4K053561Bio-Mini Revo Suture AnchorConmed Linvatec2006—
−3K091549Conmed Linvatec Soft Tissue to Bone SystemLinvatec Corp.20091
−2K163258CrossFT Knotless Suture Anchor with Disposable DriverConmed Corporation2017—
−1K220757Argo Knotless SP AnchorConmed Corporation2022—
0 (this device)K251296The Acumed Wrist Fixation SystemAcumed, LLC2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
197910
199010
199230
199310
199420
199580
1996120
1997100
199860
199950
200030
200130
200220
200410
200510
200810
FIG. 1 — Lineage of K251296, 1991–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in HRS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HRS device

  1. #1 K222465 Tyber Medical Anatomical Plating Systemscore 27.5 · 16 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K220757Argo Knotless SP Anchor
Our reading of the summary gave “K220757” beside “510(k)”; FDA’s K220757 is the only number that fits it and carries the summary’s name for the device.
Summary page 5
…PRIMARY PREDICATE/ LEGALLY MARKETED DEVICE Device Name: Argo KnotlessTM SP Anchor Company Name: CONMED Corporation 510(k) #: K220757 REFERENCE DEVICE/ LEGALLY MARKETED DEVICE Device Name: Argo Knotless® GENESYSTM Anchor Company…
Conmed Corporation2022—

Also cleared against K220757: 3 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 28, 2026Class IIZ-2532-2026Device DesignDue to manufacturing error, hexalobe screws used with wrist fixation system may break during use.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 34 years and passes through 9 companies; a new generation was cleared every 1.6 years on median.
  • 5 of the 71 ancestors have a recall of any class.
  • 3 other devices also cite its predicate K220757.

How does its category compare?

All HRS devices

HRS Plate, Fixation, Bone: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.

FDA decisions per year · HRS

54in 2025

2007: 352008: 352009: 472010: 492011: 432012: 592013: 562014: 762015: 562016: 442017: 572018: 402019: 462020: 482021: 402022: 472023: 402024: 542025: 542026 (partial year): 51
2007high 76 (2014) · low 35 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HRS

90in 2025

2007: 63 days median2008: 87 days median2009: 85 days median2010: 100 days median2011: 88 days median2012: 106 days median2013: 89 days median2014: 57 days median2015: 97 days median2016: 109 days median2017: 122 days median2018: 116 days median2019: 89 days median2020: 88 days median2021: 87 days median2022: 144 days median2023: 120 days median2024: 96 days median2025: 90 days median2026 (partial year): 82 days median
2007high 144 (2022) · low 57 (2014)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HRS

1in 2025

2007: 22008: 112009: 32010: 162011: 52012: 42013: 112015: 72016: 52017: 32018: 22019: 252020: 12021: 92022: 32023: 12025: 12026 (partial year): 3
2007high 25 (2019) · low 0 (2014)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · HRS

2,306in 2025

2007: 3392008: 3112009: 4352010: 4912011: 7732012: 1,6632013: 2,6432014: 2,4132015: 3,0952016: 2,2762017: 2,2202018: 2,6632019: 3,3592020: 2,7902021: 2,4422022: 1,7002023: 2,0492024: 1,8492025: 2,306
2007high 3,359 (2019) · low 311 (2008)2025

MAUDE reports naming product code HRS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735632339
2008358711311
200947853435
20104910016491
201143885773
20125910641,663
20135689112,643
2014765702,413
2015569773,095
20164410952,276
20175712232,220
20184011622,663
20194689253,359
2020488812,790
2021408792,442
20224714431,700
20234012012,049
2024549601,849
2025549012,306
202651823—

What is it indicated for?

The Indications for Use for The Acumed Wrist Fixation System is to provide fixation of fractures, fusions, osteotomies, and nonunions of the distal radius and ulna.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K251296: The Acumed Wrist Fixation System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K251296

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026