510(k) Report

510(k) · Cleared

Cleared

Resolve Anterior Cervical Plate System

K251436510(k) AbbreviatedCleared Jul 11, 202564-day reviewClass II21 CFR 888.3060

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Pioneer Surgical Technology, Inc. · KWQ Appliance, Fixation, Spinal Intervertebral Body

Cleared on Jul 11, 2025 as substantially equivalent to K211408 (Pioneer Surgical Technology, Inc., 2021) after a 64-day review. Its family tree reaches back 11 generations to 1992.

Decided
Jul 11, 2025
Review
64 days
Class
Class II
Product code
KWQ
Recalls
None
Predicates
1

Its longest line of descent runs back 11 generations to K922087 (1992). 36 ancestors in all, the oldest K831246 (1983). No later device cites it as a predicate yet.

  1. Root · 11 generations backK922087 Anterior Cervical Plate SystemSofamor Danek Mfg., Inc. · 1992
  2. 7 generations between
  3. 3 generations backK130427 Pioneer Aspect Anterior Cervical Plate SystemPioneer Surgical Technology, Inc. · 2013
  4. 2 generations backK183060 CervAlign™ Anterior Cervical Plate SystemPioneer Surgical Technology, Inc. (Dba Rti Surgical, Inc.) · 2019 · Recalled
  5. 1 generation backK211408 CervAlign® Anterior Cervical Plate SystemPioneer Surgical Technology, Inc. · 2021
  6. This deviceK251436 Resolve Anterior Cervical Plate SystemPioneer Surgical Technology, Inc. · 2025
Show all 11 generations
  1. Root · 11 generations backK922087 Anterior Cervical Plate SystemSofamor Danek Mfg., Inc. · 1992
  2. 10 generations backK990005 Spinal Concepts, Inc. (Sci) Anterior Cervical Plate SystemSpinal Concepts, Inc. · 1999 · Recalled
  3. 9 generations backK000486 Abc Cervical Plating SystemAesculap, Inc. · 2000
  4. 8 generations backK043066 Pioneer Anterior Cervical Plate SystemPioneer Surgical Technology · 2005
  5. 7 generations backK053053 Pacp SystemPioneer Surgical Technology · 2005
  6. 6 generations backK072703 Pioneer Slimfuse Anterior Cervical Plating SystemPioneer Surgical Technology · 2008
  7. 5 generations backK100708 Pioneer Anterior Cervical Plate Model 24-SeriesPioneer Surgical Technology · 2010
  8. 4 generations backK111528 Pioneer Aspect Anterior Cervical Plate Model 24-SeriesPioneer Surgical Technology, Inc. · 2011
  9. 3 generations backK130427 Pioneer Aspect Anterior Cervical Plate SystemPioneer Surgical Technology, Inc. · 2013
  10. 2 generations backK183060 CervAlign™ Anterior Cervical Plate SystemPioneer Surgical Technology, Inc. (Dba Rti Surgical, Inc.) · 2019 · Recalled
  11. 1 generation backK211408 CervAlign® Anterior Cervical Plate SystemPioneer Surgical Technology, Inc. · 2021
  12. This deviceK251436 Resolve Anterior Cervical Plate SystemPioneer Surgical Technology, Inc. · 2025
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−11K922087Anterior Cervical Plate SystemSofamor Danek Mfg., Inc.1992—
−10K990005Spinal Concepts, Inc. (Sci) Anterior Cervical Plate SystemSpinal Concepts, Inc.19991
−9K000486Abc Cervical Plating SystemAesculap, Inc.2000—
−8K043066Pioneer Anterior Cervical Plate SystemPioneer Surgical Technology2005—
−7K053053Pacp SystemPioneer Surgical Technology2005—
−6K072703Pioneer Slimfuse Anterior Cervical Plating SystemPioneer Surgical Technology2008—
−5K100708Pioneer Anterior Cervical Plate Model 24-SeriesPioneer Surgical Technology2010—
−4K111528Pioneer Aspect Anterior Cervical Plate Model 24-SeriesPioneer Surgical Technology, Inc.2011—
−3K130427Pioneer Aspect Anterior Cervical Plate SystemPioneer Surgical Technology, Inc.2013—
−2K183060CervAlign™ Anterior Cervical Plate SystemPioneer Surgical Technology, Inc. (Dba Rti Surgical, Inc.)20191
−1K211408CervAlign® Anterior Cervical Plate SystemPioneer Surgical Technology, Inc.2021—
0 (this device)K251436Resolve Anterior Cervical Plate SystemPioneer Surgical Technology, Inc.2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198310
199210
199310
199520
199740
199840
199930
200010
200110
200440
200510
200910
201110
FIG. 1 — Lineage of K251436, 1992–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KWQ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KWQ device

  1. #1 K233640 Segmental Plating System (SPS);IdentiTi SPS…score 21.9 · 31 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K211408CervAlign® Anterior Cervical Plate System
Summary page 5
…tebral Body 888.3060 KWQ 21 CFR 807.92(a)(3) Product Code | Legally Marketed Predicate Devices Predicate # Predicate Trade Name (Primary Predicate is listed first) K211408 KWQ 21 CFR 807.92(a)(4) CervAlign Anterior Cervical Plate System Device Description Summary The Resolve Anterior Cervical Pla…
Pioneer Surgical Technology, Inc.2021—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 33 years and passes through 6 companies; a new generation was cleared every 2.4 years on median.
  • 6 of the 36 ancestors have a recall of any class.

How does its category compare?

All KWQ devices

KWQ Appliance, Fixation, Spinal Intervertebral Body: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.

FDA decisions per year · KWQ

7in 2025

2007: 272008: 322009: 232010: 232011: 222012: 262013: 222014: 172015: 252016: 152017: 202018: 222019: 262020: 262021: 172022: 182023: 162024: 172025: 72026 (partial year): 10
2007high 32 (2008) · low 7 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KWQ

64in 2025

2007: 101 days median2008: 70 days median2009: 90 days median2010: 104 days median2011: 95 days median2012: 90 days median2013: 88 days median2014: 124 days median2015: 102 days median2016: 121 days median2017: 67 days median2018: 95 days median2019: 81 days median2020: 55 days median2021: 59 days median2022: 85 days median2023: 59 days median2024: 64 days median2025: 64 days median2026 (partial year): 76 days median
2007high 124 (2014) · low 55 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KWQ

1in 2018

2007: 22008: 332009: 352010: 22011: 52012: 72013: 62015: 22016: 92017: 82018: 1
2007high 35 (2009) · low 0 (2014)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KWQ

248in 2025

2007: 2882008: 3482009: 3552010: 6912011: 4152012: 5442013: 7962014: 5462015: 5622016: 5022017: 4922018: 3412019: 2982020: 4002021: 5912022: 3922023: 3202024: 4192025: 248
2007high 796 (2013) · low 248 (2025)2025

MAUDE reports naming product code KWQ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007271012288
2008327033348
2009239035355
2010231042691
201122955415
201226907544
201322886796
2014171240546
2015251022562
2016151219502
201720678492
201822951341
201926810298
202026550400
202117590591
202218850392
202316590320
202417640419
20257640248
202610760—

What is it indicated for?

The Resolve Anterior Cervical Plate System is intended for anterior cervical fixation of the cervical spine (C2-C7) as an adjunct to fusion in skeletally mature patients. Specific indications include: degenerative disc disease (ODD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K251436: Resolve Anterior Cervical Plate System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K251436

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026