510(k) · Cleared
ClearedVITEK 2 AST-Gram Negative Cefazolin (=<1-=>32 µg/mL)
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
bioMerieux, Inc. · LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation
Cleared on Aug 21, 2025 as substantially equivalent to K222073 (bioMerieux, Inc., 2023) after a 90-day review. Its family tree reaches back 4 generations to 2019.
- Decided
- Aug 21, 2025
- Review
- 90 days
- Class
- Class II
- Product code
- LON
- Recalls
- None
- Predicates
- 1
Its longest line of descent runs back 4 generations to K183524 (2019). No later device cites it as a predicate yet.
- Root · 4 generations backK183524 VITEK2 AST-Gram Negative Delafloxacin (<=0.06->=4 ug/mL)bioMerieux, Inc. · 2019
- 3 generations backK191766 VITEK 2 AST-Gram Negative Eravacycline (<=0.12 - >=4µg/mL)bioMerieux, Inc. · 2019
- 2 generations backK213931 VITEK 2 AST-Gram Negative Omadacycline (<=0.25 - >=16 µg/mL)bioMerieux, Inc. · 2022
- 1 generation backK222073 VITEK 2 AST-Gram Negative Cefazolin (<=1 - >=32 ug/mL)bioMerieux, Inc. · 2023
- This deviceK251579 VITEK 2 AST-Gram Negative Cefazolin (=<1-=>32 µg/mL)bioMerieux, Inc. · 2025
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −4 | K183524 | VITEK2 AST-Gram Negative Delafloxacin (<=0.06->=4 ug/mL) | bioMerieux, Inc. | 2019 | — |
| −3 | K191766 | VITEK 2 AST-Gram Negative Eravacycline (<=0.12 - >=4µg/mL) | bioMerieux, Inc. | 2019 | — |
| −2 | K213931 | VITEK 2 AST-Gram Negative Omadacycline (<=0.25 - >=16 µg/mL) | bioMerieux, Inc. | 2022 | — |
| −1 | K222073 | VITEK 2 AST-Gram Negative Cefazolin (<=1 - >=32 ug/mL) | bioMerieux, Inc. | 2023 | — |
| 0 (this device) | K251579 | VITEK 2 AST-Gram Negative Cefazolin (=<1-=>32 µg/mL) | bioMerieux, Inc. | 2025 | — |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in LON? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new LON device
- #1 K260619 VITEK 2 AST-Streptococcus Cefepime (<=0.06 –…score 13.3 · 5 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K222073 | VITEK 2 AST-Gram Negative Cefazolin (<=1 - >=32 ug/mL)Summary page 5…Code LON Common Name: VITEK® 2 AST-GN Cefazolin C. Predicate Device: VITEK® 2 AST-Gram Negative Cefazolin (K222073) D. Device Description: The principle of the VITEK® 2 AST cards is based on the microdilution minimum inhibitory concentration… | 2023 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 6 years and passes through 1 company; a new generation was cleared every 2.5 years on median.
- None of the 4 ancestors has a recall.
How does its category compare?
All LON devicesLON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.
FDA decisions per year · LON
5in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · LON
90in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · LON
1in 2025
Class I (2 since 2007) Class II/III not classified
Adverse event reports per year · LON
690in 2025
MAUDE reports naming product code LON, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 19 | 50 | 0 | 3 |
| 2008 | 7 | 66 | 1 | 1 |
| 2009 | 8 | 76 | 0 | 0 |
| 2010 | 1 | — | 116 | 3 |
| 2011 | 8 | 85 | 3 | 4 |
| 2012 | 10 | 200 | 3 | 9 |
| 2013 | 7 | 115 | 0 | 3 |
| 2014 | 4 | 90 | 2 (2) | 10 |
| 2015 | 0 | — | 0 | 41 |
| 2016 | 3 | 242 | 2 | 94 |
| 2017 | 10 | 89 | 1 | 197 |
| 2018 | 4 | 88 | 6 | 512 |
| 2019 | 8 | 84 | 1 | 315 |
| 2020 | 5 | 81 | 8 | 218 |
| 2021 | 5 | 268 | 0 | 232 |
| 2022 | 8 | 189 | 15 | 180 |
| 2023 | 11 | 210 | 0 | 453 |
| 2024 | 2 | — | 1 | 418 |
| 2025 | 5 | 90 | 1 | 690 |
| 2026 | 4 | 146 | 6 | — |
What is it indicated for?
VITEK 2 AST-Gram Negative Cefazolin is designed for antimicrobial susceptibility testing of Gram-negative bacilli and is intended for use with the VITEK 2 Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK 2 AST-Gram Negative Cefazolin (≤1-≥32 µg/mL) is a quantitative test. Testing is indicated for Enterobacterales (from infections other than uncomplicated UTI) as recognized by the FDA…
Read the full statement (137 words)
VITEK 2 AST-Gram Negative Cefazolin is designed for antimicrobial susceptibility testing of Gram-negative bacilli and is intended for use with the VITEK 2 Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents.
VITEK 2 AST-Gram Negative Cefazolin (≤1-≥32 µg/mL) is a quantitative test. Testing is indicated for Enterobacterales (from infections other than uncomplicated UTI) as recognized by the FDA Susceptibility Test Interpretive Criteria (STIC).
VITEK 2 AST-Gram Negative Cefazolin (≤1-≥32 µg/mL) has demonstrated acceptable performance with the following organisms:
Enterobacterales (Escherichia coli, Proteus mirabilis, Klebsiella pneumoniae, Klebsiella oxytoca, Citrobacter koseri)
The VITEK 2 Gram Negative Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant aerobic Gram-negative bacilli to antimicrobial agents when used as instructed.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K251579 · printed Nov 5, 2026
Cite this page
510(k) Report. “K251579: VITEK 2 AST-Gram Negative Cefazolin (=<1-=>32 µg/mL).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K251579
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026