510(k) Report

510(k) · Cleared

Cleared

Ceribell Delirium Monitor System

K251936510(k) TraditionalCleared Dec 8, 2025167-day reviewClass II21 CFR 882.1440

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ceribell, Inc. · NCG Neuropsychiatric Interpretative Electroencephalograph Assessment Aid

Cleared on Dec 8, 2025 as substantially equivalent to K222680 (Prolira B.V., 2023) and 2 other predicates after a 167-day review. Its family tree reaches back 7 generations to 1978.

Decided
Dec 8, 2025
Review
167 days
Class
Class II
Product code
NCG
Recalls
None
Predicates
3

Its longest line of descent runs back 7 generations to K780045 (1978). 16 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 7 generations backK780045 Electrode CapElectro-Cap Intl., Inc. · 1978
  2. 6 generations backK071446 Electrode Array CapElectrode Arrays · 2008
  3. 5 generations backK110223 Waveguard (TM) Eeg CapEemagine Medical Imaging Solutions GmbH · 2011
  4. 4 generations backK112319 Electro-Cap SystemElectro-Cap Intl., Inc. · 2012
  5. 3 generations backK200162 Disposable EEG Electrodes (MODEL: DL, E-CAP, FLEX-CAP)Wuhan Greentek Pty, Ltd. · 2020
  6. 2 generations backK210805 Ceribell Instant EEG HeadbandCeribell, Inc. · 2021
  7. 1 generation backK232052 Ceribell Instant EEG HeadbandCeribell, Inc. · 2023
  8. This deviceK251936 Ceribell Delirium Monitor SystemCeribell, Inc. · 2025
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K780045Electrode CapElectro-Cap Intl., Inc.1978—
−6K071446Electrode Array CapElectrode Arrays2008—
−5K110223Waveguard (TM) Eeg CapEemagine Medical Imaging Solutions GmbH2011—
−4K112319Electro-Cap SystemElectro-Cap Intl., Inc.2012—
−3K200162Disposable EEG Electrodes (MODEL: DL, E-CAP, FLEX-CAP)Wuhan Greentek Pty, Ltd.2020—
−2K210805Ceribell Instant EEG HeadbandCeribell, Inc.2021—
−1K232052Ceribell Instant EEG HeadbandCeribell, Inc.2023—
0 (this device)K251936Ceribell Delirium Monitor SystemCeribell, Inc.2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198310
200010
201010
201310
201510
201720
201910
202310
FIG. 1 — Lineage of K251936, 1978–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in NCG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NCG device

  1. #1 K222680 DeltaScan Monitorscore 7.8 · 1 generation · no Class I recall
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K222680DeltaScan Monitor
Summary page 5
…Name: Neuropsychiatric Interpretive Electroencephalograph Assessment Aid Device Class: II Product Code: NCG 3. Predicate Device: Primary Predicate: K222680: DeltaScan Monitor Secondary Predicates: K191301: Ceribell Pocket EEG Device K232052: Ceribell Instant EEG Headband…
Prolira B.V.2023—
*K191301Ceribell Pocket EEG Device
Summary page 5
…halograph Assessment Aid Device Class: II Product Code: NCG 3. Predicate Device: Primary Predicate: K222680: DeltaScan Monitor Secondary Predicates: K191301: Ceribell Pocket EEG Device K232052: Ceribell Instant EEG Headband…
Ceribell, Inc.2019—
*K232052Ceribell Instant EEG Headband
Summary page 5
…s: II Product Code: NCG 3. Predicate Device: Primary Predicate: K222680: DeltaScan Monitor Secondary Predicates: K191301: Ceribell Pocket EEG Device K232052: Ceribell Instant EEG Headband…
Ceribell, Inc.2023—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 47 years and passes through 5 companies; a new generation was cleared every 2.3 years on median.
  • None of the 16 ancestors has a recall.

How does its category compare?

All NCG devices

NCG Neuropsychiatric Interpretative Electroencephalograph Assessment Aid: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.

FDA decisions per year · NCG

1in 2025

2013: 12023: 12025: 1
2007high 1 (2013) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NCG

Too few NCG 510(k)s in any year since 2007 to give a median.

Recalls per year · NCG

No recall names NCG since 2007.

Adverse event reports per year · NCG

No MAUDE report names NCG since 2007.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20131—00
20140—00
20150—00
20160—00
20170—00
20180—00
20190—00
20200—00
20210—00
20220—00
20231—00
20240—00
20251—00
20260—0—

What is it indicated for?

The Ceribell Delirium Monitor System is intended to analyze features associated with diffuse slowing electroencephalogram (EEG) patterns that may be indicative of delirium. The Ceribell Delirium Monitor software is intended to aid in the screening and monitoring of delirium with clinical assessments in adult patients in critical care settings within hospitals.

The Ceribell Delirium Monitor System analyzes discrete segments of EEG to notify clinicians when EEG patterns associated with delirium are detected while monitoring the patient. Changes in patient condition that are detected by the device should be verified before commencing any interventions.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K251936: Ceribell Delirium Monitor System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K251936

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026