510(k) Report

510(k) · Cleared

Cleared

Healgen® AccuFluor Fentanyl Fluorescence Immunoassay (FIA)Test Kit - Qualitative; Healgen® Immunofluorescence Analyzer (OG-H180)

K251972510(k) TraditionalCleared Aug 15, 202550-day reviewClass II21 CFR 862.3650

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Healgen Scientific,, LLC · DJG Enzyme Immunoassay, Opiates

Cleared on Aug 15, 2025 as substantially equivalent to K240124 (Vivachek Biotech (Hangzhou) Co., Ltd., 2024) after a 50-day review. Its family tree reaches back 7 generations to 2000.

Decided
Aug 15, 2025
Review
50 days
Class
Class II
Product code
DJG
Recalls
None
Predicates
1

Its longest line of descent runs back 7 generations to K001178 (2000). 8 ancestors in all, the oldest K973547 (1998). No later device cites it as a predicate yet.

  1. Root · 7 generations backK001178 Cedia Dau 6-Acetylmorphine AssayMicrogenics Corp. · 2000
  2. 6 generations backK040316 Cedia Buprenorphine AssayMicrogenics Corp. · 2004
  3. 5 generations backK150606 Emit II Plus Buprenorphine Assay, Emit II Plus Specialty Drug…Siemens Healthcare Diagnostics, Inc. · 2015
  4. 4 generations backK161216 Immunalysis SEFRIA Fentanyl Urine Enzyme Immunoassay, Immunalysis…Immunalysis Corporation · 2017
  5. 3 generations backK180427 ARK Fentanyl AssayArk Diagnostics, Inc. · 2018
  6. 2 generations backK220046 Superbio Fentanyl Urine Detection Kit, Superbio Immunofluorescence…Shenzhen Superbio Technology Co., Ltd. · 2022
  7. 1 generation backK240124 BioSieve™ Fentanyl FIA Test Kit; BioSieve™ ToxiSmart FIA ReaderVivachek Biotech (Hangzhou) Co., Ltd. · 2024
  8. This deviceK251972 Healgen® AccuFluor Fentanyl Fluorescence Immunoassay (FIA)Test Kit -…Healgen Scientific,, LLC · 2025
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K001178Cedia Dau 6-Acetylmorphine AssayMicrogenics Corp.2000—
−6K040316Cedia Buprenorphine AssayMicrogenics Corp.2004—
−5K150606Emit II Plus Buprenorphine Assay, Emit II Plus Specialty Drug…Siemens Healthcare Diagnostics, Inc.2015—
−4K161216Immunalysis SEFRIA Fentanyl Urine Enzyme Immunoassay, Immunalysis Fentanyl…Immunalysis Corporation2017—
−3K180427ARK Fentanyl AssayArk Diagnostics, Inc.2018—
−2K220046Superbio Fentanyl Urine Detection Kit, Superbio Immunofluorescence Analyzer…Shenzhen Superbio Technology Co., Ltd.2022—
−1K240124BioSieve™ Fentanyl FIA Test Kit; BioSieve™ ToxiSmart FIA ReaderVivachek Biotech (Hangzhou) Co., Ltd.2024—
0 (this device)K251972Healgen® AccuFluor Fentanyl Fluorescence Immunoassay (FIA)Test Kit -…Healgen Scientific,, LLC2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199810
FIG. 1 — Lineage of K251972, 2000–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DJG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DJG device

  1. #1 K232659 Dochek® Multi-Drug Urine Test Cup Rx, Dochek®…score 17.6 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K240124BioSieve™ Fentanyl FIA Test Kit; BioSieve™ ToxiSmart FIA Reader
Summary page 5
….2560 Clinical Chemistry Fluorometer for clinical use 5. Predicate Devices: BioSieve™ Fentanyl FIA Test Kit, K240124 BioSieve™ ToxiSmart FIA Reader, K240124 6. Indications for Use Healgen AccuFluor Fentanyl Fluorescence Immunoassay (FIA) T…
Vivachek Biotech (Hangzhou) Co., Ltd.2024—

Also cleared against K240124: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 25 years and passes through 7 companies; a new generation was cleared every 1.6 years on median.
  • None of the 8 ancestors has a recall.
  • 1 other device also cites its predicate K240124.

How does its category compare?

All DJG devices

DJG Enzyme Immunoassay, Opiates: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.

FDA decisions per year · DJG

3in 2025

2007: 22008: 22009: 22010: 62011: 62012: 52013: 52014: 92015: 122016: 22017: 42018: 52019: 32020: 52021: 32022: 22023: 102024: 52025: 32026 (partial year): 3
2007high 12 (2015) · low 2 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DJG

31in 2025

2010: 187 days median2011: 233 days median2012: 95 days median2013: 60 days median2014: 62 days median2015: 141 days median2017: 101 days median2018: 134 days median2019: 235 days median2020: 30 days median2021: 422 days median2023: 140 days median2024: 150 days median2025: 31 days median2026 (partial year): 157 days median
2007high 422 (2021) · low 30 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DJG

2in 2025

2011: 12019: 12025: 2
2007high 2 (2025) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DJG

1in 2025

2011: 42012: 12013: 42015: 32016: 22017: 12018: 12021: 22022: 12024: 32025: 1
2007high 4 (2011) · low 0 (2007)2025

MAUDE reports naming product code DJG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—00
20082—00
20092—00
2010618700
2011623314
201259501
201356004
201496200
20151214103
20162—02
2017410101
2018513401
2019323510
202053000
2021342202
20222—01
20231014000
2024515003
202533121
202631570—

What is it indicated for?

The Healgen® AccuFluor Fentanyl Fluorescence Immunoassay (FIA) Test Kit-Qualitative is a fluorescence immunoassay intended for the qualitative detection of fentanyl in human urine at a cutoff concentration of 1.0 ng/mL. The assay is intended for use with Healgen® Immunofluorescence analyzer OG-H180. This in vitro diagnostic device is for prescription use only. This assay provides only a preliminary analytical test result. A more specific alternate…

Read the full statement (153 words)

The Healgen® AccuFluor Fentanyl Fluorescence Immunoassay (FIA) Test Kit-Qualitative is a fluorescence immunoassay intended for the qualitative detection of fentanyl in human urine at a cutoff concentration of 1.0 ng/mL. The assay is intended for use with Healgen® Immunofluorescence analyzer OG-H180. This in vitro diagnostic device is for prescription use only.

This assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used to obtain a confirmed analytical result. Gas Chromatography-Mass Spectrometry (GC-MS) and Liquid Chromatography-Mass Spectrometry (LC-MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to a Fentanyl test result, particularly when a preliminary positive result is obtained.

The Healgen® Immunofluorescence analyzer OG-H180 is a portable fluorescence instrument for in vitro diagnostic use only. The analyzer is designed to detect test results from in vitro diagnostic tests on clinical specimens. This analyzer can be used in a laboratory or point-of-care setting.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K251972: Healgen® AccuFluor Fentanyl Fluorescence Immunoassay (FIA)Test Kit - Qualitative; Healgen® Immunofluorescence Analyzer (OG-H180).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K251972

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026