# K252842: SafeOp 3: Neural Informatix System

- Pathway: 510(k) (Traditional)
- Applicant: Alphatec Spine, Inc.
- Decided: Jan 11, 2026 after 125 days
- Product code: GWF (Stimulator, Electrical, Evoked Response)
- Device class: 2
- Regulation: 21 CFR 882.1870
- Recalls of this device: 1 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 9 generations back to K842956 (1984). 62 ancestors in all.
No later device cites it as a predicate yet.
0 of 63 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K234092: SafeOp 3: Neural Informatix Systeem, Alphatec Spine, Inc. (2024) Summary page 5.

## Recalls of this device
- May 13, 2026 · Class II · Z-2793-2026: Intraoperative neuromonitoring device, due to a software issue, may experience loss or interruption of triggered electromyography (tEMG) functionality during use, displaying as system faults, loss of 

## Indications for use
The SafeOp 3: Neural Informatix System is intended for use in monitoring neurological status by recording transcranial motor evoked potentials (MEP), somatosensory evoked potentials (SSEP), electromyography (EMG), or assessing the neuromuscular junction (NMJ). Neuromonitoring procedures include intracranial, extracranial, intratemporal, extratemporal, neck dissections, upper and lower extremities, spinal degenerative treatments, pedicle screw fixation, intervertebral fusion cages, rhizotomy, orthopedic surgery, open/percutaneous, lumbar, thoracic, and cervical surgical procedures. SafeOp 3 Accessories: The SafeOp Accessories are utilized in spine surgical procedures to assist in location of the nerves during or after preparation and placement of implants (intervertebral fusion cages and pedicle screw fixation devices) in open and percutaneous minimally invasive approaches.