510(k) Report

510(k) · Cleared

Cleared

Single Use Digital Flexible Ureteroscope (7.5F); Videoscope Imaging Processor (Model HV-200)

K252929510(k) TraditionalCleared Apr 9, 2026206-day reviewClass II21 CFR 876.1500

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Anhui Happiness Workshop Instruments Co., Ltd. · FGB Ureteroscope And Accessories, Flexible/Rigid

Cleared on Apr 9, 2026 as substantially equivalent to K233778 (Zhuhai Pusen Medical Technology Co., Ltd., 2024) after a 206-day review. Its family tree reaches back 10 generations to 1993. 1 of the 27 devices in its family, counting itself, has a Class I recall.

Decided
Apr 9, 2026
Review
206 days
Class
Class II
Product code
FGB
Recalls
None
Predicates
1

Its longest line of descent runs back 10 generations to K931154 (1993). 26 ancestors in all, the oldest K843084 (1985). No later device cites it as a predicate yet. 1 of the 27 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year cleared 2001: 4 other ancestors2003: 1 other ancestor1997: 2 other ancestors1995: 2 other ancestors1991: 2 other ancestors, 1 with a Class I recall12000: 1 other ancestor1993: 1 other ancestor1992: 1 other ancestor1985: 2 other ancestorsReference device: K222602 · Pusen Single Use Flexible Video Cystoscope System (Single Use Flexible Video Cystoscope: PC200-AS, PC200-AR, PC200-S and PC200-R; HD Medical Video Endoscope Image Processor: PV300) · Zhuhai Pusen Medical Technology Co., Ltd. · 2022K931154 · Evis 200 System · Olympus Corp. · 1993K963033 · BF 240/P240/IT240 Bronchovideoscope & Accessories · Olympus America, Inc. · 1996 · recalledK981543 · Olympus LF-TP and LF-DP Tracheal Intubation Fiberscopes, Accessories and Ancillary Equipment · The Olympus Optical Co. · 1998 · recalledK021074 · Visera Cystovideoscope Olympus Cyf Type V/VA · Olympus America, Inc. · 2002K030960 · Ecn Electronic Video Cystonephroscope · Acmi Corporation · 2003K042225 · Acmi Invisio Icn, Models Icn-0564 and Icn-0565 · Acmi Corporation · 2004K060269 · Acmi Dur-Digital Ureteroscope and Choledochoscope System · Acmi Corporation · 2006K131369 · Flexible Video-Uretero-Renoscope System · KARL STORZ Endoscopy-America, Inc. · 2013 · recalledK172098 · Medical Video Endoscope System · Zhuhai Pusen Medical Technology Co., Ltd. · 2017K233778 · Single Use Suction Access Ureteroscope (PU400A, PU411A), Single Use Flexible Video Ureteroscope (PU3033H, PU3033AH) · Zhuhai Pusen Medical Technology Co., Ltd. · 20241990200020102020
  1. Root · 10 generations backK931154 Evis 200 SystemOlympus Corp. · 1993
  2. 6 generations between
  3. 3 generations backK131369 Flexible Video-Uretero-Renoscope SystemKARL STORZ Endoscopy-America, Inc. · 2013 · Recalled
  4. 2 generations backK172098 Medical Video Endoscope SystemZhuhai Pusen Medical Technology Co., Ltd. · 2017
  5. 1 generation backK233778 Single Use Suction Access Ureteroscope (PU400A, PU411A), Single Use…Zhuhai Pusen Medical Technology Co., Ltd. · 2024
  6. This deviceK252929 Single Use Digital Flexible Ureteroscope (7.5F); Videoscope Imaging…Anhui Happiness Workshop Instruments Co., Ltd. · 2026
Show all 10 generations
  1. Root · 10 generations backK931154 Evis 200 SystemOlympus Corp. · 1993
  2. 9 generations backK963033 BF 240/P240/IT240 Bronchovideoscope & AccessoriesOlympus America, Inc. · 1996 · Recalled
  3. 8 generations backK981543 Olympus LF-TP and LF-DP Tracheal Intubation Fiberscopes, Accessories…The Olympus Optical Co. · 1998 · Recalled
  4. 7 generations backK021074 Visera Cystovideoscope Olympus Cyf Type V/VAOlympus America, Inc. · 2002
  5. 6 generations backK030960 Ecn Electronic Video CystonephroscopeAcmi Corporation · 2003
  6. 5 generations backK042225 Acmi Invisio Icn, Models Icn-0564 and Icn-0565Acmi Corporation · 2004
  7. 4 generations backK060269 Acmi Dur-Digital Ureteroscope and Choledochoscope SystemAcmi Corporation · 2006
  8. 3 generations backK131369 Flexible Video-Uretero-Renoscope SystemKARL STORZ Endoscopy-America, Inc. · 2013 · Recalled
  9. 2 generations backK172098 Medical Video Endoscope SystemZhuhai Pusen Medical Technology Co., Ltd. · 2017
  10. 1 generation backK233778 Single Use Suction Access Ureteroscope (PU400A, PU411A), Single Use…Zhuhai Pusen Medical Technology Co., Ltd. · 2024
  11. This deviceK252929 Single Use Digital Flexible Ureteroscope (7.5F); Videoscope Imaging…Anhui Happiness Workshop Instruments Co., Ltd. · 2026

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−10K931154Evis 200 SystemOlympus Corp.1993—
−9K963033BF 240/P240/IT240 Bronchovideoscope & AccessoriesOlympus America, Inc.19966
−8K981543Olympus LF-TP and LF-DP Tracheal Intubation Fiberscopes, Accessories and…The Olympus Optical Co.19985
−7K021074Visera Cystovideoscope Olympus Cyf Type V/VAOlympus America, Inc.2002—
−6K030960Ecn Electronic Video CystonephroscopeAcmi Corporation2003—
−5K042225Acmi Invisio Icn, Models Icn-0564 and Icn-0565Acmi Corporation2004—
−4K060269Acmi Dur-Digital Ureteroscope and Choledochoscope SystemAcmi Corporation2006—
−3K131369Flexible Video-Uretero-Renoscope SystemKARL STORZ Endoscopy-America, Inc.20131
−2K172098Medical Video Endoscope SystemZhuhai Pusen Medical Technology Co., Ltd.2017—
−1K233778Single Use Suction Access Ureteroscope (PU400A, PU411A), Single Use Flexible…Zhuhai Pusen Medical Technology Co., Ltd.2024—
0 (this device)K252929Single Use Digital Flexible Ureteroscope (7.5F); Videoscope Imaging Processor…Anhui Happiness Workshop Instruments Co., Ltd.2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198520
199121
199210
199310
199520
199720
200010
200140
200310
FIG. 1 — Lineage of K252929, 1993–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FGB device

  1. #1 K251951 Single-Use Video Flexible Ureterorenoscope…score 16.2 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K233778Single Use Suction Access Ureteroscope (PU400A, PU411A), Single Use Flexible Video…
Summary page 6
…reteroscope (Model: PU3033H, PU3033AH) and Pusen Single Use Suction Access Ureteroscope (Model: PU400A, PU411A) 510(k) number: K233778 Product code: FGB This predicate device has not been subject to a design-related recall. 4.0 Indication for Use Statement…
Zhuhai Pusen Medical Technology Co., Ltd.2024—
†K222602Pusen Single Use Flexible Video Cystoscope System (Single Use Flexible Video Cystoscope…Zhuhai Pusen Medical Technology Co., Ltd.2022—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (1 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K912120Urf-P2 Ureterorenofiberscope/CholedochofiberscopeOlympus Corp. · 1991 · 8 generations back · ◆ in the 1991 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 33 years and passes through 7 companies; a new generation was cleared every 2.9 years on median.
  • 8 of the 26 ancestors have a recall of any class; 1 has a Class I recall.

How does its category compare?

All FGB devices

FGB Ureteroscope And Accessories, Flexible/Rigid: FDA decisions, review times, recalls and adverse event reports over 20 years. 2026, this device’s year, is highlighted.

FDA decisions per year · FGB

17in 2025

2012: 22013: 12014: 22015: 12017: 22018: 42019: 22020: 22021: 32022: 52023: 82024: 162025: 172026 (partial year): 9
2007high 17 (2025) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FGB

134in 2025

2018: 191 days median2021: 191 days median2022: 205 days median2023: 183 days median2024: 217 days median2025: 134 days median2026 (partial year): 140 days median
2007high 217 (2024) · low 134 (2025)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FGB

15in 2024

2022: 42023: 12024: 152026 (partial year): 1
2007high 15 (2024) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FGB

1,307in 2025

2007: 62008: 72009: 42010: 202011: 172012: 112013: 82014: 122015: 162016: 352017: 1422018: 1182019: 2072020: 4022021: 3252022: 4552023: 4582024: 5722025: 1,307
2007high 1,307 (2025) · low 4 (2009)2025

MAUDE reports naming product code FGB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—06
20080—07
20090—04
20100—020
20110—017
20122—011
20131—08
20142—012
20151—016
20160—035
20172—0142
201841910118
20192—0207
20202—0402
202131910325
202252054455
202381831458
20241621715572
20251713401,307
202691401—

What is it indicated for?

Single Use Digital Flexible Ureteroscope (7.5F) is intended to visualize organs, cavities and canals in the urinary tract (urethra, bladder, ureter, calyces and renal papillae) via transurethral or percutaneous access routes. It can also be used in conjunction with endoscopic accessories to perform various diagnostic and therapeutic procedures in the urinary tract.

The Videoscope Imaging Processor, when used with the endoscope, provides imaging for the examination, diagnosis, or treatment of urinary tract conditions.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K252929: Single Use Digital Flexible Ureteroscope (7.5F); Videoscope Imaging Processor (Model HV-200).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K252929

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026