510(k) Report

510(k) · Cleared

Cleared

Korotkoff sound blood pressure monitor

K253982510(k) TraditionalCleared Apr 21, 2026130-day reviewClass II21 CFR 870.1130

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Beijing Hanvon Health Technology Co., Ltd. · DXN System, Measurement, Blood-Pressure, Non-Invasive

Cleared on Apr 21, 2026 as substantially equivalent to K251120 (Microlife Corporation, 2025) after a 130-day review. Its family tree reaches back 12 generations to 1991.

Decided
Apr 21, 2026
Review
130 days
Class
Class II
Product code
DXN
Recalls
None
Predicates
1

Its longest line of descent runs back 12 generations to K903134 (1991). 26 ancestors in all, the oldest K872291 (1987). No later device cites it as a predicate yet.

Other ancestors, by year cleared 2015: 1 other ancestor2006: 1 other ancestor2014: 1 other ancestor2002: 3 other ancestors1999: 1 other ancestor1998: 2 other ancestors1997: 1 other ancestor1993: 1 other ancestor1996: 1 other ancestor1995: 1 other ancestor1987: 1 other ancestorK903134 · Omron Auto Oscillometric Dig BP Mont/Hem-704C 705C · Omron Marshall Products, Inc. · 1991K970211 · Microlife Automatic Blood Pressure Monitor · Micro Weiss Electronics · 1998K013485 · Microlife Upper Arm Automatic Blood Pressure Monitor, Model BP-3BTO-1, with Optional Thermal Printer, Model PR 1KA1 · Microlife Corp. · 2002K041411 · Microlife Upper Arm Blood Pressure Monitor, Model BP-3BTO-AP · Microlife Corp. · 2004K060686 · Microlife Automatic Blood Pressure Monitor, Model BP3AC1-1PC · Microlife Intellectual Property GmbH · 2006K073198 · Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model Watchbp Home (BP3MX1-1) · Microlife Intellectual Property GmbH · 2008K082357 · Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Modelwatchbp Office (TWIN200). · Microlife Intellectual Property GmbH · 2008K151869 · Microlife Upper Arm Automatic Digital Blood Pressure Monitor · Microlife Intellectual Property GmbH, Switzerland · 2015K153077 · Microlife Upper Arm Automatic Digital Blood Pressure Monitor · Microlife Intellectual Property GmbH · 2016K172498 · Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3MW1-4Y · Microlife Intellectual Property GmbH · 2017K230075 · Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3KVl-5K · Microlife Intellectual Property GmbH · 2023K251120 · Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP Progress (BP3T01-1B) · Microlife Corporation · 20251990200020102020
Other ancestors, by generation 3 generations back: 1 other ancestor4 generations back: 1 other ancestor5 generations back: 2 other ancestors6 generations back: 1 other ancestor7 generations back: 1 other ancestor8 generations back: 5 other ancestors9 generations back: 3 other ancestors K251120 · Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP Progress (BP3T01-1B) · Microlife Corporation · 2025K230075 · Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3KVl-5K · Microlife Intellectual Property GmbH · 2023K172498 · Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3MW1-4Y · Microlife Intellectual Property GmbH · 2017K153077 · Microlife Upper Arm Automatic Digital Blood Pressure Monitor · Microlife Intellectual Property GmbH · 2016K151869 · Microlife Upper Arm Automatic Digital Blood Pressure Monitor · Microlife Intellectual Property GmbH, Switzerland · 2015K082357 · Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Modelwatchbp Office (TWIN200). · Microlife Intellectual Property GmbH · 2008K073198 · Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model Watchbp Home (BP3MX1-1) · Microlife Intellectual Property GmbH · 2008K060686 · Microlife Automatic Blood Pressure Monitor, Model BP3AC1-1PC · Microlife Intellectual Property GmbH · 2006K041411 · Microlife Upper Arm Blood Pressure Monitor, Model BP-3BTO-AP · Microlife Corp. · 2004K013485 · Microlife Upper Arm Automatic Blood Pressure Monitor, Model BP-3BTO-1, with Optional Thermal Printer, Model PR 1KA1 · Microlife Corp. · 2002K970211 · Microlife Automatic Blood Pressure Monitor · Micro Weiss Electronics · 1998K903134 · Omron Auto Oscillometric Dig BP Mont/Hem-704C 705C · Omron Marshall Products, Inc. · 1991−10−50
  1. Root · 12 generations backK903134 Omron Auto Oscillometric Dig BP Mont/Hem-704C 705COmron Marshall Products, Inc. · 1991
  2. 8 generations between
  3. 3 generations backK172498 Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model…Microlife Intellectual Property GmbH · 2017
  4. 2 generations backK230075 Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model…Microlife Intellectual Property GmbH · 2023
  5. 1 generation backK251120 Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model…Microlife Corporation · 2025
  6. This deviceK253982 Korotkoff sound blood pressure monitorBeijing Hanvon Health Technology Co., Ltd. · 2026
Show all 12 generations
  1. Root · 12 generations backK903134 Omron Auto Oscillometric Dig BP Mont/Hem-704C 705COmron Marshall Products, Inc. · 1991
  2. 11 generations backK970211 Microlife Automatic Blood Pressure MonitorMicro Weiss Electronics · 1998
  3. 10 generations backK013485 Microlife Upper Arm Automatic Blood Pressure Monitor, Model…Microlife Corp. · 2002
  4. 9 generations backK041411 Microlife Upper Arm Blood Pressure Monitor, Model BP-3BTO-APMicrolife Corp. · 2004
  5. 8 generations backK060686 Microlife Automatic Blood Pressure Monitor, Model BP3AC1-1PCMicrolife Intellectual Property GmbH · 2006
  6. 7 generations backK073198 Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model…Microlife Intellectual Property GmbH · 2008
  7. 6 generations backK082357 Microlife Upper Arm Automatic Digital Blood Pressure Monitor…Microlife Intellectual Property GmbH · 2008
  8. 5 generations backK151869 Microlife Upper Arm Automatic Digital Blood Pressure MonitorMicrolife Intellectual Property GmbH, Switzerland · 2015
  9. 4 generations backK153077 Microlife Upper Arm Automatic Digital Blood Pressure MonitorMicrolife Intellectual Property GmbH · 2016
  10. 3 generations backK172498 Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model…Microlife Intellectual Property GmbH · 2017
  11. 2 generations backK230075 Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model…Microlife Intellectual Property GmbH · 2023
  12. 1 generation backK251120 Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model…Microlife Corporation · 2025
  13. This deviceK253982 Korotkoff sound blood pressure monitorBeijing Hanvon Health Technology Co., Ltd. · 2026
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−12K903134Omron Auto Oscillometric Dig BP Mont/Hem-704C 705COmron Marshall Products, Inc.1991—
−11K970211Microlife Automatic Blood Pressure MonitorMicro Weiss Electronics1998—
−10K013485Microlife Upper Arm Automatic Blood Pressure Monitor, Model BP-3BTO-1, with…Microlife Corp.2002—
−9K041411Microlife Upper Arm Blood Pressure Monitor, Model BP-3BTO-APMicrolife Corp.2004—
−8K060686Microlife Automatic Blood Pressure Monitor, Model BP3AC1-1PCMicrolife Intellectual Property GmbH2006—
−7K073198Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model Watchbp…Microlife Intellectual Property GmbH2008—
−6K082357Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Modelwatchbp…Microlife Intellectual Property GmbH2008—
−5K151869Microlife Upper Arm Automatic Digital Blood Pressure MonitorMicrolife Intellectual Property GmbH, Switzerland2015—
−4K153077Microlife Upper Arm Automatic Digital Blood Pressure MonitorMicrolife Intellectual Property GmbH2016—
−3K172498Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3MW1-4YMicrolife Intellectual Property GmbH2017—
−2K230075Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3KVl-5KMicrolife Intellectual Property GmbH2023—
−1K251120Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP…Microlife Corporation2025—
0 (this device)K253982Korotkoff sound blood pressure monitorBeijing Hanvon Health Technology Co., Ltd.2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198710
199310
199510
199610
199710
199820
199910
200230
200610
201410
201510
FIG. 1 — Lineage of K253982, 1991–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DXN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DXN device

  1. #1 K222125 Arm Blood Pressure Monitor, models AOJ-30A…score 18.5 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K251120Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP Progress…
Summary page 5
…stem System, Measurement, Blood-Pressure, Non-Invasive 870.1130 DXN, DXN Predicate Trade Name (Primary Predicate is listed first) 21 CFR 807.92(a)(3) Product Code K251120 Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP Progress (BP3T01-1B) DXN Device Description Summary 21…
Microlife Corporation2025—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 35 years and passes through 7 companies; a new generation was cleared every 2.1 years on median.
  • None of the 26 ancestors has a recall.

How does its category compare?

All DXN devices

DXN System, Measurement, Blood-Pressure, Non-Invasive: FDA decisions, review times, recalls and adverse event reports over 20 years. 2026, this device’s year, is highlighted.

FDA decisions per year · DXN

28in 2025

2007: 272008: 232009: 462010: 432011: 342012: 362013: 342014: 292015: 302016: 302017: 332018: 222019: 332020: 242021: 232022: 252023: 352024: 172025: 282026 (partial year): 29
2007high 46 (2009) · low 17 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DXN

142in 2025

2007: 118 days median2008: 82 days median2009: 73 days median2010: 94 days median2011: 72 days median2012: 116 days median2013: 150 days median2014: 133 days median2015: 97 days median2016: 144 days median2017: 184 days median2018: 235 days median2019: 156 days median2020: 198 days median2021: 190 days median2022: 137 days median2023: 182 days median2024: 107 days median2025: 142 days median2026 (partial year): 147 days median
2007high 235 (2018) · low 72 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DXN

3in 2024

2008: 12009: 32011: 12013: 12016: 12019: 12024: 3
2007high 3 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DXN

239in 2025

2007: 212008: 82009: 172010: 142011: 102012: 192013: 722014: 472015: 622016: 542017: 902018: 6172019: 962020: 1302021: 1622022: 1552023: 1622024: 2702025: 239
2007high 617 (2018) · low 8 (2008)2025

MAUDE reports naming product code DXN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200727118021
2008238218
20094673317
20104394014
20113472110
201236116019
201334150172
201429133047
20153097062
201630144154
201733184090
2018222350617
201933156196
2020241980130
2021231900162
2022251370155
2023351820162
2024171073270
2025281420239
2026291470—

What is it indicated for?

The device is a digital monitor intended for use in measuring the systolic pressure and diastolic pressure, as well as the pulse rate of adults and adolescents via a non-invasive auscultatory method in medical facilities or at home, with an inflatable cuff wrapped around the upper arm for arm circumferences ranging from 22 to 42 cm. The measurement ranges of the device are systolic pressure 40 ~ 280 mmHg (supported by automatic mode for 40 to 210 mmHg…

Read the full statement (129 words)

The device is a digital monitor intended for use in measuring the systolic pressure and diastolic pressure, as well as the pulse rate of adults and adolescents via a non-invasive auscultatory method in medical facilities or at home, with an inflatable cuff wrapped around the upper arm for arm circumferences ranging from 22 to 42 cm. The measurement ranges of the device are systolic pressure 40 ~ 280 mmHg (supported by automatic mode for 40 to 210 mmHg and manual mode for 210 to 280 mmHg), diastolic pressure 20 ~ 220 mmHg, and pulse rate 40 ~ 240 beats/min.

The device can detect Irregular Heart Beat (IHB), Atrial Fibrillation, Tachycardia, and evaluate Blood Pressure Variability (BPV) and Time in Target Range (TTR) according to the measured blood pressure values.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K253982: Korotkoff sound blood pressure monitor.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K253982

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded Oct 30, 2026
  • Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026